Enhancing the sensitivity of multidimensional NMR experiments by using triply-compensated π pulses
Open accessAbstract
In multidimensional solution NMR experiments, π pulses are used extensively for inversion and refocusing operations on 1H, 13C and 15N nuclei. Pulse miscalibration, off-resonance effects, and J-coupling evolution during π pulse execution result in severe signal losses that are exacerbated at high magnetic fields. Here, we report the implementation of a triply-compensated π pulse (G5) optimized for both inversion and refocusing in widely used 2- and 3-dimensional experiments. By replacing most of the hard π pulses, adiabatic or composite pulses on the 1H, 13C and 15N channels with G5 pulses, we obtained signal enhancements ranging from 80 to 240%. We anticipate that triply-compensated pulses will be crucial for improving the performance of multidimensional and multinuclear pulse sequences at ultra-high fields.
Overview
CoChat Summary
TL;DR
Replacing hard and selected shaped π pulses with the triply-compensated G5 pulse in 2D/3D NMR experiments improved signal intensities, with enhancements ranging from 80–240% depending on experiment and resonance type. The G5 pulse compensates off-resonance effects, RF miscalibration, and J-coupling evolution and is proposed as beneficial for multidimensional NMR at high fields.
Key findings
- In 2D 1H-15N HSQC, replacing five 1H hard π pulses with G5 gave 30–60% signal enhancement; additionally replacing four 15N hard π pulses gave 50–80%, with significant increases for Arg Nε resonances.
- In 2D 1H-13C HSQC, replacing all 1H hard π pulses gave 20–90% enhancement; replacing the shaped Crp80,0.5,20.1 and Crp80comp.4 pulses with G5 gave 60–130% for methyl resonances and 240% for aromatic groups. In a shorter Chsqc variant, enhancements were 120% for aromatic and 40–60% for aliphatic signals.
- In 3D TROSY-HNCA, replacing five 1H hard π pulses gave ~60% enhancement for downfield 1H resonances; replacing six 15N hard π pulses added ~20%, for an effective enhancement up to 70%.
- In 3D 13C-edited NOESY-HSQC, replacing six 1H hard, two 13C hard, and two 13C shaped π pulses with G5 gave 100% enhancement for downfield aromatic resonances and 200% for upfield methyl resonances.
Methods
- Design: The G5 pulse is a genetically optimized, triply-compensated π pulse with a total nutation angle of 5π and duration five times a single π pulse; it was used to replace 1H, 13C, and 15N hard π pulses and selected 13C shaped π pulses in standard Bruker pulse sequences, with acquisition and processing parameters kept constant.
- Samples and instruments: A U-[2H,15N,13C]-labeled 25 kDa Hsp90N sample was used at 900 MHz for 2D 1H-15N HSQC and 3D TROSY-HNCA; a 0.3 mM U-[15N,13C]-labeled 21 kDa Abl-1b RM sample was used at 850 MHz for 2D 1H-13C HSQC and 13C-edited 3D NOESY-HSQC; a 1.0 mM U-[13C15N]-labeled 42 kDa MBP sample was used for carbon HSQCs. Experiments used 5 mm TCI CryoProbes at 298 K.
- Analysis: Signal enhancements were assessed by comparing reference and G5-replaced spectra, including 1D projections and resonance-offset dependence. The G5 π pulse duration was 144 μs for 1H, and 112 μs in the 13C HSQC context; observed bandwidths were 23.5 ppm for 1H, 105 ppm for 15N, and 116 ppm for 13C.
Limitations
- Stated: G5 cannot replace Q3 pulses in HNCACB-type experiments because Q3 selectively inverts CO or CA/CB while leaving other 13C resonances unaffected; G5 performs inversion and refocusing only within its designed bandwidth and does not perform well outside it. The authors state a future goal is to design an improved pulse for Q3-type operations.
- Stated: For the States-TPPI 2D 1H-13C HSQC, solvent suppression is not as good as that of the Echo-Antiecho sequence, which uses PFG for coherence selection.
- Evident from the text: Tests were limited to specific proteins and to 850/900 MHz spectrometers, with no ultra-high-field data despite the anticipation of greater benefits; enhancements were offset- and resonance-type-dependent, so broad generalizability across all multidimensional experiments is not fully established.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
0 Valid · 2 Not valid · 20 Not checked · 39 Own results and interpretations
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“we observed 120% and 40–60% signal enhancement for the aromatic and aliphatic regions, respectively.”
§ Enhancing the Sensitivity of Multidimensional NMR Experiments by using Triply-Compensated π Pulses > Results and Discussion > 2D 1H-13C HSQC
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Furthermore, by replacing the 13C shaped π pulses in the 2D 1H-13C HSQC and 3D 13C-edited NOESY-HSQC, we achieved up to 200% signal enhancement.”
