STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
The DOI matches its Crossref record; 7 of the 7 references it cites were found in a registry or an index.
Abstract
In clinical measurement comparison of a new measurement technique with an established one is often needed to see whether they agree sufficiently for the new to replace the old. Such investigations are often analysed inappropriately, notably by using correlation coefficients. The use of correlation is misleading. An alternative approach, based on graphical techniques and simple calculations, is described, together with the relation between this analysis and the assessment of repeatability.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“and all parameters used for the proposition and cross-validation of the model confirmed their validity for the studied population (15, 37–39, 57)”
From Development and Cross-Validation of a Predictive Equation for Fat-Free Mass in Brazilian Adolescents by Bioelectrical Impedance (2022) · cites this paper as [39]
In the citing paper
“Each volume measurement was performed twice to determine reproducibility by the method of Bland and Altman (1986).”
From Archimedes revisited: a faster, better, cheaper method of accurately measuring the volume of small objects (2005) · cites this paper as Bland and Altman (1986)
In the citing paper
“After a fairly extensive search of physics textbooks and the scientific literature over a number of years, no direct references to the technique have been discovered 1 .”
From Archimedes revisited: a faster, better, cheaper method of accurately measuring the volume of small objects (2005) · cites this paper as [1]
In the citing paper
“However, paired t test and Bland-Altman plot are methods for analyzing agreement, and Pearson correlation coefficient is only a measure of correlation, and hence, they are nonideal measures of reliability.”
From A guideline of selecting and reporting intraclass correlation coefficients for reliability research (2016) · cites this paper as [6]
In the citing paper
“However, paired t test and Bland-Altman plot are methods for analyzing agreement, and Pearson correlation coefficient is only a measure of correlation, and hence, they are nonideal measures of reliability.”
From A guideline of selecting and reporting intraclass correlation coefficients for reliability research (2016) · cites this paper as [6]
2 more claims about this paper
Create a free account to see every claim other papers make about this one.
Citations
Cites
Measurement in Medicine: The Analysis of Method Comparison Studies
Altman, D. G., & Bland, J. M. (1983) · The Statistician · doi:10.2307/2987937 · Cited by 4,300
Self-citationStatistical Methods in Medical Research
(1972) · Postgraduate Medical Journal · doi:10.1136/pgmj.48.555.67-a · Cited by 526
RELATION BETWEEN INITIAL BLOOD PRESSURE AND ITS FALL WITH TREATMENT
Gill, J. S., Beevers, D. G., Zezulka, A. V., & Davies, P. (1985) · The Lancet · doi:10.1016/s0140-6736(85)91219-x · Cited by 156
The Nellcor N‐101 pulse oximeter: A clinical evaluation in anaesthesia and intensive care
TYTLER, J. A., & SEELEY, H. F. (1986) · Anaesthesia · doi:10.1111/j.1365-2044.1986.tb12793.x · Cited by 32
Comparison of the new miniature Wright peak flow meter with the standard Wright peak flow meter.
Oldham, H. G., Bevan, M. M., & McDermott, M. (1979) · Thorax · doi:10.1136/thx.34.6.807 · Cited by 28
Estimation of gestational age at birth. Comparison of two methods.
