Repeatability of 18F‐FDG PET radiomic features: A phantom study to explore sensitivity to image reconstruction settings, noise, and delineation method
Open accessThe DOI matches its Crossref record; 5 of the 5 references it cites were found in a registry or an index.
Abstract
Background: 18 F‐fluoro‐2‐deoxy‐D‐Glucose positron emission tomography (18 F‐FDG PET) radiomics has the potential to guide the clinical decision making in cancer patients, but validation is required before radiomics can be implemented in the clinical setting. The aim of this study was to explore how feature space reduction and repeatability of 18 F‐FDG PET radiomic features are affected by various sources of variation such as underlying data (e.g., object size and uptake), image reconstruction methods and settings, noise, discretization method, and delineation method. Methods: The NEMA image quality phantom was scanned with various sphere‐to‐background ratios (SBR), simulating different activity uptakes, including spheres with low uptake, that is, SBR smaller than 1. Furthermore, images of a phantom containing 3D printed inserts reflecting realistic heterogeneity uptake patterns were acquired. Data were reconstructed using various matrix sizes, reconstruction algorithms, and scan durations (noise). For every specific reconstruction and noise level, ten statistically equal replicates were generated. The phantom inserts were delineated using CT and PET‐based segmentation methods. A total of 246 radiomic features was extracted from each image dataset. Images were discretized with a fixed number of 64 bins (FBN) and a fixed bin width (FBW) of 0.25 for the high and a FBW of 0.05 for the low uptake data. In terms of feature reduction, we determined the impact of these factors on the composition of feature clusters, which were defined on the basis of Spearman's correlation matrices. To assess feature repeatability, the intraclass correlation coefficient was calculated over the ten replicates. Results: In general, larger spheres with high uptake resulted in better repeatability compared to smaller low uptake spheres. In terms of repeatability, features extracted from heterogeneous phantom inserts were comparable to features extracted from bigger high uptake spheres. For example, for an EARL‐compliant reconstruction, larger and smaller high uptake spheres yielded good repeatability for 32% and 30% of the features, while the heterogeneous inserts resulted in 34% repeatable features. For the low uptake spheres, this was the case for 22% and 20% of the features for bigger and smaller spheres, respectively. Images reconstructed with point‐spread‐function (PSF) resulted in the highest repeatability when compared with OSEM or time‐of‐flight, for example, 53%, 30%, and 32% of repeatable features, respectively (for unsmoothed data, discretized with FBN, 300 s scan duration). Reducing image noise (increasing scan duration and smoothing) and using CT‐based segmentation for the low uptake spheres yielded improved repeatability. FBW discretization resulted in higher repeatability than FBN discretization, for example, 89% and 35% of the features, respectively (for the EARL‐compliant reconstruction and larger high uptake spheres). Conclusion: Feature space reduction and repeatability of 18 F‐FDG PET radiomic features depended on all studied factors. The high sensitivity of PET radiomic features to image quality suggests that a high level of image acquisition and preprocessing standardization is required to be used as clinical imaging biomarker.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
A guideline of selecting and reporting intraclass correlation coefficients for reliability research
Koo, T. K., & Li, M. Y. (2016) · Journal of Chiropractic Medicine · doi:10.1016/j.jcm.2016.02.012 · Cited by 31,709
Decoding tumour phenotype by noninvasive imaging using a quantitative radiomics approach
Aerts, H. J. W. L., Velazquez, E. R., Leijenaar, R. T. H., Parmar, C., Grossmann, P., Carvalho, S., Bussink, J., Monshouwer, R., Haibe-Kains, B., Rietveld, D., et al. (2014) · Nature Communications · doi:10.1038/ncomms5006 · Cited by 5,132
FDG PET/CT: EANM procedure guidelines for tumour imaging: version 2.0
Boellaard, R., Delgado-Bolton, R., Oyen, W. J. G., Giammarile, F., Tatsch, K., Eschner, W., Verzijlbergen, F. J., Barrington, S. F., Pike, L. C., Weber, W. A., et al. (2015) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-014-2961-x · Cited by 3,303 · cited 1×
“Images were smoothed with Gaussian filters of 0, 2, 4, 6, and 8 mm full width at half maximum (FWHM) and were converted to SUV, so that the mean phantom background SUV was equal to 1.(20)”
MethodSelf-citationRadiomics: the process and the challenges
Kumar, V., Gu, Y., Basu, S., Berglund, A., Eschrich, S. A., Schabath, M. B., Forster, K., Aerts, H. J. W. L., Dekker, A., Fenstermacher, D., et al. (2012) · Magnetic Resonance Imaging · doi:10.1016/j.mri.2012.06.010 · Cited by 2,258 · cited 1×
“Possible solutions to reduce the feature space would be the use of principal component analysis or (hierarchical) clustering, based on correlation analysis or distance metrics.(10) Another challenge facing radiomic features is the establishment of their measurement error (i.”
MethodImaging Intratumor Heterogeneity: Role in Therapy Response, Resistance, and Clinical Outcome
O'Connor, J. P. B., Rose, C. J., Waterton, J. C., Carano, R. A. D., Parker, G. J. M., & Jackson, A. (2015) · Clinical Cancer Research · doi:10.1158/1078-0432.ccr-14-0990 · Cited by 677
Cited by
Mayerhoefer, M. E., Materka, A., Langs, G., Häggström, I., Szczypiński, P., Gibbs, P., & Cook, G. (2020) · Journal of Nuclear Medicine · doi:10.2967/jnumed.118.222893 · Cited by 1,777
Radiomics in medical imaging—“how-to” guide and critical reflection
van Timmeren, J. E., Cester, D., Tanadini-Lang, S., Alkadhi, H., & Baessler, B. (2020) · Insights into Imaging · doi:10.1186/s13244-020-00887-2 · Cited by 1,316
Machine and deep learning methods for radiomics
Avanzo, M., Wei, L., Stancanello, J., Vallières, M., Rao, A., Morin, O., Mattonen, S. A., & El Naqa, I. (2020) · Medical Physics · doi:10.1002/mp.13678 · Cited by 605
Scapicchio, C., Gabelloni, M., Barucci, A., Cioni, D., Saba, L., & Neri, E. (2021) · La radiologia medica · doi:10.1007/s11547-021-01389-x · Cited by 435
Bakas, S., Vollmuth, P., Galldiks, N., Booth, T. C., Aerts, H. J. W. L., Bi, W. L., Wiestler, B., Tiwari, P., Pati, S., Baid, U., et al. (2024) · The Lancet Oncology · doi:10.1016/s1470-2045(24)00315-2 · Cited by 36
Metadata
- Authors
- Pfaehler, E., Beukinga, R. J., de Jong, J. R., Slart, R. H. J. A., Slump, C. H., Dierckx, R. A. J. O., & Boellaard, R.
- Venue
- Medical Physics
- Publisher
- Wiley
- Published
- 2018-12-28
- Volume
- 46
- Issue
- 2
- Pages
- 665-678
- ISSN
- 0094-2405 (ISSN-L)
- 2473-4209
- OpenAlex
- W2902316253
- Cited by
- 116
- References
- 5
- FWCI
- 6.62
- Type
- journal article
- Language
- en
- Open access
- hybrid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.