Comparison of LC-MS-MS and GC-MS Analysis of Benzodiazepine Compounds Included in the Drug Demand Reduction Urinalysis Program
Open accessThe DOI matches its Crossref record; 7 of the 7 references it cites were found in a registry or an index; none of the 3 sources its claims cite is checked for support yet.
Abstract
Liquid chromatography-tandem mass spectrometry (LC-MS-MS) offers specific advantages over gas chromatography-mass spectrometry (GC-MS) such as the ability to identify and measure a broader range of compounds with minimal sample preparation. Comparative analysis of LC-MS-MS versus GC-MS was performed for urinalysis detection of five benzodiazepine compounds currently part of the Department of Defense (DoD) Drug Demand Reduction Program (DDRP) testing panel; alpha-hydroxyalprazolam, oxazepam, lorazepam, nordiazepam and temazepam. In the analyses of internally prepared control urine samples at concentrations around the DDRP administrative decision point for benzodiazepines (100 ng/mL), both technologies produced comparable results with average accuracies between 99.7 and 107.3% and average coefficients of variation (%CV) <9%. Analysis of service member specimens that screened positive for benzodiazepines using both technologies produced comparable results for all analytes. Different degrees of matrix effect were observed for all analytes in the LC-MS-MS analysis. However, the effects were controlled by using deuterated internal standards (ISTDs). Additionally, there was a 39% increase in nordiazepam mean concentration analyzed by LC-MS-MS due to suppression of the ISTD ion by the flurazepam metabolite 2-hydroxyethylflurazepam. The ease and speed of sample extraction, the broader range of compounds that can be analyzed and shorter run time make the LC-MS-MS technology a suitable and expedient alternative confirmation technology for benzodiazepine testing.
Overview
CoChat Summary
TL;DR
LC–MS-MS and GC–MS produced comparable urine benzodiazepine confirmation results for five DDRP panel compounds, with acceptable accuracy and precision around the 100 ng/mL administrative decision point. LC–MS-MS offered easier/faster extraction, shorter run times, and lower LOD/LOQ for some analytes, though a flurazepam metabolite interfered with nordiazepam quantitation.
Key findings
- At concentrations around the DDRP benzodiazepine decision point (100 ng/mL), both methods showed comparable performance: average accuracies between 99.7 and 107.3% and average %CV <9% in internally prepared control urine samples. At 75, 100, and 125 ng/mL, GC–MS average accuracy was 102% with CVs <5%, while LC–MS-MS average accuracy was 105% with CVs <7%; the % difference between methods ranged from 0.14 to 5.07%.
- For LORA, NORD, and TEMA, LC–MS-MS attained lower LOD/LOQ values (1.96–2.15 ng/mL; CVs 3.4–8.1%) than GC–MS (6.13–26.30 ng/mL; CVs 0.4–4.8%). AHAL and OXAZ had equivalent LOD/LOQ values for both technologies.
- LC–MS-MS matrix effects caused target ion enhancement of 2–52% for all analytes except NORD; NORD at 100 ng/mL showed <1% suppression. These effects were controlled with deuterated internal standards, with type C sample accuracies of 98.3–102.5% and CVs <5%.
- The flurazepam metabolite 2-hydroxyethylflurazepam (2HEF) suppressed the NORD ISTD ion by 36% and increased NORD mean concentration by 39% in LC–MS-MS versus GC–MS. TEMA GC–MS analysis failed in 3 of 5 replicates containing club drug mixture, while LC–MS-MS had no such problem. In service member specimens, no significant difference was found between methods for NORD, OXAZ, LORA, and AHAL; TEMA differed significantly (P = 0.025), but the mean percent difference was <3%.
Methods
- Urine samples were analyzed for five benzodiazepines: alpha-hydroxyalprazolam (AHAL), oxazepam (OXAZ), lorazepam (LORA), nordiazepam (NORD), and temazepam (TEMA). Internally prepared controls and sample material were fortified at concentrations including 20, 40, 75, 100, 125, 200, 500, and 1,000 ng/mL; GC–MS used 10 sets and LC–MS-MS used 9 sets of four replicates per target concentration.
- GC–MS used enzymatic hydrolysis, solid-phase extraction, derivatization with MTBSTFA, and selected ion monitoring on Agilent 7890 GC/5975 MS instruments. LC–MS-MS used enzymatic hydrolysis, solid-phase extraction without derivatization, and multiple reaction monitoring on a Waters ACQUITY LC coupled to a Quattro Micro triple quadrupole mass spectrometer.
