Optical performance of 3 intraocular lens designs in the presence of decentration
Abstract
PURPOSE: To study the theoretical optical performance of 3 intraocular lens (IOL) designs in the presence of IOL decentration. SETTING: Optics Center, Bausch & Lomb, Rochester, New York, USA. METHODS: A ray-tracing program was used to evaluate the effect of IOL decentration on the optical performance of 3 silicone IOLs (LI61U, Bausch & Lomb; Tecnis Z9000, Advanced Medical Optics; and a new aberration-free IOL [SofPort AO, Bausch & Lomb]) in an experimental model eye. The study was done using pupil diameters of 3.0 mm, 4.0 mm, and 5.0 mm and IOL decentrations of 0 mm, 0.25 mm, 0.50 mm, 0.75 mm, and 1.00 mm. The modulation transfer functions were computed and plotted. A Monte Carlo simulation analysis with 1000 trials with IOL decentration randomly varying for each pupil size was performed. RESULTS: Decentration of LI61U and Tecnis Z9000 IOLs led to asymmetrical higher-order aberrations that adversely affected the optical performance of the model eye; performance was not affected with the aberration-free IOL because it lacks inherent spherical aberration. Optical performance with the aberration-free IOL was better than with the LI61U IOL as the former has less spherical aberration and did not introduce other aberrations when decentered. Performance with the aberration-free IOL was better than with the Tecnis Z9000 IOL for 3.0 mm, 4.0 mm, and 5.0 mm pupils when decentration exceeded 0.15 mm, 0.30 mm, and 0.38 mm, respectively. Performance with the LI61U IOL was better than with the Tecnis Z9000 IOL for 3.0 mm, 4.0 mm, and 5.0 mm pupils when decentration exceeded 0.3 mm, 0.5 mm, and 0.5 mm, respectively. Monte Carlo simulations showed the expected postoperative results of the LI61U IOL and aberration-free IOL would be repeatable and predictable, whereas the outcomes with the Tecnis Z9000 IOL would vary widely. CONCLUSIONS: The optical performance of the model eye was not affected by decentration of an aspheric IOL designed to have no inherent spherical aberration. With decentration, the performance with the new IOL was better than with a conventional spherical IOL and an aspheric IOL designed to offset the spherical aberration of an average cornea.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Cataract in the Adult Eye Preferred Practice Pattern®
Olson, R. J., Braga-Mele, R., Chen, S. H., Miller, K. M., Pineda, R., Tweeten, J. P., & Musch, D. C. (2017) · Ophthalmology · doi:10.1016/j.ophtha.2016.09.027 · Cited by 277
Comparison of visual performance with an aspheric intraocular lens and a spherical intraocular lens
Thiagarajan, M., McClenaghan, R., & Anderson, D. F. (2011) · Journal of Cataract and Refractive Surgery · doi:10.1016/j.jcrs.2011.05.046 · Cited by 19
Metadata
- Authors
- Altmann, G. E., Nichamin, L. D., Lane, S. S., & Pepose, J. S.
- Venue
- Journal of Cataract and Refractive Surgery
- Published
- 2005-03
- Volume
- 31
- Issue
- 3
- Pages
- 574-585
- ISSN
- 0146-2776 (ISSN-L)
- 0886-3350
- OpenAlex
- W2012502763
- Cited by
- 210
- References
- 0
- FWCI
- 12.80
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Optical performance of 3 intraocular lens designs in the presence of decentration. https://app.cochat.ai/papers/10.1016/j.jcrs.2004.09.024
- The paper
- Altmann, G. E., Nichamin, L. D., Lane, S. S., & Pepose, J. S. (2005). Optical performance of 3 intraocular lens designs in the presence of decentration. Journal of Cataract and Refractive Surgery. https://doi.org/10.1016/j.jcrs.2004.09.024
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.