A transcatheter intracardiac shunt device for heart failure with preserved ejection fraction (REDUCE LAP-HF): a multicentre, open-label, single-arm, phase 1 trial
Open accessAbstract
BACKGROUND Heart failure with preserved ejection fraction (HFPEF) is a common, globally recognised, form of heart failure for which no treatment has yet been shown to improve symptoms or prognosis. The pathophysiology of HFPEF is complex but characterised by increased left atrial pressure, especially during exertion, which might be a key therapeutic target. The rationale for the present study was that a mechanical approach to reducing left atrial pressure might be effective in HFPEF. METHODS The REDUCe Elevated Left Atrial Pressure in Patients with Heart Failure (REDUCE LAP-HF) study was an open-label, single-arm, phase 1 study designed to assess the performance and safety of a transcatheter interatrial shunt device (IASD, Corvia Medical, Tewkesbury, MA, USA) in patients older than 40 years of age with symptoms of HFPEF despite pharmacological therapy, left ventricular ejection fraction higher than 40%, and a raised pulmonary capillary wedge pressure at rest (>15 mm Hg) or during exercise (>25 mm Hg). The study was done at 21 centres (all departments of cardiology in the UK, Netherlands, Belgium, France, Germany, Austria, Denmark, Australia, and New Zealand). The co-primary endpoints were the safety and performance of the IASD at 6 months, together with measures of clinical efficacy, including functional capacity and clinical status, analysed per protocol. This study is registered with ClinicalTrials.gov, number NCT01913613. FINDINGS Between Feb 8, 2014, and June 10, 2015, 68 eligible patients were entered into the study. IASD placement was successful in 64 patients and seemed to be safe and well tolerated; no patient had a peri-procedural or major adverse cardiac or cerebrovascular event or need for cardiac surgical intervention for device-related complications during 6 months of follow-up. At 6 months, 31 (52%) of 60 patients had a reduction in pulmonary capillary wedge pressure at rest, 34 (58%) of 59 had a lower pulmonary capillary wedge pressure during exertion, and 23 (39%) of 59 fulfilled both these criteria. Mean exercise pulmonary capillary wedge pressure was lower at 6 months than at baseline, both at 20 watts workload (mean 32 mm Hg [SD 8] at baseline vs 29 mm Hg [9] at 6 months, p=0·0124) and at peak exercise (34 mm Hg [8] vs 32 [8], p=0·0255), despite increased mean exercise duration (baseline vs 6 months: 7·3 min [SD 3·1] vs 8·2 min [3·4], p=0·03). Sustained device patency at 6 months was confirmed by left-to-right shunting (pulmonary/systemic flow ratio: 1·06 [SD 0·32] at baseline vs 1·27 [0·20] at 6 months, p=0·0004). INTERPRETATION Implantation of an interatrial shunt device is feasible, seems to be safe, reduces left atrial pressure during exercise, and could be a new strategy for the management of HFPEF. The effectiveness of IASD compared with existing treatment for patients with HFPEF requires validation in a randomised controlled trial. FUNDING Corvia Medical Inc.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Trends in Prevalence and Outcome of Heart Failure with Preserved Ejection Fraction
Owan, T. E., Hodge, D. O., Herges, R. M., Jacobsen, S. J., Roger, V. L., & Redfield, M. M. (2006) · New England Journal of Medicine · doi:10.1056/nejmoa052256 · Cited by 4,191
Yusuf, S., Pfeffer, M. A., Swedberg, K., Granger, C. B., Held, P., McMurray, J. J., Michelson, E. L., Olofsson, B., & Östergren, J. (2003) · The Lancet · doi:10.1016/s0140-6736(03)14285-7 · Cited by 2,994
Spironolactone for heart failure with preserved ejection fraction
Pitt, B., Pfeffer, M. A., Assmann, S. F., Boineau, R., Anand, I. S., Claggett, B., Clausell, N., Desai, A. S., Diaz, R., Fleg, J. L., et al. (2014) · New England Journal of Medicine · doi:10.1056/nejmoa1313731 · Cited by 2,809