Serfontein, G. L., & Jaroszewicz, A. M. (1978) · Archives of Disease in Childhood · doi:10.1136/adc.53.6.509 · Cited by 23
D'Arbela, P. G., Silayan, Z. M., & Bland, M. (1986) · Heart · doi:10.1136/hrt.56.5.445 · Cited by 4
Cited by
Items 47,001–47,100 of 47,868
Bui, H.-A., Bui, M.-A., Huynh, T.-T.-T., Le, H.-L.-A., Phan, P.-D., Le, N.-T.-M., Tran, T.-L., Tran, T.-B.-V., Lam, Q.-V., Pham, V.-K., et al. (2026) · Frontiers in Oral Health · doi:10.3389/froh.2026.1885521
Cicchetti, R., Pisa, S., Piuzzi, E., Pistillucci, M., Forte, A., & Testa, O. (2026) · Sensors · doi:10.3390/s26154663
Jagim, A. R., Dobbs, W. C., Magee, M. K., Fields, J. B., Sims, J. N. L., Saine, N. D., & Jones, M. T. (2026) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0000000000004071
Alvarez, A. A., Martínez, M., Redondo, J. I., Milner, P. I., & Bardell, D. (2026) · Veterinary Anaesthesia and Analgesia · doi:10.1016/j.vaa.2026.101287
Silva‐Batista, C., Scanlan, K. T., Stojak, M., Wilhelm, J. L., Carlson‐Kuhta, P., Liu, W., Chen, S., Horak, F. B., Mancini, M., & King, L. A. (2026) · Movement Disorders Clinical Practice · doi:10.1002/mdc3.70752
Rokai, J., Rácz, M., Becske, M., Csipor, J., Köllőd, C. M., Ulbert, I., & Márton, G. (2026) · Scientific Reports · doi:10.1038/s41598-026-52672-8
Physiology of the breathing rest during speech – A cross-sectional study in vocally healthy adults
Müller, C., & Lukaschyk, J. (2026) · IJVS International Journal of Voice Sciences · doi:10.2478/ijvs-2025-0005
Porteous, A. J., Beaumont, O., Blucher, N., Platt, N., & Fletcher, J. W. A. (2026) · The Knee · doi:10.1016/j.knee.2026.104569
Amjadian, M., Gok, G., Choi, N., Khan, S., Pandya, M., Jokerst, J. V., & Chen, C. (2026) · Dentomaxillofacial Radiology · doi:10.1093/dmfr/twag050 · Cited by 1
Accuracy of AI-Measured Overjet and Overbite of Cephalometry Compared to Manual Measurement
Ganjali, B., Radwan, M. T., & Altuğ, Z. (2026) · Applied Sciences · doi:10.3390/app16157404
Yalçın, M., Özden, F., Başer, M., Uysal, İ., Özkeskin, M., Baykal, D., & Sarı, Z. (2026) · Physiotherapy Theory and Practice · doi:10.1080/09593985.2026.2707231
Höchsmann, C., Diktas, H. E., Myers, C. A., Dorling, J. L., Rood, J. C., Greenway, F., Champagne, C. M., Apolzan, J. W., & Martin, C. K. (2026) · American Journal of Clinical Nutrition · doi:10.1016/j.ajcnut.2026.101452
Sayem, S. A. J., Lee, S., Gui, L., Ali, M. S., & Park, S. (2026) · Veterinary Quarterly · doi:10.1080/01652176.2026.2708928
Wu, T., Yu, H., Li, S., & Jiang, H. (2026) · Ear and Hearing · doi:10.1097/aud.0000000000001868
Adnyana, K., & Schwung, A. (2026) · Machine Learning with Applications · doi:10.1016/j.mlwa.2026.100960
Roth, R., Reinke, Z., Siedlecki, M., Luitjens, J. H., Hügle, M., Enke, J. S., Liska, C., Feitelson, L.-M., Lapa, C., Kröncke, T. J., et al. (2026) · Diagnostics · doi:10.3390/diagnostics16152329
Kim, S. J., Yang, S., Song, M., Kim, C. U., Park, Y. H., Jung, W. K., Lee, J. M., & Chung, H. C. (2026) · Frontiers in Public Health · doi:10.3389/fpubh.2026.1850944
Comparison of AIZE-EX and SITA-standard visual field measurements in patients with glaucoma
Kano, K., Nomoto, H., Matsumoto, C., Kusaka, S., & Kuwayama, Y. (2026) · Japanese Journal of Ophthalmology · doi:10.1007/s10384-026-01395-4
Fertitta, L., Coulpier, F., Oubrou, L., Decrouy, X., Audureau, É., Ortonne, N., Wolkenstein, P., & Topilko, P. (2026) · PLoS ONE · doi:10.1371/journal.pone.0354818
