CoChat

The extended surface forces apparatus. II. Precision temperature control

Heuberger, M., Vanicek, J., Zäch, M. · 2001 · Review of Scientific Instruments
Cited by 47

The DOI matches its Crossref record.

Abstract

We describe the design and performance of a low-cost, bidirectional temperature control system with very high stability and low thermal gradients in a working range from −10 to 70 °C. System elements such as thermal insulation, thermal pumps, and feedback control as well as safety features are discussed. The performance of the entire system under realistic experimental conditions is documented. Precise temperature control has led to significantly improved instrumental stability. An extension of the instrumental window towards slower measurements has been made possible. This temperature-control system is presented in the framework of the extended surface forces apparatus but is also applicable to other scientific instruments that require subnanometer positioning stability over extended periods of time.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1063/1.1392340

Cites

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

See all 47

Metadata

Authors
Heuberger, M., Vanicek, J., & Zäch, M.
Venue
Review of Scientific Instruments
Published
2001-09-01
Volume
72
Issue
9
Pages
3556-3560
ISSN
  • 0034-6748 (ISSN-L)
  • 1089-7623
OpenAlex
W2014266295
Cited by
47
References
0
FWCI
0.71
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: The extended surface forces apparatus. II. Precision temperature control. https://app.cochat.ai/papers/10.1063/1.1392340
The paper
Heuberger, M., Vanicek, J., & Zäch, M. (2001). The extended surface forces apparatus. II. Precision temperature control. Review of Scientific Instruments. https://doi.org/10.1063/1.1392340

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.