CoChat

Allowable imprecision for laboratory tests based on clinical and analytical test outcome criteria

Westgard, J O, Seehafer, J J, Barry, P L · 1994 · Clinical Chemistry
Cited by 65

The DOI matches its Crossref record.

Abstract

The allowable imprecision for laboratory tests has been estimated from criteria based on clinical and analytical test outcome. The analytical outcome criteria studied are the Clinical Laboratory Improvement Amendments (CLIA) criteria for proficiency testing. The clinical outcome criteria are estimates of medically significant changes in test results taken from a study in the literature. The estimates of allowable imprecision were obtained from quality-planning models that relate test outcome criteria to the allowable amount of imprecision and inaccuracy and to the quality control that is necessary to assure achievement of the desired outcome criteria in routine operation. These operating specifications for imprecision are consistently more demanding (require lower CVs) than the medically useful CVs originally recommended in the literature because the latter do not properly consider within-subject biological variation. In comparing estimates of allowable imprecision, the CLIA outcome criteria are more demanding than the clinical outcome criteria for aspartate aminotransferase (asymptomatic patients), cholesterol, creatinine (asymptomatic patients), glucose, thyroxine, total protein, urea nitrogen, hematocrit, and prothrombin time. The clinical outcome criteria are more demanding for bilirubin (acute illness), iron, potassium, urea nitrogen (acute illness), and leukocyte count. The estimates of allowable imprecision from analytical and clinical outcome criteria overlap for aspartate aminotransferase (acute illness), bilirubin (asymptomatic patients), calcium, creatinine (acute illness), sodium, triglyceride, and hemoglobin.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1093/clinchem/40.10.1909

Cites

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

See all 65

Metadata

Authors
Westgard, J. O., Seehafer, J. J., & Barry, P. L.
Venue
Clinical Chemistry
Published
1994-10-01
Volume
40
Issue
10
Pages
1909-1914
ISSN
  • 0009-9147 (ISSN-L)
  • 1530-8561
OpenAlex
W1921194638
Cited by
65
References
0
FWCI
2.35
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Allowable imprecision for laboratory tests based on clinical and analytical test outcome criteria. https://app.cochat.ai/papers/10.1093/clinchem/40.10.1909
The paper
Westgard, J. O., Seehafer, J. J., & Barry, P. L. (1994). Allowable imprecision for laboratory tests based on clinical and analytical test outcome criteria. Clinical Chemistry. https://doi.org/10.1093/clinchem/40.10.1909

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.