Specific absorption rate and tissue temperature in local hyperthermia

Griffiths, H., Ahmed, A., Smith, C.W., Moore, J.L., Kerby, I.J., Davies, R.M.E.·1986·International Journal of Radiation Oncology*Biology*Physics

Abstract

Specific absorption rate (SAR) and tissue temperature were measured for a total of 83 treatments in 33 patients who received local hyperthermia treatment for cancer. The patients were grouped into three categories according to tumor size. Hyperthermia was induced by 13.56 MHz electromagnetic energy applied using capacitive coupling. A method is described for evaluating SAR from the tissue temperature traces at any time in the treatment when a step change is made in applied power. The method is possible only if the temperature traces are free from interference and the total power delivered to the patient is monitored. Mean values of SAR ranged from 4.6 to 89 W kg-1 depending on the treatment site. Satisfactory heating was achieved for superficial tumors, with temperatures greater than 42 degrees C being recorded in 69% of treatments. For axillary nodes only 4% of treatments exceeded 42 degrees C. For cervix tumors an idealized tumor model was used to estimate tumor temperature from the temperature and SAR measured in the adjacent normal tissue. From the model it appears necessary either to raise the systemic temperature to 40 degrees C or to increase the SAR by at least a factor of 4 to obtain a temperature of 42 degrees C in a typical tumor. Measurements of SAR and temperature are essential for feedback control of computer models which, in principle, could provide a complete distribution of temperature during a hyperthermia treatment. Furthermore, measured SAR provides a direct comparison of the power deposition from different treatment machines in a clinical environment. The data presented form a basis for comparison with the clinical use of other heating systems.

Overview

Integrity signals

Registry

Crossref · 10.1016/0360-3016(86)90137-9

Cites

16 of 16 references verified0 citation errors · 0 not checkable

Statistics

Recomputing p-values needs the paper’s full text.

Verification

Claims the paper makes

We couldn’t retrieve this paper’s full text, so its claims and self-check weren’t run. The registry and citation checks still apply.

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Citations

Cites

Flagged first, then by citations

Show all 16 references

Research works printed but not linked: 2

  1. Kerby, 1984, RF hyperthermia in the treatment of carcinoma cervix uteri, Vol. 1, 757
  2. Roemer, 1984, Thermal modeling, Vol. 11, 293

Cited by

Sorted by citations

See all 11

Metadata

Authors
Griffiths, H., Ahmed, A., Smith, C. W., Moore, J. L., Kerby, I. J., & Davies, R. M. E.
Affiliations
Velindre Cancer Centre (GB)
Venue
International Journal of Radiation Oncology*Biology*Physics
Publisher
Elsevier BV
Published
1986-11
Volume
12
Issue
11
Pages
1997-2002
ISSN
0360-3016 (ISSN-L)
OpenAlex
W2048758955
Cited by
11
References
16
Citations by year
2024: 2 · 2022: 1 · 2021: 1 · 2020: 1
FWCI
1.51
Type
journal article
Language
en
Open access
closed
Locations
  • International Journal of Radiation Oncology*Biology*Physics (publishedVersion)
  • PubMed (publishedVersion)
Topics
  • Physical Sciences › Engineering › Biomedical Engineering › Ultrasound and Hyperthermia Applications
  • Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Infrared Thermography in Medicine
  • Health Sciences › Medicine › Hepatology › Hepatocellular Carcinoma Treatment and Prognosis
Keywords
specific absorption rate, tissue temperature, local hyperthermia, cancer hyperthermia, capacitive coupling, tumor size, axillary lymph nodes, cervical cancer
MeSH terms
Absorption, Body Temperature, Humans, Hyperthermia, Induced, Neoplasms

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.