§ Enhancing the Sensitivity of Multidimensional NMR Experiments by using Triply-Compensated π Pulses > Conclusions
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Compared to the reference spectrum, we obtained signal enhancement ranging from 30–60% (Fig. 3B).”
§ Enhancing the Sensitivity of Multidimensional NMR Experiments by using Triply-Compensated π Pulses > Results and Discussion > 2D 1H-15N HSQC
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Under these conditions, the bandwidth of G5 is ~60% wider than the Q3 pulse.”
§ Enhancing the Sensitivity of Multidimensional NMR Experiments by using Triply-Compensated π Pulses > Results and Discussion > 3D 13C-edited NOESY-HSQC
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The G5 pulse has been designed to carry out inversion and refocusing only within the given bandwidth and does not perform well outside of this range (Fig. S8).”
§ Enhancing the Sensitivity of Multidimensional NMR Experiments by using Triply-Compensated π Pulses > Results and Discussion > 3D 13C-edited NOESY-HSQC
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
31 more claims
Create a free account to see all 61 claims and to verify the 42 references still open.
- It may state two separate claims as one. 20
- Probably not a substantial claim. 9
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Moreover, these imperfections accumulate in multipulse experiments, leading to low-fidelity operations and sizable signal losses (11).”
From High-fidelity control of spin ensemble dynamics via artificial intelligence: from quantum computing to NMR spectroscopy and imaging (2022) · cites this paper as [11]
In the citing paper
“However, the level of fidelity for the standard pulses is low, constituting a significant source of signal losses, particularly for multiple simultaneous operations (11, 56, 57).”
From High-fidelity control of spin ensemble dynamics via artificial intelligence: from quantum computing to NMR spectroscopy and imaging (2022) · cites this paper as [11]
In the citing paper
“In fact, the GEN-HSQC experiment has a significantly higher RF inhomogeneity compensation relative to our previous implementation of HSQC using the universal broadband pulse G5 (11).”
From High-fidelity control of spin ensemble dynamics via artificial intelligence: from quantum computing to NMR spectroscopy and imaging (2022) · cites this paper as [11]
In the citing paper
“We were able to improve the signal intensity from 6% to 23% compared to the regular CCLS version.”
From Probing Protein-Protein Interactions Using Asymmetric Labeling and Carbonyl-Carbon Selective Heteronuclear NMR Spectroscopy (2018) · cites this paper as [76]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ Enhancing the Sensitivity of Multidimensional NMR Experiments by using Triply-Compensated π Pulses > Conclusions
“by replacing the 13C shaped π pulses in the 2D 1H-13C HSQC and 3D 13C-edited NOESY-HSQC, we achieved up to 200% signal enhancement.”
§ Enhancing the Sensitivity of Multidimensional NMR Experiments by using Triply-Compensated π Pulses > Results and Discussion > 2D 1H-13C HSQC
“after replacing both Crp80,0.5,20.1 and Crp80comp.4 with the G5 pulse, we observed 60–130% signal enhancement for the methyl resonances and 240% for aromatic groups.”
Citations
Cites
Kay, L., Keifer, P., & Saarinen, T. (1992) · Journal of the American Chemical Society · doi:10.1021/ja00052a088 · Cited by 2,247
Enhancement of nuclear magnetic resonance signals by polarization transfer
Morris, G. A., & Freeman, R. (1979) · Journal of the American Chemical Society · doi:10.1021/ja00497a058 · Cited by 2,015
Khaneja, N., Reiss, T., Kehlet, C., Schulte-Herbrüggen, T., & Glaser, S. J. (2005) · Journal of Magnetic Resonance · doi:10.1016/j.jmr.2004.11.004 · Cited by 1,854 · cited 1×
“1989) or shaped π pulses (Temps and Brewer 1984; Bauer et al. 1984; Emsley and Bodenhausen 1990, 1992; Geen and Freeman 1991; Bohlen and Bodenhausen 1993; Hwang et al. 1997; Khaneja et al. 2005), designed for either inversion or refocusing, represent alternative solutions.”
