Preconcentration and speciation of arsenic in water specimens by the combination of solidification of floating drop microextraction and electrothermal atomic absorption spectrometry
Abstract
A simple solidification of floating drop microextraction procedure for preconcentration and speciation of trace amounts of As(III) and As(V) in water samples has been proposed prior to electrothermal atomic absorption spectrometry (ETAAS). In this method, a free microdroplet of organic solvent is floated on the surface of aqueous solution while being agitated by a stirring bar placed on the bottom of the sample vial. The determination of As(III) was achieved by selective formation of the As(III)-pyrrolidine dithiocarbamate complex in the presence of 0.1M HCl while As(V) forms a weak complex with the ligand in the same pH conditions. Total inorganic As(III, V) was extracted similarly after reduction of As(V) to As(III) with potassium iodide and sodium thiosulfate and As(V) concentration was calculated by difference. Pd(NO(3))(2) was used as a chemical modifier in ETAAS. Some important extraction parameters such as the type of organic solvent, solvent volume, sample stirring rate, sample solution temperature, salt addition and the exposure time on the extraction recovery were investigated and optimized. Under the optimized extraction conditions, the detection limit of 9.2 pg mL(-1) and suitable precision (RSD<8.6%), along with enhancement factor of 1000 for As were achieved. The developed method was applied successfully to speciation of As(III), As(V) and determination of the total amount of As in water samples.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
With ammonium pyrrolidine dithiocarbamate (APDC) as chelating agent and Pd(NO3)2 as chemical modifier, the LPME-SFO-ETAAS method was successfully applied to the extraction and determination of arsenic in water specimens (tap water and well water).
“With ammonium pyrrolidine dithiocarbamate (APDC) as chelating agent and Pd(NO3)2 as chemical modifier, the method was successfully applied to the extraction and determination of arsenic in water specimens (tap water and well water) .”
From Developments in Liquid-phase Microextraction Method Based on Solidification of Floating Organic Drop (2010) · cites this paper as [33]
For the LPME-SFO-ETAAS determination of arsenic in water specimens, the RSD was below 8.6%.
“the RSD was below 8.6%.”
From Developments in Liquid-phase Microextraction Method Based on Solidification of Floating Organic Drop (2010) · cites this paper as [33]
For the LPME-SFO-ETAAS determination of arsenic in water specimens, the LOD was 9.2 ng L-1.
“The LOD was 9.2 ng L -1”
From Developments in Liquid-phase Microextraction Method Based on Solidification of Floating Organic Drop (2010) · cites this paper as [33]
Ghambarian et al. [33] developed a highly efficient LPME-SFO method for the determination of arsenic by electrothermal atomic absorption spectrometry (ETAAS).
“Ghambarian et al [33] developed a highly efficient LPME-SFO method for the determination of arsenic by electrothermal atomic absorption spectrometry (ETAAS).”
From Developments in Liquid-phase Microextraction Method Based on Solidification of Floating Organic Drop (2010) · cites this paper as [33]
Citations
Cited by
Sorted by citations
Green aspects, developments and perspectives of liquid phase microextraction techniques
Spietelun, A., Marcinkowski, Ł., de la Guardia, M., & Namieśnik, J. (2014) · Talanta · doi:10.1016/j.talanta.2013.10.050 · Cited by 333
Trends in liquid-phase microextraction, and its application to environmental and biological samples
Han, D., & Row, K. H. (2012) · Microchimica Acta · doi:10.1007/s00604-011-0678-0 · Cited by 175
Developments in Liquid-phase Microextraction Method Based on Solidification of Floating Organic Drop
WANG, Y.-Y., ZHAO, G.-Y., CHANG, Q.-Y., ZANG, X.-H., WANG, C., & WANG, Z. (2010) · Chinese Journal of Analytical Chemistry · doi:10.1016/s1872-2040(09)60073-6 · Cited by 31
Metadata
- Authors
- Ghambarian, M., Khalili-Zanjani, M. R., Yamini, Y., Esrafili, A., & Yazdanfar, N.
- Affiliations
- Tarbiat Modares University (IR)
- Academic Center for Education, Culture and Research (IR)
- Corresponding author
- Yamini, Yadollah
- Venue
- Talanta
- Publisher
- Elsevier BV
- Published
- 2010-04-15
- Volume
- 81
- Issue
- 1-2
- Pages
- 197-201
- ISSN
- 0039-9140 (ISSN-L)
- OpenAlex
- W2067083700
- Cited by
- 67
- References
- 0
- Citations by year
- 2024: 5 · 2023: 1 · 2022: 2 · 2021: 2 · 2020: 3 · 2019: 2 · 2018: 3 · 2017: 5 · 2016: 6 · 2015: 6 · 2014: 7 · 2013: 7 · 2012: 4
- FWCI
- 5.76
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Talanta (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Physical Sciences › Chemistry › Analytical Chemistry › Analytical chemistry methods development
- Physical Sciences › Environmental Science › Environmental Chemistry › Arsenic contamination and mitigation
- Physical Sciences › Chemistry › Electrochemistry › Electrochemical Analysis and Applications
- Keywords
- arsenic speciation, arsenite, arsenate, preconcentration, water analysis, pyrrolidine dithiocarbamate, graphite furnace atomic absorption spectrometry, palladium chemical modifier, trace metal analysis
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review