Torresani, M., Moudrý, V., Hakkenberg, C. R., Barták, V., Rocchini, D., Puliti, S., Hawryło, P., Stereńczak, K., Cotrina-Sánchez, A., Satta, G. G. A., et al. (2026) · Ecological Informatics · doi:10.1016/j.ecoinf.2026.103950 · Cited by 1
Collins, K. C., Clark, A., Pomeroy, V. M., & Kennedy, N. C. (2026) · Brain Topography · doi:10.1007/s10548-026-01235-1
Sun, M., Zhang, D., Zhang, W., Wu, D., Li, C., & Qiao, Y. (2026) · FirePhysChem · doi:10.1016/j.fpc.2026.07.007
Validation of the Heat3D System for Mapping U-Values in Homes Through a Two-Phase Winter Field Trial
Henshaw, G., Fitton, R., Jack, R., Bennett, S., Swan, W., Farmer, D., & Paraskevas, I. (2026) · Buildings · doi:10.3390/buildings16152968
Bermingham, K. M., Linenberg, I., Francis, L., Leeming, E. R., Wallace, S. M., Badri, F., Creedon, A. C., Hornzee, N., Bernard, H. M., Smith, H. A., et al. (2026) · Journal of Nutritional Science · doi:10.1017/jns.2026.10080 · Cited by 1
Mori, M., Masuda, T., Sugimoto, K., Matsushige, T., Takeuchi, M., Fujimoto, T., Araki, J., & Ishikawa, M. (2026) · Radiological Physics and Technology · doi:10.1007/s12194-026-01113-2
Pyo, S., 김보민, Min, J., Park, B. C., & Choi, S. (2026) · International Journal of Cosmetic Science · doi:10.1111/ics.70132
Shobi, P. M. T. (2026) · Research Square · doi:10.21203/rs.3.rs-10475315/v1
Noel-Barker, N., Hillman, C., Pollon, E., Connors, E., Christie, C., Pettitt, W., Gallagher, J., Higgins, K., Riley, M., Bestwick-Stevenson, T., et al. (2026) · Foot & Ankle Orthopaedics · doi:10.1177/24730114261460463
de León, S. C., González, M. S., Carrera, I. D. C., & Pascoal, A. G. (2026) · BMC Women s Health · doi:10.1186/s12905-026-04740-4
Real-time Cuffless Blood Pressure Estimation Using Single-lead ECG and Gaussian Process Regression
Widasari, E. R., Alhamid, F. K., Laksono, R. M., & Syauqy, D. (2026) · International journal of intelligent engineering and systems · doi:10.22266/ijies2026.0831.42
Tan, T. C., Zheng, J., Chen, T., Liu, Y., Luo, Y., Yang, Y., & Huang, T. (2026) · PeerJ · doi:10.7717/peerj.21490
Applicability of resting energy expenditure prediction equations in Chinese nursing home residents
Zhen, X., Li, T., Du, S., Mengjia, Y., Wang, L., Sun, J., Yanjuan, Z., & Chen, X. (2026) · Frontiers in Nutrition · doi:10.3389/fnut.2026.1860294
Wells, S. A., Lane, G. I., Ozkan, M., Sui, W., Cameron, A. P., & Hamilton, J. (2026) · NMR in Biomedicine · doi:10.1002/nbm.70366
Lewis, M. P., Moir, C., Scott, K. R., Mariani, S., Hänfling, B., & Dawnay, N. (2026) · Forensic Science International Genetics · doi:10.1016/j.fsigen.2026.103598
How Many Stimuli are Enough? Power and Reliability of Crowdsourced Speech Codec Evaluation
Kumar, J., & Srinivasan, S. (2026) · International Conference on Quality of Multimedia Experience (QoMEX) · doi:10.1109/qomex69967.2026.11618377
Simeoli, R., Mancini, A., Cairoli, S., Rossi, C., Calabrese, C., Trivisano, M., Salimbene, L., Dionisi‐Vici, C., Pietrafusa, N., Specchio, N., et al. (2026) · Pharmaceuticals · doi:10.3390/ph19081188
Huang, S., Tseng, H.-P., Chen, Y.-H., Tan, Y.-L., Lee, J.-P., Wu, C., & Hsu, C. H. (2026) · Journal of Imaging Informatics in Medicine · doi:10.1007/s10278-026-02167-9
Quantifying the Error in Body Composition Assessment: The Influence of Criterion Method SelectionPreprint
Fedewa, M. V., Sullivan, K. W., Metoyer, C. J., & Esco, M. R. (2026) · Research Square · doi:10.21203/rs.3.rs-9636135/v1
Lee, R. H. Y., Vukovic, S., Lucci, V.-E. M., Lando, J. T., & Claydon, V. E. (2026) · Autonomic Neuroscience · doi:10.1016/j.autneu.2026.103464