BackgroundKay, L. E., Torchia, D. A., & Bax, A. (1989) · Biochemistry · doi:10.1021/bi00449a003 · Cited by 1,768
Palmer, A. G., Cavanagh, J., Wright, P. E., & Rance, M. (1991) · Journal of Magnetic Resonance (1969) · doi:10.1016/0022-2364(91)90036-s · Cited by 948
Gradient-Enhanced Triple-Resonance Three-Dimensional NMR Experiments with Improved Sensitivity
Muhandiram, D. R., & Kay, L. E. (1994) · Journal of Magnetic Resonance, Series B · doi:10.1006/jmrb.1994.1032 · Cited by 892
Wittekind, M., & Mueller, L. (1993) · Journal of Magnetic Resonance, Series B · doi:10.1006/jmrb.1993.1033 · Cited by 859
Band-selective radiofrequency pulses
Geen, H., & Freeman, R. (1991) · Journal of Magnetic Resonance (1969) · doi:10.1016/0022-2364(91)90034-q · Cited by 828
A general enhancement scheme in heteronuclear multidimensional NMR employing pulsed field gradients
Schleucher, J., Schwendinger, M., Sattler, M., Schmidt, P., Schedletzky, O., Glaser, S. J., S�rensen, O. W., & Griesinger, C. (1994) · Journal of Biomolecular NMR · doi:10.1007/bf00175254 · Cited by 806
Levitt, M. H. (1986) · Progress in Nuclear Magnetic Resonance Spectroscopy · doi:10.1016/0079-6565(86)80005-x · Cited by 776
Zhang, O., Kay, L. E., Olivier, J. P., & Forman-Kay, J. D. (1994) · Journal of Biomolecular NMR · doi:10.1007/bf00398413 · Cited by 648
Radiofrequency pulse sequences which compensate their own imperfections
Freeman, R., Kempsell, S. P., & Levitt, M. H. (1980) · Journal of Magnetic Resonance (1969) · doi:10.1016/0022-2364(80)90327-3 · Cited by 370
(2018) · Springer eBooks · doi:10.1007/978-3-319-28388-3 · Cited by 354
Emsley, L., & Bodenhausen, G. (1990) · Chemical Physics Letters · doi:10.1016/0009-2614(90)87025-m · Cited by 345
Broadband, Narrowband, and Passband Composite Pulses for Use in Advanced NMR Experiments
Wimperis, S. (1994) · Journal of Magnetic Resonance, Series A · doi:10.1006/jmra.1994.1159 · Cited by 311
Pervushin, K., Riek, R., Wider, G., & Wüthrich, K. (1998) · Journal of the American Chemical Society · doi:10.1021/ja980742g · Cited by 310
Emsley, L., & Bodenhausen, G. (1992) · Journal of Magnetic Resonance (1969) · doi:10.1016/0022-2364(92)90242-y · Cited by 289
Experimental Aspects of Chirp NMR Spectroscopy
Böhlen, J.-M., & Bodenhausen, G. (1993) · Journal of Magnetic Resonance Series A · doi:10.1006/jmra.1993.1107 · Cited by 233
Broadband Population Inversion
Tycko, R. (1983) · Physical Review Letters · doi:10.1103/physrevlett.51.775 · Cited by 196
Bauer, C. J., Freeman, R., Frenkiel, T. A., Keeler, J., & Shaka, A. J. (1984) · Journal of Magnetic Resonance (1969) · doi:10.1016/0022-2364(84)90148-3 · Cited by 122
Composite pulses for ultra-broadband spin inversion
Shaka, A. J. (1985) · Chemical Physics Letters · doi:10.1016/0009-2614(85)87040-8 · Cited by 106
Broadband Adiabatic Refocusing without Phase Distortion
Hwang, T., van Zijl, P. C. M., & Garwood, M. (1997) · Journal of Magnetic Resonance · doi:10.1006/jmre.1996.1049 · Cited by 105
Iterative schemes for broad-band and narrow-band population inversion in NMR
Tycko, R., & Pines, A. (1984) · Chemical Physics Letters · doi:10.1016/0009-2614(84)85541-4 · Cited by 100
Radiofrequency pulse sequences which compensate their own imperfections
Freeman, R., Kempsell, S. P., & Levitt, M. H. (2011) · Journal of Magnetic Resonance · doi:10.1016/j.jmr.2011.08.032 · Cited by 76
Broadband population inversion in solid state NMR
Tycko, R., Schneider, E. N., & Pines, A. (1984) · The Journal of Chemical Physics · doi:10.1063/1.447749 · Cited by 67
Synthesis of arbitrary frequency domain transmitting pulses applicable to pulsed NMR instruments
Temps, A. J., & Brewer, C. F. (1984) · Journal of Magnetic Resonance (1969) · doi:10.1016/0022-2364(84)90300-7 · Cited by 42
Dual-compensated antisymmetric composite refocusing pulses for NMR
