The scanning electron microscope in microbiology and diagnosis of infectious disease

Golding, Christine G., Lamboo, Lindsey L., Beniac, Daniel R., Booth, Timothy F.·2016·Scientific Reports

Abstract

Despite being an excellent tool for investigating ultrastructure, scanning electron microscopy (SEM) is less frequently used than transmission electron microscopy for microbes such as viruses or bacteria. Here we describe rapid methods that allow SEM imaging of fully hydrated, unfixed microbes without using conventional sample preparation methods. We demonstrate improved ultrastructural preservation, with greatly reduced dehydration and shrinkage, for specimens including bacteria and viruses such as Ebola virus using infiltration with ionic liquid on conducting filter substrates for SEM.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1038/srep26516

Cites

Statistics

Figures

Figure 1 Comparison of conventional sputter coating SEM sample preparation methods (panels on the left-hand side) with ionic liquid treatment (centre panels) and conventional TEM (panels on the right-hand side) for the observation of microbes: (a) Leptospira biflexa, (b) Salmonella Senftenberg, (c) vaccinia, and (d) Ebola virus. The SEM images in the left-hand side panels were of specimens that were sputter coated with gold, on plain uncoated filters. Images in the centre panels were of specimens treated with ionic liquid after deposition on pre-coated aluminum filters. On the right-hand side, TEM images of similar specimens prepared using methylamine tungstate negative staining.
Figure 1 Comparison of conventional sputter coating SEM sample preparation methods (panels on the left-hand side) with ionic liquid treatment (centre panels) and conventional TEM (panels on the right-hand side) for the observation of microbes: (a) Leptospira biflexa, (b) Salmonella Senftenberg, (c) vaccinia, and (d) Ebola virus. The SEM images in the left-hand side panels were of specimens that were sputter coated with gold, on plain uncoated filters. Images in the centre panels were of specimens treated with ionic liquid after deposition on pre-coated aluminum filters. On the right-hand side, TEM images of similar specimens prepared using methylamine tungstate negative staining.CC BY · PubMed Central
Figure 2 Preparation of biological samples for SEM. (a) The components of the filter unit are shown before assembly. The inset SEM image shows a gold coated filter at high magnification before a specimen is applied. Note that the filter is clean and the pores are clearly evident. (b) The filter unit is shown after assembly, and in use with a syringe pump in a biosafety cabinet (c,d). The blue arrow in (d) points to the filter unit. (e) Images of filters that have had metal evaporated on them. The thickness of the Al is 18 nm, and 9 nm, and the Au is 27 nm and 9 nm thick. For the filter with both metals the thickness of the Al and Au are 18 nm and 27 nm, respectively. (f) SEM and (g) the corresponding elemental map generated by X-ray microanalysis of a region similar to the one highlighted by the dotted rectangle in (e). (h–k) SEM images of Salmonella stained with ionic liquid illustrating the effect of different metal types, and thickness of metal evaporated on the final images recorded (h Al 9 nm, i Au 9 nm, j Al 18 nm, k Au 27 nm). Red arrows indicate flagellae.
Figure 2 Preparation of biological samples for SEM. (a) The components of the filter unit are shown before assembly. The inset SEM image shows a gold coated filter at high magnification before a specimen is applied. Note that the filter is clean and the pores are clearly evident. (b) The filter unit is shown after assembly, and in use with a syringe pump in a biosafety cabinet (c,d). The blue arrow in (d) points to the filter unit. (e) Images of filters that have had metal evaporated on them. The thickness of the Al is 18 nm, and 9 nm, and the Au is 27 nm and 9 nm thick. For the filter with both metals the thickness of the Al and Au are 18 nm and 27 nm, respectively. (f) SEM and (g) the corresponding elemental map generated by X-ray microanalysis of a region similar to the one highlighted by the dotted rectangle in (e). (h–k) SEM images of Salmonella stained with ionic liquid illustrating the effect of different metal types, and thickness of metal evaporated on the final images recorded (h Al 9 nm, i Au 9 nm, j Al 18 nm, k Au 27 nm). Red arrows indicate flagellae.CC BY · PubMed Central

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

No references could be resolved yet.

