The DOI matches its Crossref record; 3 of the 3 references it cites were found in a registry or an index.
Small-bowel capsule endoscopy and device-assisted enteroscopy for diagnosis and treatment of small-bowel disorders: European Society of Gastrointestinal Endoscopy (ESGE) Clinical Guideline
Open accessAbstract
SMALL-BOWEL CAPSULE ENDOSCOPY (SBCE): 1: ESGE recommends that prior to SBCE patients ingest a purgative (2 L of polyethylene glycol [PEG]) for better visualization.Strong recommendation, high quality evidence.However, the optimal timing for taking purgatives is yet to be established. 2: ESGE recommends that SBCE should be performed as an outpatient procedure if possible, since completion rates are higher in outpatients than in inpatients.Strong recommendation, moderate quality evidence. 3: ESGE recommends that patients with pacemakers can safely undergo SBCE without special precautions.Strong recommendation, low quality evidence. 4: ESGE suggests that SBCE can also be safely performed in patients with implantable cardioverter defibrillators and left ventricular assist devices.Weak recommendation, low quality evidence. 5: ESGE recommends the acceptance of qualified nurses and trained technicians as prereaders of capsule endoscopy studies as their competency in identifying pathology is similar to that of medically qualified readers. The responsibility of establishing a diagnosis must however remain with the attending physician.Strong recommendation, moderate quality evidence. 6: ESGE recommends observation in cases of asymptomatic capsule retention.Strong recommendation, moderate quality evidence.In cases where capsule retrieval is indicated, ESGE recommends the use of device-assisted enteroscopy as the method of choice.Strong recommendation, moderate quality evidence. DEVICE-ASSISTED ENTEROSCOPY (DAE): 1: ESGE recommends performing diagnostic DAE as a day-case procedure in patients without significant underlying co-morbidities; in patients with co-morbidities and/or those undergoing a therapeutic procedure, an inpatient stay is recommended.Strong recommendation, low quality evidenceThe choice between different settings also depends on sedation protocols.Strong recommendation, low quality evidence. 2: ESGE suggests that conscious sedation, deep sedation, and general anesthesia are all acceptable alternatives: the choice between them should be governed by procedure complexity, clinical factors, and local organizational protocols.Weak recommendation, low quality evidence. 3: ESGE recommends that the findings of previous diagnostic investigations should guide the choice of insertion route.Strong recommendation, moderate quality evidence.If the location of the small-bowel lesion is unknown or uncertain, ESGE recommends that the antegrade route should be generally preferred.Strong recommendation, low quality evidence.In the setting of massive overt bleeding, ESGE recommends an initial antegrade approach.Strong recommendation, low quality evidence. 4: ESGE recommends that, for balloon-assisted enteroscopy (i. e., single-balloon enteroscopy [SBE] and double-balloon enteroscopy [DBE]), small-bowel insertion depth should be estimated by counting net advancement of the enteroscope during the insertion phase, with confirmation of this estimate during withdrawal.Strong recommendation, low quality evidence.ESGE recommends that, for spiral enteroscopy, insertion depth should be estimated during withdrawal.Strong recommendation, moderate quality evidence. Since the calculated insertion depth is only a rough estimate, ESGE recommends placing a tattoo to mark the identified lesion and/or the deepest point of insertion.Strong recommendation, low quality evidence. 5: ESGE recommends that all endoscopic therapeutic procedures can be undertaken at the time of DAE.Strong recommendation, moderate quality evidence.Moreover, when therapeutic interventions are performed, additional specific safety measures are needed to prevent complications.Strong recommendation, high quality evidence.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Grading quality of evidence and strength of recommendations
(2004) · BMJ · doi:10.1136/bmj.328.7454.1490 · Cited by 8,477
Recommendations of the Panel on Cost-Effectiveness in Health and Medicine
Weinstein, M. C. (1996) · JAMA: The Journal of the American Medical Association · doi:10.1001/jama.1996.03540150055031 · Cited by 2,455 · cited 1×
“Cost-effectiveness analyses are intended to support resource allocation decisions and are therefore dependent on local/regional socioeconomic perspectives [231].”
BackgroundYersinia enterocolitica Infection with Multiple Liver Abscesses Uncovering a Primary Hemochromatosis
Höpfner, M., Nitsche, R., Rohr, A., Harms, D., Schubert, S., & Fölsch, U. R. (2001) · Scandinavian Journal of Gastroenterology · doi:10.1080/003655201750066004 · Cited by 56
Cited by
Lamb, C. A., Kennedy, N. A., Raine, T., Hendy, P. A., Smith, P. J., Limdi, J. K., Hayee, B., Lomer, M. C. E., Parkes, G. C., Selinger, C., et al. (2019) · Gut · doi:10.1136/gutjnl-2019-318484 · Cited by 2,346
Arasaradnam, R. P., Brown, S., Forbes, A., Fox, M. R., Hungin, P., Kelman, L., Major, G., O’Connor, M., Sanders, D. S., Sinha, R., et al. (2018) · Gut · doi:10.1136/gutjnl-2017-315909 · Cited by 332
Self-citationArtificial intelligence in gastrointestinal endoscopy: general overview
El Hajjar, A., & Rey, J.-F. (2020) · Chinese Medical Journal · doi:10.1097/cm9.0000000000000623 · Cited by 102
Enteric neuropathy in diabetes: Implications for gastrointestinal function
Abdalla, M. M. I. (2024) · World Journal of Gastroenterology · doi:10.3748/wjg.v30.i22.2852 · Cited by 45
Defining gastrointestinal transit time using video capsule endoscopy: A study of healthy subjects
O’Grady, J., Murphy, C. L., Barry, L., Shanahan, F., & Buckley, M. (2020) · Endoscopy International Open · doi:10.1055/a-1073-7653 · Cited by 42
Metadata
- Authors
- Pennazio, M., Spada, C., Eliakim, R., Keuchel, M., May, A., Mulder, C., Rondonotti, E., Adler, S., Albert, J., Baltes, P., Barbaro, F., Cellier, C., Charton, J., Delvaux, M., Despott, E., Domagk, D., Klein, A., McAlindon, M., Rosa, B., Rowse, G., Sanders, D., Saurin, J., Sidhu, R., Dumonceau, J.-M., Hassan, C., & Gralnek, I.
- Venue
- Endoscopy
- Publisher
- Georg Thieme Verlag KG
- Published
- 2015-03-31
- Volume
- 47
- Issue
- 04
- Pages
- 352-386
- ISSN
- 0013-726X (ISSN-L)
- 1438-8812
- OpenAlex
- W2015433732
- Cited by
- 894
- References
- 3
- FWCI
- 75.65
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Small-bowel capsule endoscopy and device-assisted enteroscopy for diagnosis and treatment of small-bowel disorders: European Society of Gastrointestinal Endoscopy (ESGE) Clinical Guideline. https://app.cochat.ai/papers/10.1055/s-0034-1391855
- The paper
- Pennazio, M., Spada, C., Eliakim, R., Keuchel, M., May, A., Mulder, C., Rondonotti, E., Adler, S., Albert, J., Baltes, P., Barbaro, F., Cellier, C., Charton, J., Delvaux, M., Despott, E., Domagk, D., Klein, A., McAlindon, M., Rosa, B., . . . Gralnek, I. (2015). Small-bowel capsule endoscopy and device-assisted enteroscopy for diagnosis and treatment of small-bowel disorders: European Society of Gastrointestinal Endoscopy (ESGE) Clinical Guideline. Endoscopy. https://doi.org/10.1055/s-0034-1391855
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.