Examination of four different instruments for measuring blood lactate concentration
Open accessThe DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.
Abstract
Information on the performance of different instruments used to measure blood lactate concentration is incomplete. We therefore examined instruments from Yellow Springs Instruments (YSI 23L and YSI 1500) and three cheaper and simpler instruments: Dr. Lange's LP8+, Lactate Pro from Arkray in the KDK corporation and Accusport from Boehringer Mannheim. First, a number of blood samples were analysed by standard enzymatic photofluorometry (our reference method) and, in addition, by one or more of the instruments mentioned above. Second, measurements using two or more identical instruments were compared. Third, since Lactate Pro and Accusport are small (approximately 100 g, pocket-size), battery-driven, instruments that could be used for outdoor testing, the performance of these instruments was examined at simulated altitudes (O2 pressure of <10 kPa) and at temperatures below -20 degrees C, while screening the instruments as much as possible from the cold. Most of the different instruments showed systematically too high or too low values (10-25% deviation). The observed differences between instruments may affect the "blood lactate threshold" by 2-5%. We found different readings between "equal" YSI 1500 instruments, while we could see no difference when comparing the other instruments of the same type. Lactate Pro gave reliable results at both -21+/-1 degrees C and at simulated altitude. Accusport gave reliable results in the cold, but 1.85+/-0.08 mmol L(-1) (mean+/-SD) too high readings at the simulated altitude. Of the three simpler instruments examined, the Lactate Pro was at least as good as the YSI instruments and superior to the other two.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Cited by
Lactate Threshold Concepts: How Valid are They?
Faude, O., Kindermann, W., & Meyer, T. (2009) · Sports Medicine · doi:10.2165/00007256-200939060-00003 · Cited by 882
Seiler, K. S., & Kjerland, G. Ø. (2006) · Scandinavian Journal of Medicine & Science in Sports · doi:10.1111/j.1600-0838.2004.00418.x · Cited by 643
Blood Lactate Measurements and Analysis during Exercise: A Guide for Clinicians
Goodwin, M. L., Harris, J. E., Hernández, A., & Gladden, L. B. (2007) · Journal of Diabetes Science and Technology · doi:10.1177/193229680700100414 · Cited by 554
Edvardsen, E., Hem, E., & Anderssen, S. A. (2014) · PLoS ONE · doi:10.1371/journal.pone.0085276 · Cited by 366
Caffeine Supplementation and Multiple Sprint Running Performance
GLAISTER, M., HOWATSON, G., ABRAHAM, C. S., LOCKEY, R. A., GOODWIN, J. E., FOLEY, P., & MCINNES, G. (2008) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0b013e31817a8ad2 · Cited by 163
Metadata
- Authors
- Medbø, J. I., Mamen, A., Holt Olsen, O., & Evertsen, E.
- Venue
- Scandinavian Journal of Clinical and Laboratory Investigation
- Publisher
- Informa UK Limited
- Published
- 2000-01
- Volume
- 60
- Issue
- 5
- Pages
- 367-380
- ISSN
- 0036-5513 (ISSN-L)
- 1502-7686
- OpenAlex
- W2106989220
- Cited by
- 122
- References
- 1
- FWCI
- 0.57
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Examination of four different instruments for measuring blood lactate concentration. https://app.cochat.ai/papers/10.1080/003655100750019279
- The paper
- Medbø, J. I., Mamen, A., Holt Olsen, O., & Evertsen, E. (2000). Examination of four different instruments for measuring blood lactate concentration. Scandinavian Journal of Clinical and Laboratory Investigation. https://doi.org/10.1080/003655100750019279
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.