Brain mechanical property measurement using MRE with intrinsic activation
Open accessAbstract
Many pathologies alter the mechanical properties of tissue. Magnetic resonance elastography (MRE) has been developed to noninvasively characterize these quantities in vivo. Typically, small vibrations are induced in the tissue of interest with an external mechanical actuator. The resulting displacements are measured with phase contrast sequences and are then used to estimate the underlying mechanical property distribution. Several MRE studies have quantified brain tissue properties. However, the cranium and meninges, especially the dura, are very effective at damping externally applied vibrations from penetrating deeply into the brain. Here, we report a method, termed 'intrinsic activation', that eliminates the requirement for external vibrations by measuring the motion generated by natural blood vessel pulsation. A retrospectively gated phase contrast MR angiography sequence was used to record the tissue velocity at eight phases of the cardiac cycle. The velocities were numerically integrated via the Fourier transform to produce the harmonic displacements at each position within the brain. The displacements were then reconstructed into images of the shear modulus based on both linear elastic and poroelastic models. The mechanical properties produced fall within the range of brain tissue estimates reported in the literature and, equally important, the technique yielded highly reproducible results. The mean shear modulus was 8.1 kPa for linear elastic reconstructions and 2.4 kPa for poroelastic reconstructions where fluid pressure carries a portion of the stress. Gross structures of the brain were visualized, particularly in the poroelastic reconstructions. Intra-subject variability was significantly less than the inter-subject variability in a study of six asymptomatic individuals. Further, larger changes in mechanical properties were observed in individuals when examined over time than when the MRE procedures were repeated on the same day. Cardiac pulsation, termed intrinsic activation, produces sufficient motion to allow mechanical properties to be recovered. The poroelastic model is more consistent with the measured data from brain at low frequencies than the linear elastic model. Intrinsic activation allows MRE to be performed without a device shaking the head so the patient notices no differences between it and the other sequences in an MR examination.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The cited source says what the claim attributes to it.
In the citing paper
“Weaver and colleagues showed that intrinsic brain motion generated by cardiac pulsatility can also be exploited to estimate mechanical properties (Weaver et al., 2012; Zorgani et al., 2015).”
From 40 Years of diffusion MRI in the brain: From history to emerging frontiers (2026) · cites this paper as Weaver et al., 2012
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Elastography: A Quantitative Method for Imaging the Elasticity of Biological Tissues
Jonathan Ophir, A., Céspedes, I., Ponnekanti, H., Yazdi, Y., & Li, X. (1991) · Ultrasonic Imaging · doi:10.1177/016173469101300201 · Cited by 3,780
Magnetic Resonance Elastography by Direct Visualization of Propagating Acoustic Strain Waves
Muthupillai, R., Lomas, D. J., Rossman, P. J., Greenleaf, J. F., Manduca, A., & Ehman, R. L. (1995) · Science · doi:10.1126/science.7569924 · Cited by 2,116
Physical properties of tissue : a comprehensive reference book
Duck, F. A. (1990) · Cited by 1,675
Textbook of Medical Physiology.
(1986) · Annals of Internal Medicine · doi:10.7326/0003-4819-104-5-734_3 · Cited by 981
Elastography of breast lesions: initial clinical results.
