Multi-centre reproducibility of diffusion MRI parameters for clinical sequences in the brain

Open access
Grech‐Sollars, Matthew, Hales, Patrick W., Miyazaki, Keiko, Raschke, Felix, Rodriguez, Daniel, Wilson, Martin, Gill, Simrandip K., Banks, Tina +13 more·2015·NMR in Biomedicine

Abstract

The purpose of this work was to assess the reproducibility of diffusion imaging, and in particular the apparent diffusion coefficient (ADC), intra‐voxel incoherent motion (IVIM) parameters and diffusion tensor imaging (DTI) parameters, across multiple centres using clinically available protocols with limited harmonization between sequences. An ice–water phantom and nine healthy volunteers were scanned across fives centres on eight scanners (four Siemens 1.5T, four Philips 3T). The mean ADC, IVIM parameters (diffusion coefficient D and perfusion fraction f ) and DTI parameters (mean diffusivity MD and fractional anisotropy FA), were measured in grey matter, white matter and specific brain sub‐regions. A mixed effect model was used to measure the intra‐ and inter‐scanner coefficient of variation (CV) for each of the five parameters. ADC, D , MD and FA had a good intra‐ and inter‐scanner reproducibility in both grey and white matter, with a CV ranging between 1% and 7.4%; mean 2.6%. Other brain regions also showed high levels of reproducibility except for small structures such as the choroid plexus. The IVIM parameter f had a higher intra‐scanner CV of 8.4% and inter‐scanner CV of 24.8%. No major difference in the inter‐scanner CV for ADC, D , MD and FA was observed when analysing the 1.5T and 3T scanners separately. ADC, D , MD and FA all showed good intra‐scanner reproducibility, with the inter‐scanner reproducibility being comparable or faring slightly worse, suggesting that using data from multiple scanners does not have an adverse effect compared with using data from the same scanner. The IVIM parameter f had a poorer inter‐scanner CV when scanners of different field strengths were combined, and the parameter was also affected by the scan acquisition resolution. This study shows that the majority of diffusion MRI derived parameters are robust across 1.5T and 3T scanners and suitable for use in multi‐centre clinical studies and trials. © 2015 The Authors NMR in Biomedicine Published by John Wiley & Sons Ltd.

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1002/nbm.3269

Cites

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • Improved harmonization of multi-site diffusion data is of great interest.

    “As a result, improved harmonization of multi-site diffusion data is of great interest (Grech-Sollars et al., 2015; Pohl et al., 2016; Palacios et al., 2017).”

    From Diffusion MRI Indices and Their Relation to Cognitive Impairment in Brain Aging: The Updated Multi-protocol Approach in ADNI3 (2019) · cites this paper as Grech-Sollars et al., 2015

    The Introduction identifies comparability of diffusion data from different scanners as an important question and discusses harmonization across sites. But the Discussion (> Volunteers; page unavailable) concludes that large multi-centre studies may be feasible without extensive protocol harmonization, rather than advocating improved harmonization. (LLM, conf=0.78)