- Method comparison included LOD/LOQ/ULOL determination, matrix effect assessment by the Matuszewski method, interference testing with 40 ng/mL target analytes plus 5,000 ng/mL interferants, and side-by-side analysis of anonymized service member urine specimens that screened positive for benzodiazepines by immunoassay (Nov 2012–Jul 2013). Statistical analyses used Excel 2010 and Minitab 17, including regression ANOVA and Student’s t-tests.
Limitations
- The paper does not include a formal limitations section.
- Evident from the text: the NORD interference by 2HEF occurred at 5,000 ng/mL, which the authors note was 125 times greater than the sample concentration and higher than literature suggests would be physiologically realistic; the authors still recommend considering this effect during method development if the same parent and transition ions are chosen.
- Evident from the text: TEMA GC–MS analysis failed in 3 of 5 club-drug-mixture replicates, and TEMA results differed significantly between technologies (P = 0.025), although the mean percent difference was <3%. Some service member specimen comparisons had small sample sizes (LORA n = 7; AHAL n = 8).
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
0 Valid · 0 Not valid · 2 Not checked · 40 Own results and interpretations
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Analysis of service member specimens that screened positive for benzodiazepines using both technologies produced comparable results for all analytes.”
- Also stated in: § Conclusion · plain
“Analysis of service member specimens that screened positive for benzodiazepines using both technologies produced comparable results for all analytes.”
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Also stated in: § Conclusion · plain
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Different degrees of matrix effect were observed for all analytes in the LC–MS-MS analysis.”
- Also stated in: § Conclusion · plain
“Different degrees of ME were observed for all analytes in the LC–MS-MS analysis.”
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Also stated in: § Conclusion · plain
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“For AHAL and OXAZ, the LOD and LOQ values were equivalent for both technologies.”
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“In contrast, TEMA results differed significantly between technologies (P = 0.025), however the mean percent difference between both technologies was found to be <3% (data analysis not shown).”
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“GC–MS and LC–MS-MS measured mean concentrations for LORA and NORD differed more than 10% when 2HEF was fortified into 40 ng/mL of these analytes.”
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
26 more claims
Create a free account to see all 42 claims and to verify the 1 reference still open.
- It may state two separate claims as one. 11
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ Materials and methods
“A total of ten independent batches of four replicates of sample material for each analyte concentration were analyzed by GC–MS and LC–MS-MS.”
§ Results and discussion
“Independent sets of samples containing four replicates of the target concentrations: 20, 40, 75, 100, 125, 200, 500 and 1,000 ng/mL were analyzed by GC–MS (10 sets) and LC–MS-MS (9 sets)”
Citations
Cites
Matuszewski, B. K., Constanzer, M. L., & Chavez-Eng, C. M. (2003) · Analytical Chemistry · doi:10.1021/ac020361s · Cited by 5,335
Polydrug abuse: a review of opioid and benzodiazepine combination use
Jones, J. D., Mogali, S., & Comer, S. D. (2012) · Drug and Alcohol Dependence · doi:10.1016/j.drugalcdep.2012.07.004 · Cited by 703 · cited 1×
“Additionally, benzodiazepines are often used in combination with illicit drugs to enhance their effects making them a target for drug abuse monitoring programs (3, 4).”
BackgroundPrinciples of Forensic Toxicology
(2020) · Springer eBooks · doi:10.1007/978-3-030-42917-1 · Cited by 385
Benzodiazepine use among injecting drug users: problems and implications
Darke, S. G. (1994) · Addiction · doi:10.1111/j.1360-0443.1994.tb00909.x · Cited by 51 · cited 1×
“Additionally, benzodiazepines are often used in combination with illicit drugs to enhance their effects making them a target for drug abuse monitoring programs (3, 4).”
BackgroundStout, P. R., Bynum, N. D., Mitchell, J. M., Baylor, M. R., & Ropero-Miller, J. D. (2009) · Journal of Analytical Toxicology · doi:10.1093/jat/33.8.398 · Cited by 29 · cited 2×
“in human urine (1, 5), there are currently no studies that compare the performance of these two analytical methods in the detection and analysis of benzodiazepines.”
