Monitoring the stability of the standardization status of FT4 and TSH assays by use of daily outpatient medians and flagging frequencies

Open access
De Grande, Linde A.C., Goossens, Kenneth, Van Uytfanghe, Katleen, Das, Barnali, MacKenzie, Finlay, Patru, Maria-Magdalena, Thienpont, Linda M.·2017·Clinica Chimica Acta

Abstract

Clinicians diagnose thyroid dysfunction based on TSH and FT4 testing. However, the current lack of comparability between assays limits the optimal use of laboratory data. The International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) gave a mandate to the Committee for Standardization of Thyroid Function Tests (C-STFT) to resolve this limitation by standardization. Recently, the Committee members and their partners felt ready to set the step towards the technical recalibration. However, before implementation, they were furthered by the Food and Drugs Administration (FDA) to develop a tool to assess the sustainability of the new calibration basis. C-STFT began to use 2 online applications, i.e., the "Percentiler" and "Flagger", with the intention to assess their utility for this purpose. The tools monitor the course of instrument-specific moving medians of outpatient results (Percentiler) and flagging rates (Flagger) from data of individual laboratories grouped by instrument/assay peer. They additionally document the mid- to long-term medians, hence, are quality indicators of stability of performance of both laboratories and peers/assays. Here, the first experiences built up in the pre-standardization phase are reported. They suggest the suitability of both applications to document the sustainability of the calibration basis in the post-standardization phase.

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1016/j.cca.2016.04.032

Cites

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

The citation lists haven’t been fetched yet

References 24 · Cited by 15

Authors
De Grande, L. A. C., Goossens, K., Van Uytfanghe, K., Das, B., MacKenzie, F., Patru, M.-M., & Thienpont, L. M.
Affiliations
  • Ghent University (BE)
  • Kokilaben Dhirubhai Ambani Hospital (IN)
  • University Hospitals Birmingham NHS Foundation Trust (GB)
Corresponding author
Thienpont, Linda M.
Venue
Clinica Chimica Acta
Publisher
Elsevier BV
Published
2017-04
Volume
467
Pages
8-14
ISSN
0009-8981 (ISSN-L)
OpenAlex
W2344007603
Cited by
15
References
24
Citations by year
2026: 1 · 2025: 1 · 2023: 2 · 2022: 1 · 2021: 1 · 2020: 1 · 2019: 2 · 2018: 2 · 2017: 3 · 2016: 1
FWCI
2.14
Type
journal article
Language
en
Open access
Locations
  • Clinica Chimica Acta (publishedVersion)
  • PubMed (publishedVersion)
  • Ghent University Academic Bibliography (Ghent University) (submittedVersion, other-oa)
Topics
  • Health Sciences › Medicine › Physiology › Clinical Laboratory Practices and Quality Control
  • Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Thyroid Disorders and Treatments
  • Social Sciences › Decision Sciences › Statistics, Probability and Uncertainty › Scientific Measurement and Uncertainty Evaluation
Keywords
assay standardization, assay comparability, moving median, laboratory quality indicators
MeSH terms
Blood Chemical Analysis, Humans, Outpatients, Reference Standards, Thyrotropin, Thyroxine, Time Factors

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review