Twelve-Year Comparison of a Bjork–Shiley Mechanical Heart Valve with Porcine Bioprostheses
Open accessThe DOI matches its Crossref record; 6 of the 6 references it cites were found in a registry or an index.
Abstract
BACKGROUND: Patients undergoing heart-valve replacement may receive a mechanical prosthesis, necessitating lifelong anticoagulant treatment, or a porcine bioprosthesis, which involves no absolute need for anticoagulants. METHODS: We carried out a randomized, prospective trial to compare the durability of the Bjork-Shiley mechanical prosthesis (spherical tilting-disk model) and the incidence of valve-related complications with those variables in the Hancock and the Carpentier-Edwards porcine prostheses. The mitral valve was replaced in 261 patients, the aortic valve in 211, and both in 61; the survivors have been followed up for a mean of 12 years. RESULTS: We found a trend toward improved actuarial survival after 12 years with the Bjork-Shiley prosthesis, but this trend was not statistically significant (group with Bjork-Shiley valve vs. group with porcine valve [mean +/- SE], 51.5 +/- 3.2 vs. 44.4 +/- 3.2 percent; P = 0.08). There was no significant difference in the actuarial incidence of reoperation after 5 years, but after 12 years significantly more patients with a porcine prosthesis had undergone reoperation (8.5 +/- 2.0 vs. 37.1 +/- 4.1 percent, P less than 0.001). An analysis combining death and reoperation as end points for an actuarial assessment of survival with the original prosthesis intact confirmed that the patients with Bjork-Shiley Shiley prostheses had improved survival (48.6 +/- 3.2 vs. 30.0 +/- 3.0 percent after 12 years, P less than 0.001). Bleeding requiring hospitalization or blood transfusion was significantly more frequent in the patients with Bjork-Shiley prostheses (18.6 +/- 3.2 vs. 7.1 +/- 2.3 percent after 12 years, P less than 0.01). There was no significant difference after 12 years in the actuarial occurrence of embolism (Bjork-Shiley vs. porcine, 21.1 +/- 3.1 vs. 26.4 +/- 3.5 percent) or endocarditis (3.7 +/- 1.4 vs 4.6 +/- 1.6 percent). CONCLUSIONS: Survival with an intact valve is better among patients with the Bjork-Shiley spherical tilting-disk prosthesis than among patients with porcine bioprostheses, but use of the Bjork-Shiley valve carries an attendant increased risk of bleeding associated with the need for anticoagulant treatment.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Regression Models and Life-Tables
Cox, D. R. (1972) · Journal of the Royal Statistical Society Series B: Statistical Methodology · doi:10.1111/j.2517-6161.1972.tb00899.x · Cited by 39,793
Jamieson, W. R. E., Rosado, L. J., Munro, A. I., Gerein, A. N., Burr, L. H., Miyagishima, R. T., Janusz, M. T., & Tyers, G. F. O. (1988) · The Annals of Thoracic Surgery · doi:10.1016/s0003-4975(10)65888-2 · Cited by 191
The porcine bioprosthetic valve
Magilligan, D. J., Lewis, J. W., Tilley, B., & Peterson, E. (1985) · The Journal of Thoracic and Cardiovascular Surgery · doi:10.1016/s0022-5223(19)38753-7 · Cited by 168
The Ultimate Prognosis after Valve Replacement: An Assessment at Twenty Years
Teply, J. F., Grunkemeier, G. L., Sutherland, H. D., Lambert, L. E., Johnson, V. A., & STARR, A. M. (1981) · The Annals of Thoracic Surgery · doi:10.1016/s0003-4975(10)61017-x · Cited by 139
Perspective on valvular heart disease: An update
Rahimtoola, S. H. (1989) · Journal of the American College of Cardiology · doi:10.1016/0735-1097(89)90047-8 · Cited by 134
Is the Dose of Warfarin Prescribed by American Physicians Unnecessarily High?
Hirsh, J. (1987) · Archives of Internal Medicine · doi:10.1001/archinte.1987.00370040151026 · Cited by 60
Cited by
Bonow, R. O., Carabello, B. A., Chatterjee, K., de Leon, A. C., Faxon, D. P., Freed, M. D., Gaasch, W. H., Lytle, B. W., Nishimura, R. A., O’Gara, P. T., et al. (2006) · Circulation · doi:10.1161/circulationaha.106.176857 · Cited by 8,495
Authors/Task Force Members, Vahanian, A., Baumgartner, H., Bax, J., Butchart, E., Dion, R., Filippatos, G., Flachskampf, F., Hall, R., Iung, B., et al. (2006) · European Heart Journal · doi:10.1093/eurheartj/ehl428 · Cited by 3,593
Hench, L. L. (1998) · Journal of the American Ceramic Society · doi:10.1111/j.1151-2916.1998.tb02540.x · Cited by 2,652
Nishimura, R. A., Otto, C. M., Bonow, R. O., Carabello, B. A., Erwin, J. P., Guyton, R. A., O’Gara, P. T., Ruiz, C. E., Skubas, N. J., Sorajja, P., et al. (2014) · Circulation · doi:10.1161/cir.0000000000000029 · Cited by 2,566
Bonow, R. O., Carabello, B. A., Chatterjee, K., de Leon, A. C., Faxon, D. P., Freed, M. D., Gaasch, W. H., Lytle, B. W., Nishimura, R. A., O'Gara, P. T., et al. (2008) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2008.05.007 · Cited by 2,257
Metadata
- Authors
- Bloomfield, P., Wheatley, D. J., Prescott, R. J., & Miller, H. C.
- Venue
- New England Journal of Medicine
- Publisher
- Massachusetts Medical Society
- Published
- 1991-02-28
- Volume
- 324
- Issue
- 9
- Pages
- 573-579
- ISSN
- 0028-4793 (ISSN-L)
- 1533-4406
- OpenAlex
- W2327876284
- Cited by
- 428
- References
- 6
- FWCI
- 13.26
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.