Assessment of two glycated hemoglobin immunoassays
Abstract
Introduction Glycated hemoglobin (HbA1c) level reflects chronic glycemic status if reliable tests are used, however, in some regions worldwide high performing assays might not be readily available. This study aimed to asses two HbA1c immunoassays, comparing them with high-performance liquid chromatography (HPLC) assay, three methods available in Ecuador. Material and methods HbA1c were measured in 114 fresh whole blood-samples by DCA-Vantage point-of-care analyzer, I-Chroma portable fluorescent scanner immunoassay and BioRad Variant II Turbo HPLC. Normal and pathological HbA1c ranges were included. Blood samples with variants of hemoglobin were excluded. HbA1c values were expressed in National Glycohemoglobin Standardization Program percentages and mmol/mol, as mean±standard deviation. Results HbA1c results by HPLC and DCA-Vantage were similar: 6.3±1.7% (45±18.6mmol/mol) vs. 6.3±1.8% (45±19.7mmol/mol), respectively, P=0.057; while HbA1c values by I-Chroma were lower than HPLC, 5.8±1.9% (40±20.8mmol/mol), P Conclusion HbA1c immunoassays DCA-Vantage was comparable to HPLC assay, showing good correlation, appropriate precision and low bias, whereas I-Chroma assay was precise but inaccurate. Therefore, DCA-Vantage has better performance than I-Chroma. These findings suggest that is convenient to assess the HbA1c immunoassays commercially available in our country, Ecuador.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The DCA Vantage Analyzer point-of-care immunoassay has shown sufficient correlation with the reference method, high-performance liquid chromatography.
“We used capillary blood samples and equipment to conduct point-of-care immunoassay (DCA Vantage Analyzer, Siemens) which has shown sufficient correlation with the reference method, high-performance liquid chromatography.39”
The systematic review and meta-analysis included 11 studies involving 1,593 participants and 3,285 paired HbA1c measurements comparing POC devices with reference HPLC-based laboratory methods.
“A total of 11 studies involving 1,593 participants and 3,285 paired measurements were included.”
From Comparison of point‑of‑care HbA1c devices with laboratory HPLC methods: A systematic review and meta‑analysis (2026) · cites this paper as [30]
Some included studies evaluated POC devices using venous blood instead of capillary fingerstick samples, despite the devices being designed for the latter.
“From a practical standpoint, some studies evaluated POC devices using venous blood instead of capillary fingerstick samples, despite the devices being designed for the latter (30,33).”
From Comparison of point‑of‑care HbA1c devices with laboratory HPLC methods: A systematic review and meta‑analysis (2026) · cites this paper as [30]
The citation lists haven’t been fetched yet
References 27 · Cited by 13
- Authors
- Guadalupe Vargas, M., Pazmiño Gomez, B. J., Vera Lorenti, F. E., Álvarez Condo, G. M., Rodas Neira, E. I., Veron, D., Fernández Veron, M., Cercado, A. G., Bahar, B., Tufro, A., & Veron, D.
- Affiliations
- Universidad Estatal de Milagro (EC)
- Yale University (US)
- ORCID
- Bahar, B.
- Venue
- Endocrinología, Diabetes y Nutrición
- Publisher
- Elsevier BV
- Published
- 2020-05
- Volume
- 67
- Issue
- 5
- Pages
- 297-303
- ISSN
- 2530-0164 (ISSN-L)
- OpenAlex
- W2994856420
- Cited by
- 13
- References
- 27
- Citations by year
- 2026: 1 · 2025: 4 · 2024: 3 · 2023: 2 · 2021: 3
- FWCI
- 0.52
- Type
- journal article
- Language
- es
- Open access
- closed
- Locations
- Endocrinología Diabetes y Nutrición (publishedVersion)
- PubMed (publishedVersion)
- Funders
- Secretary of Education Superior, Science, Technology and Innovation of Ecuador
- State University of Milagro (UNEMI-OCAS-SO-27092013-N°26 DV2013–2014, UNEMI-OCAS-SO-03072014-N8-DV2014-2015, UNEMI-OCAS-SO-30052016-N5-DV2015-2017)
- Topics
- Health Sciences › Medicine › Physiology › Clinical Laboratory Practices and Quality Control
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Life Sciences › Biochemistry, Genetics and Molecular Biology › Clinical Biochemistry › Advanced Glycation End Products research
- Keywords
- HbA1c, point-of-care testing, HPLC comparison, assay precision and accuracy, diabetes monitoring, method comparison, resource-limited settings
- MeSH terms
- Adult, Aged, Chromatography, High Pressure Liquid, Ecuador, Female, Glycated Hemoglobin, Humans, Immunoassay, Male, Middle Aged
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review