BackgroundMethodStout, P. R., Bynum, N. D., Lewallen, C. M., Mitchell, J. M., Baylor, M. R., & Ropero‐Miller, J. D. (2010) · Journal of Analytical Toxicology · doi:10.1093/jat/34.8.430 · Cited by 17 · cited 1×
“in human urine (1, 5), there are currently no studies that compare the performance of these two analytical methods in the detection and analysis of benzodiazepines.”
BackgroundConcentrations of phenobarbital, flurazepam, and flurazepam metabolites in autopsy cases
Ferrara, S., Tedeschi, L., Marigo, M., & Castagna, F. (1979) · Journal of Forensic Sciences · doi:10.1520/jfs10790j · Cited by 9
Cited by
Analytical techniques for the determination of acetaminophen: A review
Montaseri, H., & Forbes, P. B. C. (2018) · TrAC Trends in Analytical Chemistry · doi:10.1016/j.trac.2018.08.023 · Cited by 138
Hamd, W. S., Daher, E. A., Tofa, T. S., & Dutta, J. (2022) · Frontiers in Marine Science · doi:10.3389/fmars.2022.885614 · Cited by 124
Abdelrahman, M., Burritt, D. J., & Tran, L. P. (2017) · Seminars in Cell and Developmental Biology · doi:10.1016/j.semcdb.2017.06.020 · Cited by 114
Bridging Targeted and Untargeted Mass Spectrometry-Based Metabolomics via Hybrid Approaches
Chen, L., Zhong, F., & Zhu, J. (2020) · Metabolites · doi:10.3390/metabo10090348 · Cited by 113
“Meanwhile, UHPLC/HPLC-MS gained popularity in hybrid approaches because of its quicker and less extensive extraction procedures and ability to identify and measure a broader range of compounds [49,50].”
BackgroundKumar, A., Green, K. M., & Rawat, M. (2024) · Foods · doi:10.3390/foods13182937 · Cited by 119
Micro-QuEChERS extraction coupled to GC–MS for a fast determination of Bisphenol A in human urine
Correia‐Sá, L., Norberto, S., Delerue-Matos, C. M., Calhau, C., & Domingues, V. F. (2017) · Journal of Chromatography B · doi:10.1016/j.jchromb.2017.10.060 · Cited by 68
Analytical Methods Used for the Detection and Quantification of Benzodiazepines
Qriouet, Z., Qmichou, Z., Bouchoutrouch, N., Mahi, H., Cherrah, Y., & Sefrioui, H. (2019) · Journal of Analytical Methods in Chemistry · doi:10.1155/2019/2035492 · Cited by 76
“-e search for the identity of the specific ions obtained is performed by comparison with a given reference library [10, 13, 15, 18, 60].”
BackgroundMethodA Facile Profiling Method of Short Chain Fatty Acids Using Liquid Chromatography-Mass Spectrometry
Song, H. E., Lee, H. Y., Kim, S. J., Back, S. H., & Yoo, H. J. (2019) · Metabolites · doi:10.3390/metabo9090173 · Cited by 65
“Liquid chromatography-mass spectrometry (LC-MS) has been often used in metabolomics studies with minimal sample preparation as compared with GC or GC-MS [9,10].”
