Evaluation of the Orbitrap Mass Spectrometer for the Molecular Fingerprinting Analysis of Natural Dissolved Organic Matter
The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.
Abstract
We investigated the application of the LTQ-Orbitrap mass spectrometer (LTQ-Velos Pro, Thermo Fisher) for resolving complex mixtures of natural aquatic dissolved organic matter (DOM) and compared this technique to the more established state-of-the-art technique, Fourier transform ion cyclotron resonance mass spectrometry (FTICR-MS, Bruker Daltonics), in terms of the distribution of molecular masses detected and the reproducibility of the results collected. The Orbitrap was capable of excellent reproducibility: Bray-Curtis dissimilarity between duplicate measurements was 2.85 ± 0.42% (mean ± standard deviation). The Orbitrap was also capable of the detection of most major ionizable organic molecules in typical aquatic mixtures, with the exception of most sulfur and phosphorus containing masses. This result signifies that the Orbitrap is an appropriate technique for the investigation of very subtle biogeochemical processing of bulk DOM. The lower costs (purchase and maintenance) and wider availability of Orbitrap mass spectrometers in university departments means that the tools necessary for research into DOM processing at the molecular level should be accessible to a much wider group of scientists than before. The main disadvantage of the technique is that substantially fewer molecular formulas can be resolved from a complex mixture (roughly one third as many), meaning some loss of information. In balance, most biogeochemical studies that aim at molecularly fingerprinting the source of natural DOM could be satisfactorily carried out with Orbitrap mass spectrometry. For more targeted metabolomic studies where individual compounds are traced through natural systems, FTICR-MS remains advantageous.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
High Resolution Mass Spectrometry
Xian, F., Hendrickson, C. L., & Marshall, A. G. (2012) · Analytical Chemistry · doi:10.1021/ac203191t · Cited by 281
Cited by
Zhang, B., Shan, C., Wang, S., Fang, Z., & Pan, B. (2021) · Water Research · doi:10.1016/j.watres.2020.116484 · Cited by 276
Tolu, J., Rydberg, J., Meyer-Jacob, C., Gerber, L., & Bindler, R. (2017) · Biogeosciences · doi:10.5194/bg-14-1773-2017 · Cited by 30
“Advanced ultra-high-resolution MS techniques, i.e., Fourier transform ion cyclotron resonance mass spectrometry (FTICR-MS) or linear trap quadruple Orbitrap MS, enable the determination of a large number of organic molecular formulas in liquid samples (> 1000; e.g., Hawkes et al., 2016).”
Method
Metadata
- Authors
- Hawkes, J. A., Dittmar, T., Patriarca, C., Tranvik, L., & Bergquist, J.
- Venue
- Analytical Chemistry
- Publisher
- American Chemical Society (ACS)
- Published
- 2016-07-22
- Volume
- 88
- Issue
- 15
- Pages
- 7698-7704
- ISSN
- 0003-2700 (ISSN-L)
- 1520-6882
- OpenAlex
- W2471379059
- Cited by
- 172
- References
- 1
- FWCI
- 6.16
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.