The application of video image processing to quantitative analysis of extremity tremor in humans
Abstract
A new method for the detection and quantification of extremity tremor is described, based on video image processing. A single CCD camera recorded the movement of the extremity. A passive marker in the shape of a black annulus was placed on the forearm and the movement of the annulus analysed. The framestore digitised the video signal at a sample rate of 10 Hz. The time period of the movement analysis was delta t = 6.4 s. A total of 32 adults with alcoholism and 22 controls participated in this study. The movement of the extremity was recorded during the usual neurological test (sitting posture, feet together, upper extremities directly in front of subject) for both extremities. In this study, it was assumed that the probability density function f(d) of some variable D is characteristic of the tremor. This function is formed by a finite mixture of bivariate continuous distributions. The results suggest that f(d) characterises patients with alcoholism and distinguishes them from control subjects with only physiological tremor. The results demonstrate the capacity of the measuring system based on video imaging to quantifying motor impairments in clinical neurology.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Measuring devices based on video imaging are effective in quantifying motor impairments in clinical settings.
“Measuring devices based on video imaging are effective in quantifying motor impairments in clinical settings [34].”
From Neurological Tremor: Sensors, Signal Processing and Emerging Applications (2010) · cites this paper as [34]
Cites
van Someren, E. J. W., van Gool, W. A., Vonk, B. F. M., Mirmiran, M., Speelman, J. D., Bosch, D. A., & Swaab, D. F. (1993) · Journal of the Neurological Sciences · doi:10.1016/0022-510x(93)90148-r · Cited by 56 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Neurological Tremor: Sensors, Signal Processing and Emerging Applications
Grimaldi, G., & Manto, M. (2010) · Sensors · doi:10.3390/s100201399 · Cited by 138 · see citing works
“Measuring devices based on video imaging are effective in quantifying motor impairments in clinical settings [34].”
Background
- Authors
- Świder, M.
- Venue
- Journal of Neuroscience Methods
- Publisher
- Elsevier BV
- Published
- 1998-10
- Volume
- 84
- Issue
- 1-2
- Pages
- 167-172
- ISSN
- 0165-0270 (ISSN-L)
- OpenAlex
- W2160322927
- Cited by
- 11
- References
- 1
- Citations by year
- 2022: 2 · 2014: 1 · 2012: 1
- FWCI
- 0.00
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Journal of Neuroscience Methods (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Life Sciences › Neuroscience › Cognitive Neuroscience › Motor Control and Adaptation
- Health Sciences › Medicine › Neurology › Alcoholism and Thiamine Deficiency
- Health Sciences › Health Professions › Physical Therapy, Sports Therapy and Rehabilitation › Balance, Gait, and Falls Prevention
- Keywords
- tremor quantification, video processing, alcoholism, physiological tremor, probability density function, finite mixture distributions, motor impairment assessment, clinical neurology
- MeSH terms
- Forearm, Humans, Motor Activity, Movement, Probability, Tremor, Video Recording
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review