Farawahida, A. H., Palmer, J. S., & Flint, S. (2022) · International Journal of Food Microbiology · doi:10.1016/j.ijfoodmicro.2022.109829 · Cited by 65
Mass Spectrometry-based Metabolomics in Translational Research
Kim, S. J., Song, H. E., Lee, H. Y., & Yoo, H. J. (2021) · Advances in experimental medicine and biology · doi:10.1007/978-981-33-6064-8_19 · Cited by 62
Samiec, P., Švorc, Ĺ., Stanković, D. M., Vojs, M., Marton, M., & Navrátilová, Z. (2017) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2017.02.023 · Cited by 51
“Many analytical methods have been recently developed for determination of BDZs, in particular, high performance liquid chromatography (HPLC) coupled with diode array detector (DAD) [5, 6] and tandem mass spectrometry (MS/MS) [7, 8], gas chromatography (GC) in combination with MS [9], electrophoresis [10] and spectrofl…”
MethodLee, B. C. Y., Lim, F. Y., Loh, W. H., Ong, S. L., & Hu, J. (2021) · Water · doi:10.3390/w13172340 · Cited by 54
Metabolomics of asthma, COPD, and asthma-COPD overlap: an overview
Dasgupta, S., Ghosh, N., Bhattacharyya, P., Chowdhury, S. R., & Chaudhury, K. (2022) · Critical Reviews in Clinical Laboratory Sciences · doi:10.1080/10408363.2022.2140329 · Cited by 56
Earwax as an alternative specimen for forensic analysis
Shokry, E., Júnior, J. G. M., Ragazzo, P. C., Pereira, N. Z., & Filho, N. R. A. (2017) · Forensic Toxicology · doi:10.1007/s11419-017-0363-z · Cited by 41
Pretorius, C. J., Tugizimana, F., Steenkamp, P. A., Piater, L. A., & Dubery, I. A. (2021) · Metabolites · doi:10.3390/metabo11030165 · Cited by 38
Liu, J., Wurita, A., Xue-jun, W., Hongkun, Y., Gu, J., & Chen, L. (2020) · Legal Medicine · doi:10.1016/j.legalmed.2020.101822 · Cited by 31
Fiorentin, T. R., D’Avila, F. B., Comiran, E., Zamboni, A., Scherer, J. N., Pechansky, F., Mayorga, P., Fröehlich, P. E., & Limberger, R. P. (2017) · Journal of Pharmacological and Toxicological Methods · doi:10.1016/j.vascn.2017.04.003 · Cited by 39
Verhoeven, M., Bonetti, J. L., Kranenburg, R. F., et al. (2023) · TrAC Trends in Analytical Chemistry · doi:10.1016/j.trac.2023.117157 · Cited by 31
Sarp, G., & Yılmaz, E. (2022) · ACS Omega · doi:10.1021/acsomega.2c01311 · Cited by 31
Nicolaou, A. G., Christodoulou, M. C., Stavrou, I. J., & Kapnissi‐Christodoulou, C. P. (2021) · Journal of Chromatography A · doi:10.1016/j.chroma.2021.462277 · Cited by 29
Tarmizi, A. H. A., & Kuntom, A. (2021) · Critical Reviews in Food Science and Nutrition · doi:10.1080/10408398.2020.1865264 · Cited by 24
“In addition, LC-MS-MS exclude sample volatilization stage which further minimizes chemical breakdown and tendency to develop new compounds at high temperature (Perez et al. 2016).”
BackgroundReinvestigation of Ferrate(VI) Oxidation of Bisphenol A over a Wide pH Range
Wang, D., Liang, J., Zhang, H., & Zhang, J. (2021) · ACS ES&T Water · doi:10.1021/acsestwater.1c00302 · Cited by 22
Recent advances in mass spectrometry for the detection of doping
Thomas, A., & Thevis, M. (2024) · Expert Review of Proteomics · doi:10.1080/14789450.2024.2305432 · Cited by 26
Development of a New LC-MS/MS Screening Method for Detection of 120 NPS and 43 Drugs in Blood
Vaiano, F., Bertol, E., Mineo, M., Pietrosemoli, L., Rubicondo, J., Supuran, C. T., & Carta, F. (2021) · Separations · doi:10.3390/separations8110221 · Cited by 27
Brownstein, K. J., Tushingham, S., Damitio, W. J., et al. (2020) · Frontiers in Molecular Biosciences · doi:10.3389/fmolb.2020.00133 · Cited by 22
Melo, W. O. D. S., Júnior, E. C. F., Freire, J. C. P., Costa, B. P., Lira, A. B., Freires, I. A., Cavalcanti, Y. W., Lopes, W. S., Tavares, J. F., Pêssoa, H. D. L. F., et al. (2021) · Archives of Oral Biology · doi:10.1016/j.archoralbio.2020.104994 · Cited by 29
Almofti, N., Ballesteros‐Gómez, A., Rubio, S., & Girela-López, E. (2022) · Talanta · doi:10.1016/j.talanta.2022.123713 · Cited by 22
An, J., Dong, Z., Zhang, W., Yan, Y., Kang, W., & Lian, K. (2021) · Microchemical Journal · doi:10.1016/j.microc.2021.106475 · Cited by 20
Dixit, D., Reddy, C. R. K., Trivedi, M. H., & Gadhavi, D. K. (2020) · Separation Science Plus · doi:10.1002/sscp.201900109 · Cited by 17
Use of micro, capillary, and nano liquid chromatography for forensic analysis
Langford, J. B., & Lurie, I. S. (2021) · Journal of Separation Science · doi:10.1002/jssc.202100631 · Cited by 19
Analysis of triterpenes and triterpenoids
Garg, A., Sharma, R., Dey, P. K., Kundu, A. B., Kim, H. S., Bhakta, T., & Kumar, A. (2020) · Elsevier eBooks · doi:10.1016/b978-0-12-816455-6.00011-1 · Cited by 28
Adeyeye, S. A. O., Sivapriya, T., & Sankarganesh, P. (2025) · Discover Food · doi:10.1007/s44187-025-00391-w · Cited by 16
“However, GC–MS has two main drawbacks: first, electron ionization (EI) mass spectra often reveal weak or missing molecular ions; second, it can only study a very limited range of volatile, thermally stable compounds [120].”