Paulus, W. J., Tschöpe, C., Sanderson, J. E., Rusconi, C., Flachskampf, F. A., Rademakers, F. E., Marino, P., Smiseth, O. A., De Keulenaer, G., Leite-Moreira, A. F., et al. (2007) · European Heart Journal · doi:10.1093/eurheartj/ehm037 · Cited by 2,679
Irbesartan in patients with heart failure and preserved ejection fraction
Massie, B. M., Carson, P. E., McMurray, J. J., Komajda, M., McKelvie, R., Zile, M. R., Anderson, S., Donovan, M., Iverson, E., Staiger, C., et al. (2008) · New England Journal of Medicine · doi:10.1056/nejmoa0805450 · Cited by 1,960
The perindopril in elderly people with chronic heart failure (PEP-CHF) study
Cleland, J. G. F. (2006) · European Heart Journal · doi:10.1093/eurheartj/ehl250 · Cited by 1,508
Self-citationCongestive heart failure in subjects with normal versus reduced left ventricular ejection fraction
Vasan, R. S., Larson, S., Benjamin, E. J., Evans, J. C., Reiss, C. K., & Levy, D. (1999) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(99)00118-7 · Cited by 1,331
Redfield, M. M., Chen, H. H., Borlaug, B. A., Semigran, M. J., Lee, K. L., Lewis, G., LeWinter, M. M., Rouleau, J. L., Bull, D. A., Mann, D. L., et al. (2013) · JAMA · doi:10.1001/jama.2013.2024 · Cited by 1,155
Exercise Hemodynamics Enhance Diagnosis of Early Heart Failure With Preserved Ejection Fraction
Borlaug, B. A., Nishimura, R. A., Sorajja, P., Lam, C. S. P., & Redfield, M. M. (2010) · Circulation: Heart Failure · doi:10.1161/circheartfailure.109.930701 · Cited by 1,118
Yancy, C. W., Lopatin, M., Stevenson, L. W., De Marco, T., & Fonarow, G. C. (2006) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2005.09.022 · Cited by 927
Cited by
Metra, M., & Teerlink, J. R. (2017) · The Lancet · doi:10.1016/s0140-6736(17)31071-1 · Cited by 673
Evaluation and management of heart failure with preserved ejection fraction
Borlaug, B. A. (2020) · Nature Reviews Cardiology · doi:10.1038/s41569-020-0363-2 · Cited by 675
Gorter, T. M., van Veldhuisen, D. J., Bauersachs, J., Borlaug, B. A., Celutkiene, J., Coats, A. J. S., Crespo-Leiro, M. G., Guazzi, M., Harjola, V.-P., Heymans, S., et al. (2018) · European Journal of Heart Failure · doi:10.1002/ejhf.1029 · Cited by 389
Heart failure with preserved ejection fraction: from mechanisms to therapies
Lam, C. S. P., Voors, A. A., de Boer, R. A., Solomon, S. D., & van Veldhuisen, D. J. (2018) · European Heart Journal · doi:10.1093/eurheartj/ehy301 · Cited by 355
Feldman, T., Mauri, L., Kahwash, R., Litwin, S., Ricciardi, M. J., van der Harst, P., Penicka, M., Fail, P. S., Kaye, D. M., Petrie, M. C., et al. (2018) · Circulation · doi:10.1161/circulationaha.117.032094 · Cited by 269
Self-citation
Metadata
- Authors
- Hasenfuß, G., Hayward, C., Burkhoff, D., Silvestry, F. E., McKenzie, S., Gustafsson, F., Malek, F., Van der Heyden, J., Lang, I., Petrie, M. C., Cleland, J. G. F., Leon, M., & Kaye, D. M.
- Venue
- The Lancet
- Publisher
- Elsevier BV
- Published
- 2016-03
- Volume
- 387
- Issue
- 10025
- Pages
- 1298-1304
- ISSN
- 0099-5355 (ISSN-L)
- 0140-6736
- OpenAlex
- W2311933573
- Cited by
- 294
- References
- 27
- FWCI
- 28.14
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: A transcatheter intracardiac shunt device for heart failure with preserved ejection fraction (REDUCE LAP-HF): a multicentre, open-label, single-arm, phase 1 trial. https://app.cochat.ai/papers/10.1016/s0140-6736(16)00704-2
- The paper
- Hasenfuß, G., Hayward, C., Burkhoff, D., Silvestry, F. E., McKenzie, S., Gustafsson, F., Malek, F., Van der Heyden, J., Lang, I., Petrie, M. C., Cleland, J. G. F., Leon, M., & Kaye, D. M. (2016). A transcatheter intracardiac shunt device for heart failure with preserved ejection fraction (REDUCE LAP-HF): a multicentre, open-label, single-arm, phase 1 trial. The Lancet. https://doi.org/10.1016/s0140-6736(16)00704-2
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.