Minimizing Back Exchange in the Hydrogen Exchange-Mass Spectrometry Experiment
Open accessThe DOI matches its Crossref record; 28 of the 28 references it cites were found in a registry or an index.
Abstract
The addition of mass spectrometry (MS) analysis to the hydrogen exchange (HX) proteolytic fragmentation experiment extends powerful HX methodology to the study of large biologically important proteins. A persistent problem is the degradation of HX information due to back exchange of deuterium label during the fragmentation-separation process needed to prepare samples for MS measurement. This paper reports a systematic analysis of the factors that influence back exchange (solution pH, ionic strength, desolvation temperature, LC column interaction, flow rates, system volume). The many peptides exhibit a range of back exchange due to intrinsic amino acid HX rate differences. Accordingly, large back exchange leads to large variability in D-recovery from one residue to another as well as one peptide to another that cannot be corrected for by reference to any single peptide-level measurement. The usual effort to limit back exchange by limiting LC time provides little gain. Shortening the LC elution gradient by 3-fold only reduced back exchange by ~2%, while sacrificing S/N and peptide count. An unexpected dependence of back exchange on ionic strength as well as pH suggests a strategy in which solution conditions are changed during sample preparation. Higher salt should be used in the first stage of sample preparation (proteolysis and trapping) and lower salt (<20 mM) and pH in the second stage before electrospray injection. Adjustment of these and other factors together with recent advances in peptide fragment detection yields hundreds of peptide fragments with D-label recovery of 90% ± 5%.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Primary structure effects on peptide group hydrogen exchange
Bai, Y., Milne, J. S., Mayne, L., & Englander, S. W. (1993) · Proteins: Structure, Function, and Bioinformatics · doi:10.1002/prot.340170110 · Cited by 1,902 · cited 2×
“[34]), and has been directly measured more recently [36].”
Self-citationZhang, Z., & Smith, D. L. (1993) · Protein Science · doi:10.1002/pro.5560020404 · Cited by 1,012 · cited 1×
“Hydrogen exchange investigations of larger and biologically more interesting protein systems can be achieved by a proteolytic fragmentation method [1] followed by mass spectrometry analysis [2–5].”
Hydrogen exchange mass spectrometry for studying protein structure and dynamics
Konermann, L., Pan, J., & Liu, Y.-H. (2011) · Chem. Soc. Rev. · doi:10.1039/c0cs00113a · Cited by 838 · cited 2×
“Hydrogen exchange investigations of larger and biologically more interesting protein systems can be achieved by a proteolytic fragmentation method [1] followed by mass spectrometry analysis [2–5].”
Primary structure effects on peptide group hydrogen exchange
Molday, R. S., Englander, S. W., & Kallen, R. G. (1972) · Biochemistry · doi:10.1021/bi00752a003 · Cited by 471 · cited 2×
“[34]), and has been directly measured more recently [36].”
Self-citationIsotope effects in peptide group hydrogen exchange
Connelly, G. P., Bai, Y., Jeng, M., & Englander, S. W. (1993) · Proteins: Structure, Function, and Bioinformatics · doi:10.1002/prot.340170111 · Cited by 376 · cited 6×
“The earlier HX rate calibrations [9, 33] were done in high salt (0.”
InfluentialSelf-citationRand, K. D., Zehl, M., Jensen, O. N., & Jørgensen, T. J. D. (2009) · Analytical Chemistry · doi:10.1021/ac9008447 · Cited by 220 · cited 1×
“A related HX MS method uses electron transfer or capture dissociation to produce a series of protein fragments that differ by only one terminal residue [3, 10–20].”
Pan, J., Han, J., Borchers, C. H., & Konermann, L. (2009) · Journal of the American Chemical Society · doi:10.1021/ja904379w · Cited by 201 · cited 1×
“A related HX MS method uses electron transfer or capture dissociation to produce a series of protein fragments that differ by only one terminal residue [3, 10–20].”
Hydrogen exchange mass spectrometry: what is it and what can it tell us?