Yang, S., Zhao, Q., Wang, D., Zhang, T., Zhong, Z., Kwok, L., Bai, M., & Sun, Z.-H. (2024) · Journal of Dairy Science · doi:10.3168/jds.2024-25217 · Cited by 17
Medina, R., Franco, M. E. E., Cabral, L. D. C., et al. (2020) · Australasian Plant Pathology · doi:10.1007/s13313-020-00753-1 · Cited by 14
Aamir, M., Irum, S., Siddique, A., et al. (2021) · Analytical Biochemistry · doi:10.1016/j.ab.2021.114496 · Cited by 14
Systematic Evaluation and Validation of Benzodiazepines Confirmation Assay Using LC–MS-MS
Kang, M.-G., & Lin, H. (2018) · Journal of Analytical Toxicology · doi:10.1093/jat/bky071 · Cited by 14
Analysis of Benzodiazepines for Drug-Facilitated Assaults and Abuse Settings (Urine)
Drummer, O. H., Rago, M. D., & Gerostamoulos, D. (2018) · Methods in molecular biology · doi:10.1007/978-1-4939-8823-5_3 · Cited by 14
He, X., Wang, J., Teng, X., Fan, L., Li, X., Wei, C., & Zhang, C. (2020) · Microchemical Journal · doi:10.1016/j.microc.2020.105853 · Cited by 12
Kuwayama, K., Miyaguchi, H., Kanamori, T., et al. (2020) · Forensic Toxicology · doi:10.1007/s11419-020-00553-8 · Cited by 13
“Multi-target SRM using a tandem quadrupole LC–MS/MS instrument is used as a general drug screening method in most laboratories [1, 4, 5, 24, 25].”
MethodFormation and identification of novel derivatives of primary amine and zwitterionic drugs
Ash, J., Hickey, L. D., & Goodpaster, J. V. (2018) · Forensic Chemistry · doi:10.1016/j.forc.2018.07.004 · Cited by 12
“Lorazepam has been analyzed in urine utilizing both MTBSTFA derivatization [23] and a dual derivatization procedure with MTBSTFA and TFAA [24].”
MethodKim, H., Sung, D., Yu, H., et al. (2021) · Toxics · doi:10.3390/toxics9100230 · Cited by 12
“In the case of pharmaceuticals and environmental samples, some studies have compared GC and LC methods simultaneously [19,25,26].”
BackgroundBioinformatic Analysis of Metabolomic Data: From Raw Spectra to Biological Insight
Santamaria, G., & Pinto, F. R. (2024) · BioChem · doi:10.3390/biochem4020005 · Cited by 18
Choucair, I., Shang, E., Tran, M. N., Cassella, G., & El-Khoury, J. M. (2024) · Clinica Chimica Acta · doi:10.1016/j.cca.2024.119939 · Cited by 13
Adewuyi, A., & Li, Q. (2024) · Journal of Hazardous Materials Advances · doi:10.1016/j.hazadv.2024.100465 · Cited by 25
Yen, Y., Chang, Y., Lai, P.-J., et al. (2020) · Molecules · doi:10.3390/molecules25040972 · Cited by 11
“The results revealed biases in the range of −4% to 9% within ±20% of the target concentration [39].”
BackgroundHösli, R., König, S., & Mühlebach, S. (2018) · Journal of Analytical Methods in Chemistry · doi:10.1155/2018/8274131 · Cited by 9
“While the GC-MS has long been the standard method in forensic testing, LC-MS/MS methods have become more common, as they generally offer some advantages over GC-MS [9].”
