Examination of four different instruments for measuring blood lactate concentration

Open access
Medbø, J. I., Mamen, A., Holt Olsen, O., Evertsen, E. · 2000 · Scandinavian Journal of Clinical and Laboratory Investigation
Cited by 122

The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.

Abstract

Information on the performance of different instruments used to measure blood lactate concentration is incomplete. We therefore examined instruments from Yellow Springs Instruments (YSI 23L and YSI 1500) and three cheaper and simpler instruments: Dr. Lange's LP8+, Lactate Pro from Arkray in the KDK corporation and Accusport from Boehringer Mannheim. First, a number of blood samples were analysed by standard enzymatic photofluorometry (our reference method) and, in addition, by one or more of the instruments mentioned above. Second, measurements using two or more identical instruments were compared. Third, since Lactate Pro and Accusport are small (approximately 100 g, pocket-size), battery-driven, instruments that could be used for outdoor testing, the performance of these instruments was examined at simulated altitudes (O2 pressure of <10 kPa) and at temperatures below -20 degrees C, while screening the instruments as much as possible from the cold. Most of the different instruments showed systematically too high or too low values (10-25% deviation). The observed differences between instruments may affect the "blood lactate threshold" by 2-5%. We found different readings between "equal" YSI 1500 instruments, while we could see no difference when comparing the other instruments of the same type. Lactate Pro gave reliable results at both -21+/-1 degrees C and at simulated altitude. Accusport gave reliable results in the cold, but 1.85+/-0.08 mmol L(-1) (mean+/-SD) too high readings at the simulated altitude. Of the three simpler instruments examined, the Lactate Pro was at least as good as the YSI instruments and superior to the other two.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1080/003655100750019279

Cites

1 of 1 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

See all 122

Metadata

Authors
Medbø, J. I., Mamen, A., Holt Olsen, O., & Evertsen, E.
Venue
Scandinavian Journal of Clinical and Laboratory Investigation
Published
2000-01
Volume
60
Issue
5
Pages
367-380
ISSN
  • 0036-5513 (ISSN-L)
  • 1502-7686
OpenAlex
W2106989220
Cited by
122
References
1
FWCI
0.57
Type
journal article
Language
en
Open access
green

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Examination of four different instruments for measuring blood lactate concentration. https://app.cochat.ai/papers/10.1080/003655100750019279
The paper
Medbø, J. I., Mamen, A., Holt Olsen, O., & Evertsen, E. (2000). Examination of four different instruments for measuring blood lactate concentration. Scandinavian Journal of Clinical and Laboratory Investigation. https://doi.org/10.1080/003655100750019279

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.