Marcsisin, S. R., & Engen, J. R. (2010) · Analytical and Bioanalytical Chemistry · doi:10.1007/s00216-010-3556-4 · Cited by 189
Rand, K. D., Adams, C. M., Zubarev, R. A., & Jørgensen, T. J. D. (2008) · Journal of the American Chemical Society · doi:10.1021/ja076448i · Cited by 189 · cited 1×
“A related HX MS method uses electron transfer or capture dissociation to produce a series of protein fragments that differ by only one terminal residue [3, 10–20].”
Zehl, M., Rand, K. D., Jensen, O. N., & Jørgensen, T. J. D. (2008) · Journal of the American Chemical Society · doi:10.1021/ja805573h · Cited by 172 · cited 1×
“A related HX MS method uses electron transfer or capture dissociation to produce a series of protein fragments that differ by only one terminal residue [3, 10–20].”
Cited by
James, E. I., Murphree, T. A., Vorauer, C., Engen, J. R., & Guttman, M. (2022) · Chemical Reviews · doi:10.1021/acs.chemrev.1c00279 · Cited by 305
The diverse and expanding role of mass spectrometry in structural and molecular biology
Lössl, P., van de Waterbeemd, M., & Heck, A. J. (2016) · The EMBO Journal · doi:10.15252/embj.201694818 · Cited by 222
“In view of this obstacle, it is clear that increasing speed and efficiency of the bottom-up analysis has been a key to make HDX-MS a broadly applicable strategy (Konermann et al , 2011; Walters et al , 2012).”
A protein fold switch joins the circadian oscillator to clock output in cyanobacteria
Chang, Y.-G., Cohen, S. E., Phong, C., Myers, W. K., Kim, Y.-I., Tseng, R., Lin, J., Zhang, L., Boyd, J. S., Lee, Y., et al. (2015) · Science · doi:10.1126/science.1260031 · Cited by 212
Petrie, E. J., Sandow, J. J., Jacobsen, A. V., Smith, B. J., Griffin, M. D. W., Lucet, I. S., Dai, W., Young, S. N., Tanzer, M. C., Wardak, A., et al. (2018) · Nature Communications · doi:10.1038/s41467-018-04714-7 · Cited by 189
“f Liposomes containing self-quenching dye 5(6)-carboxyfluorescein were added to 1 μM of wild-type hMLKL (dark blue), wild-type hMLKL+ATP (red), E351K hMLKL (orange), E351K hMLKL+ATP (light blue), and hMLKL(2-154) (green), and dye release was monitored by spectrophotometry at 485 nm over 30min.”
MethodStepwise protein folding at near amino acid resolution by hydrogen exchange and mass spectrometry
Hu, W., Walters, B. T., Kan, Z.-Y., Mayne, L., Rosen, L. E., Marqusee, S., & Englander, S. W. (2013) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.1305887110 · Cited by 172
“Previous papers describe methods for obtaining and analyzing many fragments (10) and for minimizing back exchange of the D-label during sample preparation (11).”
BackgroundMethodInfluentialSelf-citationKoenig, P., Lee, C. V., Walters, B. T., Janakiraman, V., Stinson, J., Patapoff, T. W., & Fuh, G. (2017) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.1613231114 · Cited by 107
Self-citationChetty, P. S., Mayne, L., Lund-Katz, S., Englander, S. W., & Phillips, M. C. (2017) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.1617523114 · Cited by 58
Self-citationThe Role of Mass Spectrometry in Structural Studies of Flavin-Based Electron Bifurcating Enzymes
Tokmina-Lukaszewska, M., Patterson, A., Berry, L., Scott, L., Balasubramanian, N., & Bothner, B. (2018) · Frontiers in Microbiology · doi:10.3389/fmicb.2018.01397 · Cited by 15
Metadata
- Authors
- Walters, B. T., Ricciuti, A., Mayne, L., & Englander, S. W.
- Venue
- Journal of the American Society for Mass Spectrometry
- Publisher
- American Chemical Society (ACS)
- Published
- 2012-09-11
- Volume
- 23
- Issue
- 12
- Pages
- 2132-2139
- ISSN
- 1044-0305 (ISSN-L)
- OpenAlex
- W2059003438
- Cited by
- 182
- References
- 28
- FWCI
- 3.99
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.