The effect of a polyurethane coating incorporating both a thrombin inhibitor and nitric oxide on hemocompatibility in extracorporeal circulation
Open accessAbstract
Nitric oxide (NO) releasing (NORel) materials have been extensively investigated to create localized increases in NO concentration by the proton driven diazeniumdiolate-containing polymer coatings and demonstrated to improve extracorporeal circulation (ECC) hemocompatibility. In this work, the NORel polymeric coating composed of a diazeniumdiolated dibutylhexanediamine (DBHD-N2O2)-containing hydrophobic Elast-eon™ (E2As) polyurethane was combined with a direct thrombin inhibitor, argatroban (AG), and evaluated in a 4 h rabbit thrombogenicity model without systemic anticoagulation. In addition, the immobilizing of argatroban to E2As polymer was achieved by either a polyethylene glycol-containing (PEGDI) or hexane methylene (HMDI) diisocyanate linker. The combined polymer film was coated on the inner walls of ECC circuits to yield significantly reduced ECC thrombus formation compared to argatroban alone ECC control after 4 h blood exposure (0.6 ± 0.1 AG/HMDI/NORel vs 1.7 ± 0.2 cm(2) AG/HMDI control). Platelet count (2.8 ± 0.3 AG/HMDI/NORel vs 1.9 ± 0.1 × 10(8)/ml AG/HMDI control) and plasma fibrinogen levels were preserved after 4 h blood exposure with both the NORel/argatroban combination and the AG/HMDI control group compared to baseline. Platelet function as measured by aggregometry remained near normal in both the AG/HMDI/NORel (63 ± 5%) and AG/HMDI control (58 ± 7%) groups after 3 h compared to baseline (77 ± 1%). Platelet P-selectin mean fluorescence intensity (MFI) as measured by flow cytometry also remained near baseline levels after 4 h on ECC to ex vivo collagen stimulation (16 ± 3 AG/HMDI/NORel vs 11 ± 2 MFI baseline). These results suggest that the combined AG/HMDI/NORel polymer coating preserves platelets in blood exposure to ECCs to a better degree than AG/PEGDI/NORel, NORel alone or AG alone. These combined antithrombin, NO-mediated antiplatelet effects were shown to improve thromboresistance of the AG/HMDI/NORel polymer-coated ECCs and move potential nonthrombogenic polymers closer to mimicking vascular endothelium.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Combined AG/DBHD polymer coatings better prevent thrombus formation after 4 h of blood exposure compared with control ECCs or ECCs coated with only DBHD/N2O2 or AG alone.
“The results showed that the combined AG/DBHD polymer coatings can better prevent thrombus formation after 4 h of blood exposure compared to control ECCs or ECCs with coatings containing only DBHD/N2O2 or AG alone.43”
From Recent advances in thromboresistant and antimicrobial polymers for biomedical applications: just say yes to nitric oxide (NO) (2016) · cites this paper as [43]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
The citation lists haven’t been fetched yet
References 50 · Cited by 83
Research works printed but not linked: 2
- [31]Noguchi H Iwata H Ikada Y Synthesis of monomeric and polymeric conjugates carrying a thrombin inhibitor through an ester bond J Biomed Mater Res 1998 39 621 9 9492224
- [42]Annich GM Meinhardt JP Mowery KA Ashton BA Merz SI Hirschl RB Reduced platelet activation and thrombosis in extracorporeal circuits coated with nitric oxide release polymers Crit Care Med 2000 28 915 20 10809259
- Authors
- Major, T. C., Brisbois, E. J., Jones, A. M., Zanetti, M. E., Annich, G. M., Bartlett, R. H., & Handa, H.
- Affiliations
- University of Michigan (US)
- Corresponding author
- Major, Terry C.
- Handa, Hitesh
- Venue
- Biomaterials
- Publisher
- Elsevier BV
- Published
- 2014-08
- Volume
- 35
- Issue
- 26
- Pages
- 7271-7285
- ISSN
- 0142-9612 (ISSN-L)
- PMC
- PMC4099065
- OpenAlex
- W2014082100
- Cited by
- 83
- References
- 50
- Citations by year
- 2026: 4 · 2025: 1 · 2024: 3 · 2023: 7 · 2022: 10 · 2021: 6 · 2020: 10 · 2019: 8 · 2018: 9 · 2017: 4 · 2016: 15 · 2015: 5 · 2014: 1
- FWCI
- 3.28
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Locations
- Biomaterials (publishedVersion)
- PubMed (publishedVersion)
- PubMed Central (acceptedVersion)
- Funders
- National Institutes of Health (R21EB016236, K25HL111213)
- Federal Drug Administration (2P50FD003787-03)
- Topics
- Physical Sciences › Materials Science › Surfaces, Coatings and Films › Polymer Surface Interaction Studies
- Physical Sciences › Engineering › Biomedical Engineering › Mechanical Circulatory Support Devices
- Health Sciences › Medicine › Physiology › Nitric Oxide and Endothelin Effects
- Keywords
- hemocompatibility, extracorporeal circulation, diazeniumdiolates, argatroban, thrombin inhibitors, polyurethane coatings, antithrombotic coatings, platelet storage, P-selectin, fibrinogen, thrombus formation, polyethylene glycol linker
- MeSH terms
- Animals, Antithrombins, Arginine, Blood Coagulation, Blood Platelets, Extracorporeal Circulation, Nitric Oxide, Pipecolic Acids, Platelet Function Tests, Polyethylene Glycols, Polymers, Polyurethanes, Rabbits, Sulfonamides, Thrombin, Thrombosis, Platelet Activation, Isocyanates, Coated Materials, Biocompatible
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review