Research works printed but not linked: 34

  1. van Rooyen, C. E. & SCOTT, G. D. Smallpox diagnosis with special reference to electron microscopy. Can. J. Public Health 39, 467–477 (1948).
  2. Goldsmith, C. S. & Miller, S. E. Modern uses of electron microscopy for detection of viruses. Clin. Microbiol. Rev. 22, 552–563, 10.1128/CMR.00027-09 (2009). · DOI
  3. Hazelton, P. R. & Gelderblom, H. R. Electron microscopy for rapid diagnosis of infectious agents in emergent situations. Emerg. Infect. Dis. 9, 294–303 (2003). · DOI
  4. Nitsche, A., Gelderblom, H. R., Eisendle, K., Romani, N. & Pauli, G. Pitfalls in diagnosing human poxvirus infections. J. Clin. Virol. 38, 165–168 (2007). · DOI
  5. Biel, S. S. et al. Detection of human polyomaviruses in urine from bone marrow transplant patients: comparison of electron microscopy with PCR. Clin. Chem. 50, 306–312 (2004). · DOI
  6. Biel, S. S. & Gelderblom, H. R. Diagnostic electron microscopy is still a timely and rewarding method. J. Clin. Virol. 13, 105–119 (1999). · DOI
  7. Beniac, D. R., Siemens, C. G., Wright, C. J. & Booth, T. F. A filtration based technique for simultaneous SEM and TEM sample preparation for the rapid detection of pathogens. Viruses. 6, 3458–3471, 10.3390/v6093458 (2014). · DOI
  8. Beniac, D. R., Hiebert, S. L., Siemens, C. G., Corbett, C. R. & Booth, T. F. A mobile biosafety microanalysis system for infectious agents. Sci. Rep. 5, 9505, 10.1038/srep09505 (2015). · DOI
  9. Thiberge, S. et al. Scanning electron microscopy of cells and tissues under fully hydrated conditions. Proc. Natl. Acad. Sci. USA 101, 3346–3351 (2004). · DOI
  10. de Jonge, N., Peckys, D. B., Kremers, G. J. & Piston, D. W. Electron microscopy of whole cells in liquid with nanometer resolution. Proc. Natl. Acad. Sci. USA 106, 2159–2164, 10.1073/pnas.0809567106 (2009). · DOI
  11. Ogura, T. Nanoscale analysis of unstained biological specimens in water without radiation damage using high-resolution frequency transmission electric-field system based on FE-SEM. Biochem. Biophys. Res. Commun. 459, 521–528, 10.1016/j.bbrc.2015.02.140 (2015). · DOI
  12. Bergmans, L., Moisiadis, P., Van, M. B., Quirynen, M. & Lambrechts, P. Microscopic observation of bacteria: review highlighting the use of environmental SEM. Int. Endod. J. 38, 775–788 (2005). · DOI
  13. Walther, P. & Muller, M. Biological ultrastructure as revealed by high resolution cryo-SEM of block faces after cryo-sectioning. J. Microsc. 196, 279–287 (1999). · DOI
  14. Hayles, M. F., Stokes, D. J., Phifer, D. & Findlay, K. C. A technique for improved focused ion beam milling of cryo-prepared life science specimens. J. Microsc. 226, 263–269 (2007). · DOI
  15. Katsen-Globa, A., Puetz, N., Gepp, M. M., Neubauer, J. C. & Zimmermann, H. Study of SEM preparation artefacts with correlative microscopy: Cell shrinkage of adherent cells by HMDS-drying. Scanning, 10.1002/sca.21310 (2016). · DOI
  16. Beniac, D. R. et al. The organisation of Ebola virus reveals a capacity for extensive, modular polyploidy. PLoS. One. 7, e29608, 10.1371/journal.pone.0029608 (2012). · DOI
  17. Cotelesage, J. J. et al. Hepatitis B Virus Genotype G forms core-like particles with unique structural properties. J. Viral Hepat. 18, 443–448, 10.1111/j.1365-2893.2010.01330.x (2010). · DOI
