Performance of miniature scalar atomic magnetometers for magnetocardiography in an unshielded hospital laboratory for clinical electrophysiology

Open access
Fenici, R, Mashkar, R, Brisinda, D·2020·European Heart Journal

Abstract

Background: Magnetocardiography (MCG) is under clinical evaluation as a novel, sensitive, radiation-free method for first-level non-invasive diagnosis of ischemic heart disease (IHD) and for rule-out of acute coronary syndromes (ACS) in chest pain patients with still normal ECG and serum markers. So far, multichannel MCG has been mainly performed with cryogenic sensors (SQUIDs). Alternatively, to reduce costs and/or to bring MCG to the patient bedside, novel non-cryogenic technologies have been developed, based on small induction coils or optically pumped magnetometers (OPMs). The magnetic field (MF) resolution of OPMs is nowadays comparable to that of the SQUIDs. However, OPMs operate in zero-field mode and consequently need magnetic shielding. Thus, presently used OPMs technology is not suitable for unshielded MCG. On the contrary, Miniature Scalar Atomic Magnetometers (MFAM™) operating in the total-field mode, with a sensitivity better than 2pT/√Hz to about 400Hz, were reported to detect the cardiac MF in an unshielded environment. This study aimed to investigate whether the MFAM, originally developed for geophysics and magnetic anomaly detection, could be suitable for MCG in an unshielded hospital laboratory for clinical electrophysiology (UHEPL). Method: Recordings were performed in the north-south direction, with sensors parallel to sources and tilted 30°. A dual-sensor MFAM module was used: one sensor, placed as close as possible to the source of interest, to detect its MF, and the other, more distant, to detect and subtract in real-time the background electromagnetic noise. The magnetic signal was acquired at 1-kHz sampling frequency (bandwidth DC-250 Hz). Signal averaging (SA) and digital filtering of RF noise (notch 50 Hz, LP 100 Hz) were used to improve the signal to noise ratio (S/Nr). First, the recordings were performed with a simplified phantom, filled with saline, using our amagnetic catheter to simulate current dipoles (CDs) of different geometry. Then MCG was attempted in 15 heathy volunteers (HV) (age 22–73 years) and compared with their SQUID MCG (Figure 1C, D). Results: After SA of 200 beats and filtering average peak-to-peak noise was 2–4 pT. S/N ratio (≥10 in the phantom (with 1 mA CDs) was twice lower in MFAM MCG. However ventricular MCG was feasible in all HV. On the contrary atrial MCG was adequate in one case only. Figure 1 summarizes the performance of the MFAM in UHEPL: (A) typical MFAM noise spectra. (B) 36-point SQUID MCG at the R wave peak, (C) MFAM-MCG signal taken at the fifth right parasternal intercostal space, D) SQUID MCG at same position for comparison. Conclusions: MFAM sensors were sufficiently stable to record MCG of several minutes in our UHEPL. The present sensitivity of MFAM is adequate for clinical MCG of ventricular de/repolarization (i.e. for detection/rule out of IHD/ACS), but not of atrial electrophysiology. A better performance is foreseen with development of more efficient gradiometer technology. Figure 1. Comparison between MFAM and SQUID Funding Acknowledgement: Type of funding source: None

Overview

Integrity signals

Registry

Crossref · 10.1093/ehjci/ehaa946.0386

Cites

Statistics

Recomputing p-values needs the paper’s full text.

Verification

Claims the paper makes

We couldn’t retrieve this paper’s full text, so its claims and self-check weren’t run. The registry and citation checks still apply.

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Citations

The citation lists haven’t been fetched yet

Cited by 7

Metadata

Authors
Fenici, R., Mashkar, R., & Brisinda, D.
Affiliations
  • Biomagnetism and Clinical Physiology International Center (BSSS), Rome, Italy
  • Geometrics, inc,, S, Josè, United States of America
  • Agostino Gemelli University Polyclinic (IT)
Venue
European Heart Journal
Published
2020-11-01
Volume
41
Issue
Supplement_2
ISSN
  • 0195-668X (ISSN-L)
  • 1522-9645
OpenAlex
W3106907842
Cited by
7
References
0
Citations by year
2025: 3 · 2024: 1 · 2023: 1 · 2022: 2
FWCI
0.30
Type
journal article
Language
en
Open access
Locations
European Heart Journal (publishedVersion)
Topics
  • Physical Sciences › Physics and Astronomy › Atomic and Molecular Physics, and Optics › Atomic and Subatomic Physics Research
  • Physical Sciences › Engineering › Electrical and Electronic Engineering › Magnetic Field Sensors Techniques
  • Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
Keywords
magnetocardiography, SQUID comparison, cardiac magnetic field, current dipoles, signal averaging, ischemic heart disease, acute coronary syndrome, clinical electrophysiology, magnetic shielding, signal-to-noise ratio

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review