Elecsys® Total-Tau and Phospho-Tau (181P) CSF assays: Analytical performance of the novel, fully automated immunoassays for quantification of tau proteins in human cerebrospinal fluid

Open access
Lifke, Valeria, Kollmorgen, Gwendlyn, Manuilova, Ekaterina, Oelschlaegel, Tobias, Hillringhaus, Lars, Widmann, Monika, von Arnim, Christine A.F., Otto, Markus, Christenson, Robert H., Powers, Jennifer L., Shaw, Leslie M., Hansson, Oskar, Doecke, James D., Li, Qiao-Xin, Teunissen, Charlotte, Tumani, Hayrettin, Blennow, Kaj · 2019 · Clinical Biochemistry
Cited by 91

The DOI matches its Crossref record; 14 of the 14 references it cites were found in a registry or an index.

Abstract

Total tau (tTau) and phosphorylated 181P tau (pTau) are supportive diagnostic cerebrospinal fluid (CSF) biomarkers for Alzheimer's disease. Manual CSF tau assays are limited by lot-to-lot and between-laboratory variability and long incubation/turnaround times. Elecsys® Total-Tau CSF and Phospho-Tau (181P) CSF immunoassays were developed for fully automated cobas e analyzers, allowing broader access in clinical practice and trials. Analytical performance, reproducibility, method comparisons with commercially available assays, and lot-to-lot and platform comparability (cobas e 601/411) of the Elecsys® CSF assays were assessed. Tau distributions and concentration ranges were evaluated in CSF samples from two clinical cohorts. Both assays showed high sensitivity (limit of quantitation [LoQ]: 63 pg/mL [tTau]; 4 pg/mL [pTau]) and linearity over the measuring range (80–1300 pg/mL; 8–120 pg/mL), which covered the entire concentration range measured in clinical samples. Lot-to-lot and platform comparability demonstrated good consistency (Pearson's r: 0.998; 1.000). Multicenter evaluation coefficients of variation (CVs): repeatability, < 1.8%; intermediate precision, < 2.8%; between-laboratory variability, < 2.7% (both assays); and total reproducibility, < 6.7% (tTau) and < 4.7% (pTau). Elecsys® CSF assays demonstrated good correlation with commercially available tau assays. Elecsys® Total-Tau CSF and Phospho-Tau (181P) CSF assays demonstrate good analytical performance with clinically relevant measuring ranges; data support their use in clinical trials and practice.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1016/j.clinbiochem.2019.05.005

Cites

14 of 14 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 14 references

Cited by

Metadata

Authors
Lifke, V., Kollmorgen, G., Manuilova, E., Oelschlaegel, T., Hillringhaus, L., Widmann, M., von Arnim, C. A. F., Otto, M., Christenson, R. H., Powers, J. L., Shaw, L. M., Hansson, O., Doecke, J. D., Li, Q.-X., Teunissen, C., Tumani, H., & Blennow, K.
Venue
Clinical Biochemistry
Publisher
Elsevier BV
Published
2019-10
Volume
72
Pages
30-38
ISSN
0009-9120 (ISSN-L)
OpenAlex
W2945475774
Cited by
91
References
14
FWCI
6.57
Type
journal article
Language
en
Open access
hybrid

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.