  18. Frank, J. et al. SPIDER and WEB: processing and visualization of images in 3D electron microscopy and related fields. J. Struct. Biol. 116, 190–199 (1996). · DOI
  19. Kuwabata, S., Kongkanand, A., Oyamatsu, D. & Torimoto, T. Observation of ionic liquid by scanning electron microscope. Chemistry Letters 35, 600–601, 10.1246/cl.2006.600 (2006). · DOI
  20. Dwiranti, A. et al. Chromosome observation by scanning electron microscopy using ionic liquid. Microsc. Res. Tech. 75, 1113–1118, 10.1002/jemt.22038 (2012). · DOI
  21. Takahashi, C., Shirai, T. & Fuji, M. FE-SEM observation of swelled seaweed using hydrophilic ionic liquid; 1-butyl-3-methylimidazolium tetrafluoroborate. Microsc. Res. Tech. 76, 66–71, 10.1002/jemt.22137 (2013). · DOI
  22. Ishigaki, Y. et al. Ionic liquid enables simple and rapid sample preparation of human culturing cells for scanning electron microscope analysis. Microsc. Res. Tech. 74, 415–420, 10.1002/jemt.20924 (2011). · DOI
  23. Kawai, K., Kaneko, K., Kawakami, H., Narushima, T. & Yonezawa, T. Simple pretreatment of non-conductive small hydrous bio-samples with choline-type ionic liquid and membrane filter for microsample mounting. Colloids Surf. B Biointerfaces. 102, 9–12, 10.1016/j.colsurfb.2012.08.019 (2013). · DOI
  24. Torimoto, T., Tsuda, T., Okazaki, K. & Kuwabata, S. New frontiers in materials science opened by ionic liquids. Adv. Mater. 22, 1196–1221, 10.1002/adma.200902184 (2010). · DOI
  25. Arimoto, S., Oyamatsu, D., Torimoto, T. & Kuwabata, S. Development of in situ electrochemical scanning electron microscopy with ionic liquids as electrolytes. Chemphyschem. 9, 763–767, 10.1002/cphc.200700758 (2008). · DOI
  26. Klungthong, C. et al. The impact of primer and probe-template mismatches on the sensitivity of pandemic influenza A/H1N1/2009 virus detection by real-time RT-PCR. J Clin. Virol. 48, 91–95, 10.1016/j.jcv.2010.03.012 (2010). · DOI
  27. Pyne, M. T., Brown, K. L. & Hillyard, D. R. Evaluation of the Roche Cobas AmpliPrep/Cobas TaqMan HIV-1 test and identification of rare polymorphisms potentially affecting assay performance. J Clin. Microbiol. 48, 2852–2858, 10.1128/JCM.00776-10 (2010). · DOI
  28. Mahony, J. B. & Richardson, S. Molecular diagnosis of severe acute respiratory syndrome: the state of the art. J Mol. Diagn. 7, 551–559 (2005). · DOI
  29. Gelderblom, H. R. & Hazelton, P. R. Specimen collection for electron microscopy. Emerg. Infect. Dis. 6, 433–434 (2000). · DOI
  30. Goldstein, J. I. et al. Scanning Electron Microscopy and X-ray Microanalysis (Springer, New York 2003). · DOI
  31. Levett, P. N. Leptospirosis. Clin. Microbiol. Rev. 14, 296–326 (2001). · DOI
  32. Fàbrega, A. & Vila, J. Salmonella enterica serovar Typhimurium skills to succeed in the host: virulence and regulation. Clin. Microbiol. Rev. 26, 308–341, 10.1128/CMR.00066-12 (2013). · DOI
  33. Halfmann, P. et al. Replication-deficient ebolavirus as a vaccine candidate. J. Virol. 83, 3810–3815, 10.1128/JVI.00074-09 (2009). · DOI
  34. Abramoff, M. D., Magelhaes, P. J. & Ram, S. J. Image Processing with ImageJ. Biophotonics International 11, 36–42 (2004).

Cited by

Sorted by citations

Metadata

Authors
Golding, C. G., Lamboo, L. L., Beniac, D. R., & Booth, T. F.
Venue
Scientific Reports
Published
2016-05-23
Volume
6
Issue
1
ISSN
2045-2322
Cited by
236
References
0
Language
en

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review