Garra, B. S., Céspedes, E. I., Jonathan Ophir, A., Spratt, S. R., Zuurbier, R. A., Magnant, C. M., & Pennanen, M. F. (1997) · Radiology · doi:10.1148/radiology.202.1.8988195 · Cited by 940
High-resolution tensor MR elastography for breast tumour detection
Sinkus, R., Lorenzen, J., Schrader, D., Lorenzen, M., Dargatz, M., & Holz, D. (2000) · Physics in Medicine and Biology · doi:10.1088/0031-9155/45/6/317 · Cited by 481
Tissue response to mechanical vibrations for “sonoelasticity imaging”
Parker, K. J., Huang, S. R., Musulin, R. A., & Lerner, R. M. (1990) · Ultrasound in Medicine & Biology · doi:10.1016/0301-5629(90)90003-u · Cited by 478
MR Elastography of the Liver: Preliminary Results
Rouvière, O., Yin, M., Dresner, M. A., Rossman, P. J., Burgart, L. J., Fidler, J. L., & Ehman, R. L. (2006) · Radiology · doi:10.1148/radiol.2402050606 · Cited by 446
Magnetic resonance elastography of the brain
Kruse, S. A., Rose, G. H., Glaser, K. J., Manduca, A., Felmlee, J. P., Jack, C. R., & Ehman, R. L. (2007) · NeuroImage · doi:10.1016/j.neuroimage.2007.08.030 · Cited by 426
The impact of aging and gender on brain viscoelasticity
Sack, I., Beierbach, B., Wuerfel, J., Klatt, D., Hamhaber, U., Papazoglou, S., Martus, P., & Braun, J. (2009) · NeuroImage · doi:10.1016/j.neuroimage.2009.02.040 · Cited by 423
Cited by
Sorted by citations
Items 101–102 of 102
Optimal conditions and sensitivity of brain MRE : from homogeneous to heterogeneous media
Andoh, F. (2021) · theses.fr (ABES) · doi:10.70675/8295b051z541cz4fa9z833ezf06b80b85bff
In vivo elastography of the human retina using light-evoked intrinsic actuationPreprint
Liu, T., Li, H., Pandiyan, V. P., Chen, K. C., Bharadwaj, P., Wendel, B. J., Mustafi, D., Chao, J. R., Ling, T., & Sabesan, R. (2026) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.64898/2026.05.01.722017
Metadata
- Authors
- Weaver, J. B., Pattison, A. J., McGarry, M. D., Perreard, I. M., Swienckowski, J. G., Eskey, C. J., Lollis, S. S., & Paulsen, K. D.
- Affiliations
- Dartmouth College (US)
- Dartmouth–Hitchcock Medical Center (US)
- Corresponding author
- Weaver, John B
- Venue
- Physics in Medicine and Biology
- Publisher
- IOP Publishing
- Published
- 2012-10-18
- Volume
- 57
- Issue
- 22
- Pages
- 7275-7287
- ISSN
- 0031-9155 (ISSN-L)
- 1361-6560
- PMC
- PMC3797022
- OpenAlex
- W1996982466
- Semantic Scholar
- 52ea9a96ab7cb4286c5ce8f4b1798a8f232ef152
- Cited by
- 102
- References
- 30
- Citations by year
- 2026: 5 · 2025: 8 · 2024: 8 · 2023: 4 · 2022: 3 · 2021: 16 · 2020: 7 · 2019: 7 · 2018: 9 · 2017: 6 · 2016: 7 · 2015: 6 · 2014: 8 · 2013: 8
- FWCI
- 5.31
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Physics in Medicine and Biology (publishedVersion)
- PubMed (publishedVersion)
- Europe PMC (PubMed Central) (acceptedVersion)
- PubMed Central (acceptedVersion)
- Funders
- National Institutes of Health
- National Cancer Institute (P01CA080139)
- National Institute of Diabetes and Digestive and Kidney Diseases (R01DK063013)
- National Institute of Biomedical Imaging and Bioengineering (R01EB004632)
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Ultrasound Imaging and Elastography
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Advanced MRI Techniques and Applications
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Advanced Neuroimaging Techniques and Applications
- Keywords
- magnetic resonance elastography, brain tissue mechanical properties, cardiac pulsation, shear modulus, poroelastic model, linear elastic model, phase-contrast MR angiography, tissue velocity, reproducibility
- MeSH terms
- Biomechanical Phenomena, Blood Vessels, Brain, Magnetic Resonance Imaging, Movement, Imaging, Three-Dimensional, Elasticity Imaging Techniques, Mechanical Phenomena
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review