    Experimental — accuracy not yet measured

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

The citation lists haven’t been fetched yet

References 32 · Cited by 213

Research works printed but not linked: 26

  1. [1]Stejskal EO Tanner JE Spin diffusion measurements: spin echoes in the presence of a time-dependent field gradient J. Chem. Phys 1965 42 288 292 DOI: 10.1063/1.1695690 · DOI
  2. [2]Basser PJ Mattiello J LeBihan D MR diffusion tensor spectroscopy and imaging Biophys. J 1994 66 259 267 DOI: 10.1016/S0006-3495(94)80775-1 8130344 PMC1275686 · DOI · Paper page
  3. [4]Cercignani M Bammer R Sormani MP Fazekas F Filippi M Inter-sequence and inter-imaging unit variability of diffusion tensor MR imaging histogram-derived metrics of the brain in healthy volunteers Am. J. Neuroradiol 2003 24 638 643 12695195 PMC8148683
  4. [5]Hales PW Clark CA Combined arterial spin labeling and diffusion-weighted imaging for noninvasive estimation of capillary volume fraction and permeability–surface product in the human brain J. Cereb. Blood Flow Metab 2013 33 67 75 DOI: 10.1038/jcbfm.2012.125 22990418 PMC3597361 · DOI
  5. [6]Veenith TV Carter E Grossac J Newcombe VFJ Outtrim JG Lupson V Williams GB Menon DK Coles JP Inter subject variability and reproducibility of diffusion tensor imaging within and between different imaging sessions PLoS One 2013 8 e65941 DOI: 10.1371/journal.pone.0065941 23840380 PMC3696006 · DOI
  6. [7]Pfefferbaum A Adalsteinsson E Sullivan EV Replicability of diffusion tensor imaging measurements of fractional anisotropy and trace in brain J. Magn. Reson. Imaging 2003 18 427 433 DOI: 10.1002/jmri.10377 14508779 · DOI
  7. [8]Sasaki M Yamada K Watanabe Y Matsui M Ida M Fujiwara S Shibata E Variability in absolute apparent diffusion coefficient values across different platforms may be substantial: a multivendor, multi-institutional comparison study Radiology 2008 249 624 630 DOI: 10.1148/radiol.2492071681 18936317 · DOI
  8. [9]Vollmar C O'Muircheartaigh J Barker GJ Symms MR Thompson P Kumari V Duncan JS Richardson MP Koepp MJ Identical, but not the same: intra-site and inter-site reproducibility of fractional anisotropy measures on two 3.0T scanners Neuroimage 2010 51 1384 1394 DOI: 10.1016/j.neuroimage.2010.03.046 20338248 PMC3163823 · DOI
  9. [10]Teipel SJ Reuter S Stieltjes B Acosta-Cabronero J Ernemann U Fellgiebel A Filippi M Frisoni G Hentschel F Jessen F Klöppel S Meindl T Pouwels PJW Hauenstein K-H Hampel H Multicenter stability of diffusion tensor imaging measures: a European clinical and physical phantom study Psychiatry Res 2011 DOI: 10.1016/j.pscychresns.2011.05.012 22078796 · DOI
  10. [11]Magnotta VA Matsui JT Liu D Johnson HJ Long JD Bolster BD Mueller BA Lim K Mori S Helmer KG Turner JA Reading S Lowe MJ Aylward E Flashman LA Bonett G Paulsen JS Multicenter reliability of diffusion tensor imaging Brain Connect 2012 2 345 55 DOI: 10.1089/brain.2012.0112 23075313 PMC3623569 · DOI
  11. [12]Walker L Curry M Nayak A Lange N Pierpaoli C A framework for the analysis of phantom data in multicenter diffusion tensor imaging studies Hum. Brain Mapp 2013 34 2439 2454 DOI: 10.1002/hbm.22081 22461391 PMC3458186 · DOI
  12. [13]Malyarenko D Galbán CJ Londy FJ Meyer CR Johnson TD Rehemtulla A Ross BD Chenevert TL Multi-system repeatability and reproducibility of apparent diffusion coefficient measurement using an ice–water phantom J. Magn. Reson. Imaging 2013 37 1238 1246 DOI: 10.1002/jmri.23825 23023785 PMC3548033 · DOI
  13. [14]Tofts PS Lloyd D Clark CA Barker GJ Parker GJ McConville P Baldock C Pope JM Test liquids for quantitative MRI measurements of self-diffusion coefficient in vivo Magn. Reson. Med 2000 43 368 374 10725879
  14. [17]Jenkinson M Beckmann CF Behrens TEJ Woolrich MW Smith SM FSL NeuroImage 2012 62 782 790 DOI: 10.1016/j.neuroimage.2011.09.015 21979382 · DOI · Paper page
  15. [19]Dale AM Fischl B Sereno MI Cortical surface-based analysis. I. Segmentation and surface reconstruction Neuroimage 1999 194 179 194 9931268
  16. [20]Fischl B Sereno MI Dale AM Cortical surface-based analysis II: Inflation, Flattening, Surf-based Coordinate System 1999 207 195 207 9931269
  17. [23]Pekar J Moonen CTW van Zijl PCM On the precision of diffusion/perfusion imaging by gradient sensitization Magn. Reson. Med 1992 23 122 129 1734174
  18. [25]Mori S Oishi K Jiang H Jiang L Li X Akhter K Hua K Faria AV Mahmood A Woods R Toga AW Pike GB Neto PR Evans A Zhang J Huang H Miller MI van Zijl P Mazziotta J Stereotaxic white matter atlas based on diffusion tensor imaging in an ICBM template Neuroimage 2008 40 570 582 DOI: 10.1016/j.neuroimage.2007.12.035 18255316 PMC2478641 · DOI
  19. [28]Jones DK Basser PJ “Squashing peanuts and smashing pumpkins”: how noise distorts diffusion-weighted MR data Magn. Reson. Med 2004 52 979 993 DOI: 10.1002/mrm.20283 15508154 · DOI
  20. [29]Zhou F-L Hubbard PL Eichhorn SJ Parker GJM Coaxially electrospun axon-mimicking fibers for diffusion magnetic resonance imaging ACS Appl. Mater. Interfaces 2012 4 6311 6316 DOI: 10.1021/am301919s 23135104 · DOI
  21. [30]Maier SE Sun Y Mulkern RV Diffusion imaging of brain tumors NMR Biomed 2010 23 849 864 DOI: 10.1002/nbm.1544 20886568 PMC3000221 · DOI
  22. [31]Wirestam R Borg M Brockstedt S Lindgren A Holtås S Ståhlberg F Perfusion-related parameters in intravoxel incoherent motion MR imaging compared with CBV and CBF measured by dynamic susceptibility-contrast MR technique Acta Radiol 2001 42 123 128 DOI: 10.1080/028418501127346459 11281143 · DOI
  23. [32]Lemke A Laun FB Simon D Stieltjes B Schad LR An in vivo verification of the intravoxel incoherent motion effect in diffusion-weighted imaging of the abdomen Magn. Reson. Med 2010 64 1580 1585 DOI: 10.1002/mrm.22565 20665824 · DOI
  24. [33]Leenders KL Perani D Lammertsma AA Heather JD Buckingham P Healy MJ Gibbs JM Wise RJ Hatazawa J Herold S Cerebral blood flow, blood volume and oxygen utilization. Normal values and effect of age Brain 1990 113 27 47 2302536
  25. [34]Clark CA Le Bihan D Water diffusion compartmentation and anisotropy at high b values in the human brain Magn. Reson. Med 2000 44 852 859 11108621
  26. [35]Wilson M Morgan PS Blumhardt LD Quantitative diffusion characteristics of the human brain depend on MRI sequence parameters Neuroradiology 2002 44 586 591 DOI: 10.1007/s00234-002-0797-8 12136360 · DOI

