The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.
Long-term monitoring and data analysis of the Tamar Bridge
Open accessAbstract
Abstract A sound understanding of a structure's normal condition, including its response to normal environmental and operational variations is desirable for structural health monitoring and necessary for performance monitoring of civil structures. The current paper outlines the extensive monitoring campaign of the Tamar Suspension Bridge as well as analysis carried out in an attempt to understand the bridge's normal condition. Specifically the effects of temperature, traffic loading and wind speed on the structure's dynamic response are investigated. Finally, initial steps towards the development of a structural health monitoring system for the Tamar Bridge are addressed.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Traffic-induced variability in dynamic properties of cable-stayed bridge
Zhang, Q. W., Fan, L. C., & Yuan, W. C. (2002) · Earthquake Engineering & Structural Dynamics · doi:10.1002/eqe.204 · Cited by 73
Cited by
Hassani, S., & Dackermann, U. (2023) · Sensors · doi:10.3390/s23042204 · Cited by 492
Recent progress and future trends on damage identification methods for bridge structures
An, Y., Chatzi, E., Sim, S., Laflamme, S., Blachowski, B., & Ou, J. (2019) · Structural Control and Health Monitoring · doi:10.1002/stc.2416 · Cited by 338
Vibration-based monitoring of civil infrastructure: challenges and successes
Brownjohn, J. M. W., De Stefano, A., Xu, Y.-L., Wenzel, H., & Aktan, A. E. (2011) · Journal of Civil Structural Health Monitoring · doi:10.1007/s13349-011-0009-5 · Cited by 336
Self-citationValidation of vibration-based structural health monitoring on retrofitted railway bridge KW51
Maes, K., Van Meerbeeck, L., Reynders, E. P. B., & Lombaert, G. (2022) · Mechanical Systems and Signal Processing · doi:10.1016/j.ymssp.2021.108380 · Cited by 155
Worden, K., & Cross, E. J. (2018) · Mechanical Systems and Signal Processing · doi:10.1016/j.ymssp.2017.04.022 · Cited by 135
“In this case, it was clear from the coefficient estimates in the original study [4] that the dominant effect on the natural frequency was from traffic loading; Fig.”
BackgroundMethodResultInfluentialSelf-citation
Metadata
- Authors
- Cross, E. J., Koo, K. Y., Brownjohn, J. M. W., & Worden, K.
- Venue
- Mechanical Systems and Signal Processing
- Publisher
- Elsevier BV
- Published
- 2013-02
- Volume
- 35
- Issue
- 1-2
- Pages
- 16-34
- ISSN
- 0888-3270 (ISSN-L)
- OpenAlex
- W2094460095
- Cited by
- 392
- References
- 1
- FWCI
- 32.13
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Long-term monitoring and data analysis of the Tamar Bridge. https://app.cochat.ai/papers/10.1016/j.ymssp.2012.08.026
- The paper
- Cross, E. J., Koo, K. Y., Brownjohn, J. M. W., & Worden, K. (2013). Long-term monitoring and data analysis of the Tamar Bridge. Mechanical Systems and Signal Processing. https://doi.org/10.1016/j.ymssp.2012.08.026
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.