Proton range verification through prompt gamma-ray spectroscopy
The DOI matches its Crossref record; 9 of the 9 references it cites were found in a registry or an index.
Abstract
We present an experimental study of a novel method to verify the range of proton therapy beams. Differential cross sections were measured for 15 prompt gamma-ray lines from proton-nuclear interactions with (12)C and (16)O at proton energies up to 150 MeV. These cross sections were used to model discrete prompt gamma-ray emissions along proton pencil-beams. By fitting detected prompt gamma-ray counts to these models, we simultaneously determined the beam range and the oxygen and carbon concentration of the irradiated matter. The performance of the method was assessed in two phantoms with different elemental concentrations, using a small scale prototype detector. Based on five pencil-beams with different ranges delivering 5 × 10(8) protons and without prior knowledge of the elemental composition at the measurement point, the absolute range was determined with a standard deviation of 1.0-1.4 mm. Relative range shifts at the same dose level were detected with a standard deviation of 0.3-0.5 mm. The determined oxygen and carbon concentrations also agreed well with the actual values. These results show that quantitative prompt gamma-ray measurements enable knowledge of nuclear reaction cross sections to be used for precise proton range verification in the presence of tissue with an unknown composition.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Robust estimation of a location parameter
Huber, P. J. (1964) · The Annals of Mathematical Statistics · doi:10.1214/aoms/1177703732 · Cited by 7,228
Geant4 developments and applications
Allison, J., Amako, K., Apostolakis, J., Araujo, H., Arce Dubois, P., Asai, M., Barrand, G., Capra, R., Chauvie, S., Chytracek, R., et al. (2006) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.2006.869826 · Cited by 6,946
Relative biological effectiveness (RBE) values for proton beam therapy
Paganetti, H., Niemierko, A., Ancukiewicz, M., Gerweck, L. E., Goitein, M., Loeffler, J. S., & Suit, H. D. (2002) · International Journal of Radiation Oncology*Biology*Physics · doi:10.1016/s0360-3016(02)02754-2 · Cited by 1,364
NESTA: A Fast and Accurate First-Order Method for Sparse Recovery
Becker, S., Bobin, J., & Candès, E. J. (2009) · Cited by 897
In vivoproton range verification: a review
Knopf, A. C., & Lomax, A. (2013) · Physics in Medicine and Biology · doi:10.1088/0031-9155/58/15/r131 · Cited by 620
Parodi, K., Paganetti, H., Shih, H. A., Michaud, S., Loeffler, J. S., DeLaney, T. F., Liebsch, N. J., Munzenrider, J. E., Fischman, A. J., Knopf, A., et al. (2007) · International Journal of Radiation Oncology*Biology*Physics · doi:10.1016/j.ijrobp.2007.01.063 · Cited by 409
Mitigating the risk of radiation-induced cancers: limitations and paradigms in drug development
Yoo, S. S., Jorgensen, T. J., Kennedy, A. R., Boice Jr, J. D., Shapiro, A., Hu, T. C.-C., Moyer, B. R., Grace, M. B., Kelloff, G. J., Fenech, M., et al. (2014) · Journal of Radiological Protection · doi:10.1088/0952-4746/34/2/r25 · Cited by 221
Simulation of prompt gamma-ray emission during proton radiotherapy
Verburg, J., Shih, H. A., & Seco, J. C. (2012) · Physics in Medicine and Biology · doi:10.1088/0031-9155/57/17/5459 · Cited by 111 · cited 2×
“We have previously compared literature data and simulations of cross sections for reactions with 12C, 14N and 16O (Verburg et al 2012).”
BackgroundResultSelf-citationEspaña, S., & Paganetti, H. (2010) · Physics in Medicine and Biology · doi:10.1088/0031-9155/55/24/011 · Cited by 72 · cited 1×
“This leads to uncertainty because of the lack of a one-to-one relation between the computed tomography images used for treatment planning and elemental composition (Parodi et al 2007, Espana and Paganetti 2010).”
Method
Cited by
Nuclear physics in particle therapy: a review
Durante, M., & Paganetti, H. (2016) · Reports on Progress in Physics · doi:10.1088/0034-4885/79/9/096702 · Cited by 345
A full-scale clinical prototype for proton range verification using prompt gamma-ray spectroscopy
Hueso-González, F., Rabe, M., Ruggieri, T. A., Bortfeld, T., & Verburg, J. M. (2018) · Physics in Medicine & Biology · doi:10.1088/1361-6560/aad513 · Cited by 197
Self-citation
Metadata
- Authors
- Verburg, J. M., & Seco, J.
- Venue
- Physics in Medicine and Biology
- Publisher
- IOP Publishing
- Published
- 2014-11-03
- Volume
- 59
- Issue
- 23
- Pages
- 7089-7106
- ISSN
- 0031-9155 (ISSN-L)
- 1361-6560
- OpenAlex
- W2039699511
- Cited by
- 201
- References
- 9
- FWCI
- 14.30
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Proton range verification through prompt gamma-ray spectroscopy. https://app.cochat.ai/papers/10.1088/0031-9155/59/23/7089
- The paper
- Verburg, J. M., & Seco, J. (2014). Proton range verification through prompt gamma-ray spectroscopy. Physics in Medicine and Biology. https://doi.org/10.1088/0031-9155/59/23/7089
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.