MethodChemical Profiles of Terminalia catappa LINN Nut and Terminalia subspathulata KING Fruit
Yakubu, Y., Lee, S. Y., & Shaari, K. (2021) · Pertanika journal of tropical agricultural science · doi:10.47836/pjtas.44.4.06 · Cited by 14
“Generally, LC-MS allows the analysis of compounds with a broader range of polarity and minimal sample preparation (Perez et al., 2016).”
Caballero‐Casero, N., Mihretu, L. D., & Rubio, S. (2021) · Journal of Analytical Toxicology · doi:10.1093/jat/bkab023 · Cited by 10
Droplet probe: A non-destructive residue analysis of Wari ceramics from the imperial heartland
Cank, K. B., Henkin, J. M., Cook, A. G., et al. (2021) · Journal of Archaeological Science · doi:10.1016/j.jas.2021.105468 · Cited by 9
Steuer, C., & Raeber, J. (2025) · European Journal of Pharmaceutical Sciences · doi:10.1016/j.ejps.2025.107136 · Cited by 10
Using targeted metabolomics to elucidate the indole auxin network in plants
Cohen, J. D., Tang, Q., & Hegeman, A. D. (2022) · Methods in enzymology on CD-ROM/Methods in enzymology · doi:10.1016/bs.mie.2022.07.038 · Cited by 9
Human hair as a diagnostic tool in medicine
Florou, V. A., Manaprasertsak, A., Slyusarenko, M., Amend, S. R., Kazi, J. U., Hammarlund, E. U., & Pienta, K. J. (2025) · Biochemistry and Biophysics Reports · doi:10.1016/j.bbrep.2025.102129 · Cited by 19
“In addition, LC-MS is superior in analyzing a broader range of metabolites and offers an easier and faster sample preparation compared to GC-MS [146].”
Lock-in Amplifier Dairy Sensor for Detection of Ciprofloxacin
Devasagayam, J., Leclerc, C. A., Bosma, R., et al. (2023) · IEEE Access · doi:10.1109/access.2023.3270136 · Cited by 7
“The GC-MS technique can achieve a similar LOD [55],”
MethodTargeted Metabolomics Using LC‐MS in Neurospora crassa
Carrillo, A. J., Halilovic, L., Hur, M., Kirkwood, J. S., & Borkovich, K. A. (2022) · Current Protocols · doi:10.1002/cpz1.454 · Cited by 7
Chemical sensor technology for detection of 3-monochloropropane-1,2-diol in foodstuffs
Saputra, H. A., Andreas, A., Karim, M. M., et al. (2025) · Critical Reviews in Analytical Chemistry · doi:10.1080/10408347.2025.2565254 · Cited by 6
Chen, S., Han, Y., Zhang, S., et al. (2024) · Heliyon · doi:10.1016/j.heliyon.2024.e27897 · Cited by 6
“…the LC-MS/MS method does not require sample volatilization, this not only reduces sample consumption but also prevents the loss of the analytical target or the generation of new substances during the complex pretreatment process [25,26], thereby minimizing interference with the analysis.”
BackgroundNarcotic Analgesics and Common Drugs of Abuse
Bluth, M. H., & Pincus, M. R. (2016) · Clinics in Laboratory Medicine · doi:10.1016/j.cll.2016.07.013 · Cited by 10
Fagiola, M. (2019) · Legal Medicine · doi:10.1016/j.legalmed.2019.02.004 · Cited by 6
Kim, S. Y., Shin, D. W., Hyun, J., et al. (2023) · Molecules · doi:10.3390/molecules28196803 · Cited by 5
“ADHD diagnoses have recently increased, along with the use of prescription psychostimulants, particularly prescription ADHD medications [3,4].”