Saltarelli, G., Cerbo, G. D., Innocenzi, A., Chellis, M. D., Bruno, F., Splendiani, A., & Cesare, E. D. (2026) · Diagnostics · doi:10.3390/diagnostics16152383 · Cited by 1
Jandewerth, T., Kaiser, K. P., Golschan, E., Dávidová, P., & Kohnen, T. (2026) · Research Square · doi:10.21203/rs.3.rs-10467160/v1
Jiang, L., Shu, H., Wiloth, S., Hüsler, S.-S., Zhang, P., Shen, C., Wang, R., & Kasahara, S. (2026) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-026-02109-0
Uellendahl, M. M. (2026) · ARQUIVOS BRASILEIROS DE CARDIOLOGIA - IMAGEM CARDIOVASCULAR · doi:10.36660/abcimg.20260087i
Uellendahl, M. M. (2026) · ARQUIVOS BRASILEIROS DE CARDIOLOGIA - IMAGEM CARDIOVASCULAR · doi:10.36660/abcimg.20260087
Yuan, R., & Lu, X. (2026) · Journal of Ultrasound in Medicine · doi:10.1002/jum.70392
Vlad-Alexandru, A., Munteanu, M., & Lodin, A. (2026) · doi:10.3390/engproc2026148040
Agreement Coefficients for Continuous Variables: A Review
Vallejos, R. O. (2026) · Environmetrics · doi:10.1002/env.70117
Creation of na image capture device for collective monitoring of water quality
Camacho, F. G. F., Kapp, M. N., Boroski, M., & Steiger, N. D. L. (2026) · Ambiente Construído · doi:10.1590/s1678-86212026000100999
Dominguez, R. E., Yang, V., Fu, A., Cheng, E. H., Terrera, M., Hota, P., Pradhan, A., Miranda, S., Snozek, C., Thomas, L. F., et al. (2026) · Analytica—A Journal of Analytical Chemistry and Chemical Analysis · doi:10.3390/analytica7030051
Böker, K., Vogel, M., Grundmann, A., & Kieß, W. (2026) · European Journal of Clinical Nutrition · doi:10.1038/s41430-026-01788-1
Daikuya, S., & Okayama, Y. (2026) · International Journal of Sports Physical Therapy · doi:10.26603/001c.165045
Advancing boundary-layer physics via Doppler wind LiDAR dynamic-thermodynamic profiling
Xian, J., Yang, H., Qiu, Z., Tian, Z., Li, L., Guo, J., Zhang, N., Wang, S., Chan, P. W., Xin, J., et al. (2026) · npj Climate and Atmospheric Science · doi:10.1038/s41612-026-01494-6
Liu, D., Liu, J., Yang, K., Hu, D., Ma, J., Ran, H., & Song, Y. (2026) · Frontiers in Psychology · doi:10.3389/fpsyg.2026.1863652
Evaluating Reliability and Bias of Large Language Models in Automated Essay Scoring
Lee, Y. (2026) · Computers in the Schools · doi:10.1080/07380569.2026.2712300
Validity and reliability of England Hockey’s 16-pitch length aerobic capacity test
Spooner, T. W., Wallis, P. S., & West, A. T. (2026) · The Journal of Sport and Exercise Science · doi:10.36905/jses.2026.01.05
Dulanská, S., Silliková, V., Gebeová, K., Goneková, Z., Trnka, M., & Pánik, J. (2026) · Nuclear Engineering and Technology · doi:10.1016/j.net.2026.104633
Kabba, D., Goyal, K., & Kalra, R. (2026) · Physiotherapy Research International · doi:10.1002/pri.70265
Li, S., Shrestha, L., Tuychiev, M., Jung, T. Y., Choi, R. W., & Miyamoto, K. (2026) · Journal of Cosmetics Dermatological Sciences and Applications · doi:10.4236/jcdsa.2026.163012
Assessing Sprint Mechanical Outputs Derived from LPS, Radar and Laser Technologies in Basketball
Irid, Y., Fenaux, E., Legoupil, R., Tachdjian, M., Leduc, C., Toussaint, J., & Sedeaud, A. (2026) · Sensors · doi:10.3390/s26154845
Long, E., Dunn, R., & Androshchuk, V. (2026) · BMC Cardiovascular Disorders · doi:10.1186/s12872-026-06358-8
Dickinson, B., Jones, P. A., & Comfort, P. (2026) · International Journal of Sports Science & Coaching · doi:10.1177/17479541261469980
The Effect of Using Write‐Zoom for Volume Flow Measurements in Haemodialysis Fistulas
