A Prospective Multicenter Evaluation of the Accuracy and Safety of an Implanted Continuous Glucose Sensor: The PRECISION Study
Open accessThe DOI matches its Crossref record; 6 of the 6 references it cites were found in a registry or an index.
Abstract
Background: A prior study (PRECISE II) demonstrated that an implantable continuous glucose monitoring (CGM) system (Eversense ® CGM System) provided accurate glucose readings through the 90-day sensor life with a favorable safety profile in participants with type 1 or type 2 diabetes (T1D, T2D). This study was performed to further characterize the accuracy of the system. Methods: PRECISION was a prospective multicenter study that evaluated the accuracy and safety of Eversense among adults with T1D or T2D through 90 days (NCT02647905). Accuracy measures included percentage system agreement and mean absolute relative difference (MARD) between Eversense and Yellow Springs Instrument reference measurements from 40 to 400 mg/dL. The primary safety endpoint was incidence of device-related or sensor insertion/removal procedure-related serious adverse events (SAEs) through 90 days. An updated glucose calculation algorithm was also applied to the sensor data from the PRECISE II study to evaluate consistency of accuracy results. Results: Thirty-five participants received the CGM system. Eighty-five percent of CGM values were within 15/15% of reference and the MARD value against reference was 9.6% (95% confidence interval [CI]: 8.9–10.4). All sensors were functional through day 90. No device- or procedure-related SAEs occurred. Application of the updated algorithm to PRECISE II sensor data resulted in 87% of readings within 15/15% of reference and an MARD value against reference of 8.5% (95% CI: 8.0%–9.1%). Conclusions: PRECISION corroborated prior accuracy and safety findings of the Eversense CGM System through the 90-day sensor life. The updated algorithm improved accuracy of measurements in PRECISE II.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Beck, R. W., Riddlesworth, T., Ruedy, K., Ahmann, A., Bergenstal, R., Haller, S., Kollman, C., Kruger, D., McGill, J. B., Polonsky, W., et al. (2017) · JAMA · doi:10.1001/jama.2016.19975 · Cited by 1,185
Lind, M., Polonsky, W., Hirsch, I. B., Heise, T., Bolinder, J., Dahlqvist, S., Schwarz, E., Ólafsdóttir, A. F., Frid, A., Wedel, H., et al. (2017) · JAMA · doi:10.1001/jama.2016.19976 · Cited by 705
Beck, R. W., Riddlesworth, T. D., Ruedy, K., Ahmann, A., Haller, S., Kruger, D., McGill, J. B., Polonsky, W., Price, D., Aronoff, S., et al. (2017) · Annals of Internal Medicine · doi:10.7326/m16-2855 · Cited by 635
the SWITCH Study Group, Battelino, T., Conget, I., Olsen, B., Schütz-Fuhrmann, I., Hommel, E., Hoogma, R., Schierloh, U., Sulli, N., & Bolinder, J. (2012) · Diabetologia · doi:10.1007/s00125-012-2708-9 · Cited by 494
Ueland, G. Å., Methlie, P., Kellmann, R., Bjørgaas, M., Åsvold, B. O., Thorstensen, K., Kelp, O., Thordarson, H. B., Mellgren, G., Løvås, K., et al. (2017) · European Journal of Endocrinology · doi:10.1530/eje-17-0078 · Cited by 94
Menke, A., Arloth, J., Best, J., Namendorf, C., Gerlach, T., Czamara, D., Lucae, S., Dunlop, B. W., Crowe, T. M., Garlow, S. J., et al. (2016) · Psychoneuroendocrinology · doi:10.1016/j.psyneuen.2016.04.003 · Cited by 35 · cited 1×
“This level is eight times lower than the 1 ng/mL level that is required to have a therapeutic effect, 12,13 providing evidence that the risk of chronic dexamethasone exposure with the Eversense sensor is low.”
Cited by
Opportunities and challenges in the diagnostic utility of dermal interstitial fluid
Friedel, M., Thompson, I. A. P., Kasting, G., Polsky, R., Cunningham, D., Soh, H. T., & Heikenfeld, J. (2023) · Nature Biomedical Engineering · doi:10.1038/s41551-022-00998-9 · Cited by 360
Freckmann, G., Eichenlaub, M., Waldenmaier, D., Pleus, S., Wehrstedt, S., Haug, C., Witthauer, L., Jendle, J., Hinzmann, R., Thomas, A., et al. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231190941 · Cited by 86
“90,96,123,127,131,138,140,155,158,160 However, this cannot be recommended as the CGM value distribution is dependent on CGM accuracy and thus impairs the comparison of testing procedures among studies.”
BackgroundDeiss, D., Irace, C., Carlson, G., Tweden, K. S., & Kaufman, F. R. (2020) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2019.0159 · Cited by 44
Self-citationReal-World Data from the First U.S. Commercial Users of an Implantable Continuous Glucose Sensor
Sanchez, P., Ghosh-Dastidar, S., Tweden, K. S., & Kaufman, F. R. (2019) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2019.0234 · Cited by 40
“10,23 In addition, the wear time of the transmitter indicates that patients value the real-world use of the Eversense CGM system.”
Self-citationA Prospective Multicenter Clinical Performance Evaluation of the C-CGM System
Rebec, M., Cai, K., Dutt-Ballerstadt, R., & Anderson, E. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820964574 · Cited by 11
Metadata
- Authors
- Christiansen, M. P., Klaff, L. J., Bailey, T. S., Brazg, R., Carlson, G., & Tweden, K. S.
- Venue
- Diabetes Technology & Therapeutics
- Publisher
- SAGE Publications
- Published
- 2019-05
- Volume
- 21
- Issue
- 5
- Pages
- 231-237
- ISSN
- 1520-9156 (ISSN-L)
- 1557-8593
- OpenAlex
- W2932140587
- Cited by
- 62
- References
- 6
- FWCI
- 6.37
- Type
- journal article
- Language
- en
- Open access
- hybrid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: A Prospective Multicenter Evaluation of the Accuracy and Safety of an Implanted Continuous Glucose Sensor: The PRECISION Study. https://app.cochat.ai/papers/10.1089/dia.2019.0020
- The paper
- Christiansen, M. P., Klaff, L. J., Bailey, T. S., Brazg, R., Carlson, G., & Tweden, K. S. (2019). A Prospective Multicenter Evaluation of the Accuracy and Safety of an Implanted Continuous Glucose Sensor: The PRECISION Study. Diabetes Technology & Therapeutics. https://doi.org/10.1089/dia.2019.0020
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.