Confident identification of isomeric N ‐glycan structures by combined ion mobility mass spectrometry and hydrophilic interaction liquid chromatography
The DOI matches its Crossref record.
Abstract
RATIONALE: A central issue in glycan mass analysis is the ambiguity of structural assignments due to the heterogeneity and complexity of glycan structures. Ion mobility mass spectrometry (IM‐MS) has the potential to separate isomeric glycans depending on their unique collisional cross section especially when coupled with hydrophilic interaction liquid chromatography (HILIC). METHODS: Ten pyridylaminated biantennary N ‐glycans including isomeric structures were measured by electrospray ionization quadrupole‐time‐of‐flight mass spectrometry with an ion mobility phase. We investigated which adduct ions would be suitable for good separation in the ion mobility phase. The differences in observed drift time of isomeric glycans were assessed by molecular dynamics (MD) simulations in vacuum. Connecting an HILIC system with IM‐MS provided another, augmented separation mode. RESULTS: By selecting doubly protonated precursor ion species, we succeeded in separating a pair of isomeric glycans in the ion mobility phase with reasonable resolution. MD simulations of monogalactosylated glycan isomers indicate that the galactosylated Man α1‐3 branch preferentially folds back to the core chitobiose portion to form a compact structure. IM‐MS combined with HILIC resulted in even clearer separation of isomeric glycans within 15 min. CONCLUSIONS: A combination of IM‐MS with an HILIC system is eminently suitable for the confident and rapid distinction of glycan structures within a defined mixture. Copyright © 2012 John Wiley & Sons, Ltd.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Mass spectrometry approaches to glycomic and glycoproteomic analyses
Ruhaak, L. R., Xu, G., Li, Q., Goonatilleke, E., & Lebrilla, C. B. (2018) · Chemical Reviews · doi:10.1021/acs.chemrev.7b00732 · Cited by 395
Oligosaccharide Analysis by Mass Spectrometry: A Review of Recent Developments
Kailemia, M. J., Ruhaak, L. R., Lebrilla, C. B., & Amster, I. J. (2014) · Analytical Chemistry · doi:10.1021/ac403969n · Cited by 377
Zhang, P., Woen, S., Wang, T., Liau, B., Zhao, S., Chen, C., Yang, Y., Song, Z., Wormald, M. R., Yu, C., et al. (2016) · Drug Discovery Today · doi:10.1016/j.drudis.2016.01.006 · Cited by 213
Glycan analysis of therapeutic glycoproteins
Zhang, L., Luo, S., & Zhang, B. (2016) · mAbs · doi:10.1080/19420862.2015.1117719 · Cited by 207
Characterization of isomeric glycan structures by LC‐MS/MS
Veillon, L., Huang, Y., Peng, W., Dong, X., Cho, B. G., & Mechref, Y. (2017) · ELECTROPHORESIS · doi:10.1002/elps.201700042 · Cited by 163
Jooß, K., Meckelmann, S. W., Klein, J., Schmitz, O. J., & Neusüß, C. (2019) · Analytical and Bioanalytical Chemistry · doi:10.1007/s00216-018-1515-7 · Cited by 41
Metadata
- Authors
- Yamaguchi, Y., Nishima, W., Re, S., & Sugita, Y.
- Venue
- Rapid Communications in Mass Spectrometry
- Publisher
- Wiley
- Published
- 2012-11-05
- Volume
- 26
- Issue
- 24
- Pages
- 2877-2884
- ISSN
- 0951-4198 (ISSN-L)
- 1097-0231
- OpenAlex
- W2142335003
- Cited by
- 56
- References
- 0
- FWCI
- 2.13
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.