Odedra, S., Thrippleton, M. J., & Wimperis, S. (2012) · Journal of Magnetic Resonance · doi:10.1016/j.jmr.2012.10.003 · Cited by 31
Genetic algorithm optimized triply compensated pulses in NMR spectroscopy
Manu, V. S., & Veglia, G. (2015) · Journal of Magnetic Resonance · doi:10.1016/j.jmr.2015.09.010 · Cited by 24
Self-citationNielsen, N. C., Bildsøe, H., JAKOBSEN, H. J., & Sørensen, O. W. (1989) · Journal of Magnetic Resonance (1969) · doi:10.1016/0022-2364(89)90149-2 · Cited by 18
Manu, V. S., & Veglia, G. (2016) · Journal of Magnetic Resonance · doi:10.1016/j.jmr.2016.09.021 · Cited by 13
Self-citationSubrahmanian, M. V., Dregni, A. J., & Veglia, G. (2018) · Springer eBooks · doi:10.1007/978-3-319-28388-3_71 · Cited by 3
Self-citationXia, Y., Rossi, P., Tonelli, M., Huang, C., Kalodimos, C. G., & Veglia, G. (2017) · Journal of Biomolecular NMR · doi:10.1007/s10858-017-0133-6 · Cited by 2
Self-citationSubrahmanian, M. V., Dregni, A. J., & Veglia, G. (2017) · Modern Magnetic Resonance · doi:10.1007/978-3-319-28275-6_71-1
Self-citation
Research works printed but not linked: 1
- Cavanagh J, Fairbrother W, Palmer A III, Rance M, Skelton N (2007)Protein NMR spectroscopy. Principles and practice, 2nd edn. Elselvier Academic Press, New York, pp 796–813
Cited by
Alik, A., Bouguechtouli, C., Julien, M., Bermel, W., Ghouil, R., Zinn‐Justin, S., & Theillet, F. (2020) · Angewandte Chemie International Edition · doi:10.1002/anie.202002288 · Cited by 31
Manu, V. S., Olivieri, C., Pavuluri, K., & Veglia, G. (2022) · Physical Chemistry Chemical Physics · doi:10.1039/d2cp01744j · Cited by 25
Self-citationSubrahmanian, M. V., Pavuluri, K., Olivieri, C., & Veglia, G. (2022) · PNAS Nexus · doi:10.1093/pnasnexus/pgac133 · Cited by 24
Self-citationComposite pulses with errant phases
Torosov, B. T., & Vitanov, N. V. (2019) · Physical Review A · doi:10.1103/physreva.100.023410 · Cited by 23
Zhan, H., Huang, Y., & Chen, Z. (2026) · Applied Physics Reviews · doi:10.1063/5.0277355 · Cited by 22
Metadata
- Authors
- Xia, Y., Rossi, P., Subrahmanian, M. V., Huang, C., Saleh, T., Olivieri, C., Kalodimos, C. G., & Veglia, G.
- Affiliations
- University of Minnesota (US)
- Corresponding author
- Kalodimos, Charalampos G.
- Veglia, Gianluigi
- ORCID
- Venue
- Journal of Biomolecular NMR
- Published
- 2017-11-21
- Volume
- 69
- Issue
- 4
- Pages
- 237-243
- ISSN
- 0925-2738 (ISSN-L)
- 1573-5001
- PMC
- PMC5738291
- OpenAlex
- W2768348132
- Semantic Scholar
- ffb22fc61acf3d02309d8b5f6aaa362d4acb6ef7
- Cited by
- 20
- References
- 33
- Citations by year
- 2026: 2 · 2025: 4 · 2024: 1 · 2023: 2 · 2022: 4 · 2021: 1 · 2020: 3 · 2019: 2 · 2018: 1
- FWCI
- 0.71
- Type
- journal article
- Language
- en
- Open access
- green
- europepmc.org
- Licence
- www.springer.com/tdm (tdm)
- www.springer.com/tdm (vor)
- Locations
- Journal of Biomolecular NMR (publishedVersion)
- PubMed (publishedVersion)
- Europe PMC (PubMed Central) (acceptedVersion)
- PubMed Central (acceptedVersion)
- Funders
- National Institute of General Medical Sciences (GM100310)
- National Institute on Aging (AI094623)
- Topics
- Physical Sciences › Chemistry › Spectroscopy › Advanced NMR Techniques and Applications
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Advanced MRI Techniques and Applications
- Life Sciences › Biochemistry, Genetics and Molecular Biology › Molecular Biology › Metabolomics and Mass Spectrometry Studies
- Keywords
- multidimensional NMR, signal amplification, off-resonance effects, inversion, refocusing, adiabatic pulses, composite pulses, ultra-high field NMR
- MeSH terms
- Nuclear Magnetic Resonance, Biomolecular
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
A language model drew the claims from the paper on 2026-11-18, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-11-18 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.