Kong, N., Curry, G., Thwaites, S. E., Pereira, I., & Uribe, S. A. (2026) · Sonography · doi:10.1002/sono.70095
Tan, K. Z., Pai, S. G. S., Kim, Y., Frautschi, A., Gwerder, M., Tan, K. Y., Koh, V. J. W., Ravi, D. K., Taylor, W. R., Malhotra, R., et al. (2026) · medRxiv · doi:10.64898/2026.08.03.26359452
Paperlein, F., Blaurock, M., Fenske, B., Möller, H., Wagner, V., Schneider, L., Ivanovska, T., Busch, C., Scharf, C., & Beule, A. G. (2026) · Journal of Imaging · doi:10.3390/jimaging12110535
Repeatability of Foveal Sensitivity Measured Using Adaptive Optics Microperimetry
Govindasamy, N., Tuten, W. S., Brainard, D. H., & Morgan, J. I. W. (2026) · Translational Vision Science & Technology · doi:10.1167/tvst.15.8.3
Telles-Araújo, G. D. T., Neto, R. S. D. O., Soares, M. Q. S., Honório, H. M., Sarmento, V. A., Duarte, M. A. H., & Rubira‐Bullen, I. R. F. (2026) · Biomaterial Investigations in Dentistry · doi:10.2340/biid.v13.46618
Berrow, E. J. (2012) · Aston Publications Explorer (Aston University) · doi:10.48780/publications.aston.ac.uk.00016328 · Cited by 2
Measuring the influence of stress on eating: Taking stock of the Salzburg Stress Eating Scale
Reichenberger, J., Fröse, N. V., & Meule, A. (2026) · Appetite · doi:10.1016/j.appet.2026.108751
Li, Y., 李海鹏, Song, Y., Ren, Z., & Zhang, B. (2026) · BMC Sports Science Medicine and Rehabilitation · doi:10.1186/s13102-026-01931-2
Moon, S. I., & Kwon, N. K. (2026) · Electronics · doi:10.3390/electronics15153477
A exploratory pilot method comparison study of two viscoelastic coagulation analyzers
Argumedo, D. R., Maranho, J. G., & Ortiz-Posadas, M. R. (2026) · Frontiers in Medicine · doi:10.3389/fmed.2026.1874412
Lee, D.-G., & Park, J. (2026) · Aquaculture International · doi:10.1007/s10499-026-02633-9
Development and Bench-to-Human Feasibility Evaluation of a Wireless Pinch-and-Grip Dynamometer
Pourhossein, A., Ouerghi, M. M., Kuborn, E., Tasdemir, D., Warwas, V., & Hughes, C. M. L. (2026) · IEEE Journal of Translational Engineering in Health and Medicine · doi:10.1109/jtehm.2026.3720716 · Cited by 1
Tume, S. C., O'Byrne, M. L., Fudge, C., Anders, M. M., & Gupta, D. (2026) · Pediatric Critical Care Medicine · doi:10.1097/pcc.0000000000003986
Mata, F., Cerqueira, J. O. L., Vaz, P. S., Paixão, G., & Araújo, J. P. (2026) · Ruminants · doi:10.3390/ruminants6030065
Chu, Y.-L., Ko, Y., Shih, H.-C., Muthusamy, A., Shiao, M., & Chiang, Y. (2026) · Biology Direct · doi:10.1186/s13062-026-00935-9
How useful is the VisTech MCTS000 for contrast sensitivity and glare testing?
Elliott, D. B. (2026) · Canadian journal of optometry/CJO. Canadian journal of optometry · doi:10.65636/cjo.v54i4.7214
Özcan, D., Nielsen, P. S., Alsner, J., & Tramm, T. (2026) · Biomarkers in Medicine · doi:10.1080/17520363.2026.2714196
Burch, K., Hamkins, J., McDaniel, L. T., Costa, A. R. C. E., Yang, Z., Stenum, J., Pagliocchini, A., Szczesny, C., Langdon, J. M., Chellappa, R. R., et al. (2026) · medRxiv · doi:10.64898/2026.08.04.26359742
Mazurek, M., Domagała, Z., & Matkowski, R. (2026) · Journal of Clinical Medicine · doi:10.3390/jcm15166154
Avaliação de dois relógios inteligentes de baixo custo para a aferição da pressão arterial
Ferreira, R. M., Stochero, L., Henning, E., Daumas, R. P., Matos, J., & Aranha, R. N. (2026) · Revista da Faculdade de Ciências Médicas de Sorocaba · doi:10.23925/1984-4840.2026v28a18
Tarhan, M., Meller, D., & Hammer, M. (2026) · BMC Ophthalmology · doi:10.1186/s12886-026-05216-7