Authors
Grech‐Sollars, M., Hales, P. W., Miyazaki, K., Raschke, F., Rodriguez, D., Wilson, M., Gill, S. K., Banks, T., Saunders, D. E., Clayden, J. D., Gwilliam, M. N., Barrick, T. R., Morgan, P. S., Davies, N. P., Rossiter, J., Auer, D. P., Grundy, R., Leach, M. O., Howe, F. A., Peet, A. C., & Clark, C. A.
Affiliations
  • University College London (GB)
  • Engineering and Physical Sciences Research Council (GB)
  • University of London (GB)
  • Division of Clinical Neuroscience, School of Medicine University of Nottingham Nottingham UK
  • University of Birmingham (GB)
  • Great Ormond Street Hospital (GB)
  • CR UK and EPSRC Cancer Imaging Centre Institute of Cancer Research and Royal Marsden Foundation Trust Belmont Surrey UK
  • University of Nottingham (GB)
  • University Hospitals Birmingham NHS Foundation Trust (GB)
Corresponding author
Grech‐Sollars, Matthew
Venue
NMR in Biomedicine
Publisher
Wiley
Published
2015-03-19
Volume
28
Issue
4
Pages
468-485
ISSN
  • 0952-3480 (ISSN-L)
  • 1099-1492
OpenAlex
W1617250309
Cited by
213
References
32
Citations by year
2026: 19 · 2025: 14 · 2024: 14 · 2023: 25 · 2022: 17 · 2021: 23 · 2020: 20 · 2019: 24 · 2018: 27 · 2017: 15 · 2016: 14 · 2015: 1
FWCI
15.40
Type
journal article
Language
en
Open access
Locations
  • NMR in Biomedicine (publishedVersion, cc-by)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • St George's Online Research Archive (St George's University of London) (acceptedVersion)
  • UCL Discovery (University College London) (submittedVersion)
  • University of Birmingham Research Portal (University of Birmingham) (submittedVersion)
Funders
Cancer Research UK (C7809/A10342)
Topics
  • Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Advanced Neuroimaging Techniques and Applications
  • Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Advanced MRI Techniques and Applications
  • Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › MRI in cancer diagnosis
Keywords
apparent diffusion coefficient, intravoxel incoherent motion, diffusion tensor imaging, inter-scanner variability, white matter, perfusion fraction
MeSH terms
Adult, Body Water, Brain, Diffusion, Humans, Ice, Models, Theoretical, Motion, Water, Reproducibility of Results, Anisotropy, Phantoms, Imaging, Diffusion Magnetic Resonance Imaging, Diffusion Tensor Imaging, Neuroimaging, White Matter

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review