The Host Response to Endotoxin-Contaminated Dermal Matrix
Abstract
Biologic scaffold materials composed of extracellular matrix (ECM) have been shown to promote the formation of site-specific, functional, host tissue following placement in a number of preclinical and clinical studies. Endotoxin contamination of biomaterials is thought to result in deleterious immune responses that may affect the remodeling outcome when present in significant quantities. However, the exact amount of endotoxin contamination within or upon an ECM-based biologic scaffold that is required to elicit adverse effects in recipients is currently unknown. The present study examined the in vitro and in vivo effects of endotoxin contamination within an ECM scaffold derived from porcine dermis upon the host immune response and the downstream ability of the scaffold material to promote constructive tissue remodeling. Test articles with endotoxin values that exceed the current U.S. Food and Drug Administration (FDA) limit had similar or decreased immune responses both in vitro and in vivo when compared with devices that were below the current FDA limit. Dermal matrices spiked with large doses of endotoxin (100 ng/mL), equivalent to 10-20 times the FDA limit, elicited a robust immune response in vitro. However, by 35 days postimplantation, no difference in tissue remodeling was detected, regardless of the amount of endotoxin present within the material. These results suggest that current endotoxin standards may fall well below levels that induce an adverse acute proinflammatory response and associated long-term deleterious effects upon tissue remodeling outcomes.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The FDA recommends specific endotoxin limits depending on intended use, but some commercial and investigational ECM products exceed these thresholds.
“The U.S. Food and Drug Administration (FDA) recommends specific endotoxin limits depending on the intended use of the device [146,147], but studies have shown that some commercial and investigational ECM products exceed these thresholds [148].”
From Immune Response to Extracellular Matrix Bioscaffolds: A Comprehensive Review (2025) · cites this paper as [148]
The citation lists haven’t been fetched yet
References 36 · Cited by 46
- Authors
- Daly, K. A., Liu, S., Agrawal, V., Brown, B. N., Huber, A., Johnson, S. A., Reing, J., Sicari, B., Wolf, M., Zhang, X., & Badylak, S. F.
- Affiliations
- University of Pittsburgh (US)
- Corresponding author
- Badylak, Stephen F.
- ORCID
- Venue
- Tissue Engineering Part A
- Publisher
- SAGE Publications
- Published
- 2012-06
- Volume
- 18
- Issue
- 11-12
- Pages
- 1293-1303
- ISSN
- 1937-3341 (ISSN-L)
- 1937-335X
- OpenAlex
- W2038108644
- Cited by
- 46
- References
- 36
- Citations by year
- 2026: 2 · 2025: 6 · 2024: 3 · 2023: 2 · 2022: 3 · 2021: 1 · 2020: 1 · 2018: 2 · 2017: 3 · 2016: 4 · 2015: 7 · 2014: 4 · 2013: 6 · 2012: 1
- FWCI
- 5.87
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Tissue Engineering Part A (publishedVersion)
- PubMed (publishedVersion)
- Funders
- U.S. Food and Drug Administration
- Topics
- Health Sciences › Medicine › Surgery › Tissue Engineering and Regenerative Medicine
- Physical Sciences › Engineering › Biomedical Engineering › Bone Tissue Engineering Materials
- Health Sciences › Medicine › Rehabilitation › Wound Healing and Treatments
- Keywords
- endotoxin contamination, biological scaffolds, extracellular matrix, porcine dermis, host immune response, tissue remodeling, pro-inflammatory response, biomaterial safety, dermal matrix
- MeSH terms
- Animals, Body Temperature, Cell Count, Cell Line, Endotoxins, Extracellular Matrix, Female, Humans, Monocytes, RNA, Messenger, Staining and Labeling, Tumor Necrosis Factor-alpha, Interleukin-10, Cell Death, Rats, Sprague-Dawley, Chemokine CCL2, Prosthesis Implantation, Implants, Experimental, Dermis, Sus scrofa, Cell Proliferation, Toll-Like Receptor 4, Rats, Chemokine CCL4
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›