Limited Interchangeability of Shunt‐Side SpO 2 Measurements During Hemodialysis
Kojima, K., Hirate, Y., Wada, Y., Nakajima, T., & Nakai, K. (2026) · Therapeutic Apheresis and Dialysis · doi:10.1002/1744-9987.70207
Obregón-Rosas, S., Loughlin, M. J. M., Carlos-Solís, P. F., Vera-Ávila, N., Rojas-Bergés, I., & Navarrete-Punaro, J. P. (2026) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2026.154424
Xu, Y., Xie, S., Wei, C., Yin, C., 雷晓东, Liu, Y., Mai, G., & Liu, H. (2026) · Research Square · doi:10.21203/rs.3.rs-10405040/v1
Yanzón, E. V. A., Ferre-Martínez, J., López-Rodríguez, D., Cervigón, R., Camacho-García, A., & Llopis, R. L. (2026) · Sensors · doi:10.3390/s26165036
Ho, Q. T., Bui, N. H., Tran, T. D., Khuat, Q. H., Tran, N. T., Le, X. C., Nguyen, H. Q., Nguyen, T. T., Nguyen, V. T., Mai, D. T., et al. (2026) · Journal of Imaging · doi:10.3390/jimaging12080364
Detector-Guided Multi-View Visual Monitoring of Bolt Loosening in Hydropower Generator Rotors
Lyu, J., Guo, J., Yuan, F., Ran, Y., Gong, H., Wu, T., & Zhang, T. (2026) · Applied Sciences · doi:10.3390/app16167930
Tanigawa, M., Kato, C., Nakamura, K., Kim, Y., Yoshizaki, Y., Kokai, S., & ONO, T. (2026) · BMC Oral Health · doi:10.1186/s12903-026-10126-7
Hsieh, K. Y., & Zafar, S. (2026) · ACS Omega · doi:10.1021/acsomega.6c05076
Limited predictive utility of spinal cord stimulation trial pain metrics for post-implant outcomes
Keith, A. D., Kilgore, J. S., & Provenzano, D. A. (2026) · Regional Anesthesia & Pain Medicine · doi:10.1136/rapm-2026-108028
Sun, F.-Y., Zhao, T., Shen, J.-Q., & Pan, X.-F. (2026) · Echocardiography · doi:10.1111/echo.70585
Wang, Y., Zeng, L., Xie, Y., Liu, Y., Zhang, L., & Fu, Z. (2026) · Atmospheric Environment · doi:10.1016/j.atmosenv.2026.122291
Belmans, F., Llambrich, S., Vos, W. G., Maes, F., Martínez‐Abadías, N., & Velde, G. V. (2026) · Frontiers in Medical Technology · doi:10.3389/fmedt.2026.1823048
David, E., Cabezas, L., Lara, B., Monterroso, L., & Lencina, A. (2026) · SSRN Electronic Journal · doi:10.2139/ssrn.7267482
Feng, D., Lind, A., & He, W. (2026) · Journal of Biopharmaceutical Statistics · doi:10.1080/10543406.2026.2703576
Julian, P., Walker, W. W., & Surratt, D. D. (2026) · Research Square · doi:10.21203/rs.3.rs-10589631/v1
Metadata
- Authors
- Martin Bland, J., & Altman, D.
- Venue
- The Lancet
- Publisher
- Elsevier BV
- Published
- 1986-02
- Volume
- 327
- Issue
- 8476
- Pages
- 307-310
- ISSN
- 0099-5355 (ISSN-L)
- 0140-6736
- OpenAlex
- W2015795623
- Cited by
- 47,965
- References
- 7
- Citations by year
- 2026: 1541 · 2025: 1310 · 2024: 1377 · 2023: 1382 · 2022: 1513 · 2021: 1827 · 2020: 1878 · 2019: 1797 · 2018: 1725 · 2017: 1906 · 2016: 1876 · 2015: 1983 · 2014: 2081 · 2013: 2031 · 2012: 2082
- FWCI
- 149.89
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.elsevier.com/tdm/userlicense/1.0/ (tdm)
- Locations
- The Lancet (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Social Sciences › Decision Sciences › Statistics, Probability and Uncertainty › Reliability and Agreement in Measurement
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Physical Sciences › Mathematics › Statistics and Probability › Statistical Methods in Epidemiology
- Keywords
- agreement between measurement methods, method comparison, correlation coefficient limitations, repeatability, clinical measurement
- MeSH terms
- Diagnosis, Humans, Peak Expiratory Flow Rate, Statistics as Topic
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.