BackgroundA review on advancement of biosensors for benzodiazepines detection
Mehta, Y., Suleman, S., Anzar, N., Sharma, R., & Narang, J. (2023) · International Journal of Environmental & Analytical Chemistry · doi:10.1080/03067319.2023.2261008 · Cited by 5
Zhao, X., Zhang, Y., Wang, Q., Huang, J., She, B., Shi, Q., Xiang, X., Su, C., Zhang, L., Huang, R., et al. (2021) · Phytochemical Analysis · doi:10.1002/pca.3054 · Cited by 5
Moini, M. (2018) · Methods in molecular biology · doi:10.1007/978-1-4939-8579-1_4 · Cited by 5
Haššo, M., Kekeľáková, A., Hanko, M., et al. (2024) · Journal of The Electrochemical Society · doi:10.1149/1945-7111/ad3c95 · Cited by 4
Kamal, H. J., Gandhi, V., Akil, L., et al. (2025) · Molecules · doi:10.3390/molecules30173451 · Cited by 4
New Trends in the Methodologies of Determination of Benzodiazepine Residues in Biological Samples
Ibarra, I. S., Vázquez-Garrido, I., Islas, G., & Flores-Aguilar, J. F. (2025) · Separations · doi:10.3390/separations12040095 · Cited by 4
“This technique provides accurate and precise results while minimizing the need for sample preparation [34].”
Zhang, L., Si, X., Yan, X., et al. (2020) · Current Analytical Chemistry · doi:10.2174/1573411016999200414145325 · Cited by 4
Marley, M. B., Brownstein, K. J., Lepper, B. T., Tushingham, S., Leone, K. L., Rafferty, S. M., Gang, D. R., Pansing, L. L., Pickard, W. H., & Six, J. (2018) · Journal of Archaeological Science Reports · doi:10.1016/j.jasrep.2018.03.008 · Cited by 4
Analytical Methods used for the detection and quantification of benzodiazepinesPreprint
Qriouet, Z. (2019) · OSF Preprints (OSF Preprints) · doi:10.31219/osf.io/4d8fw · Cited by 6
Coleman, J. L., Wang, S. Y., Marek, P. E., Morrison, C. R., Sedio, B. E., & Cannatella, D. C. (2025) · Journal of Chemical Ecology · doi:10.1007/s10886-025-01584-4 · Cited by 3
“…data we report for S. flotator was collected using a type of liquid-chromatography mass-spectrometry (LC–MS) instrument, which tend to have lower detection limits (i.e., higher sensitivities to detect more compounds) compared to GC–MS instruments (Perez et al. 2016; Topolewska and Haliński 2024).”
Singh, T., Sharma, U., Kumar, R., & Agrawal, V. (2025) · In Vitro Cellular & Developmental Biology - Plant · doi:10.1007/s11627-025-10526-4 · Cited by 3
“This technique can identify a wide dynamic range of biochemicals having extreme molecular complexity, with lower detection limit and provide reproducible quantitative data (Lu et al 2008; Perez et al 2016).”
State-of-the-art methods for quantifying microbial polyhydroxyalkanoates
Conners, E. M., & Bose, A. (2025) · Applied and Environmental Microbiology · doi:10.1128/aem.00274-25 · Cited by 2
“Basic protocols require spectrofluorome-ters and/or fluorescent microplate readers to measure the Ex/Em spectra of stained cells (69).”
Zheng, Y. Z., Bunch, D. R., Lembright, K., & Wang, S. (2018) · The Journal of Applied Laboratory Medicine · doi:10.1373/jalm.2018.026658 · Cited by 2
Toxicologia forense: o estudo dos agentes tóxicos nas ciências forenses
Pritsch, I. C. (2020) · Revista Criminalistica e Medicina Legal · doi:10.51147/rcml039.2020 · Cited by 5
Christopher, F., Ndakidemi, P. A., Nyalala, S., & Mbega, E. R. (2023) · Journal of Natural Pesticide Research · doi:10.1016/j.napere.2023.100057 · Cited by 7
Ahmed, M. D., Maraz, K. M., & Mazumder, S. (2023) · Materials Engineering Research · doi:10.25082/mer.2023.01.005 · Cited by 2
“The technique’s capacity to analyze a broad range of compounds, from small molecules to complex biomolecules, is relevant in diverse applications within these fields [85,86].”
BackgroundElectrochemical aptasensor coated with Zno@Au bi-nanostructured for the detection of alprazolam
Mehta, Y., Sharma, R., Suleman, S., et al. (2025) · International Journal of Environmental & Analytical Chemistry · doi:10.1080/03067319.2025.2501190 · Cited by 1
Lin, C., Yoshinaga, K., Imasaka, T., et al. (2025) · Microchemical Journal · doi:10.1016/j.microc.2025.114356 · Cited by 1
Msobo, A., Tsotetsi, T. N., Koorsen, G., Piater, L. A., & Mhlongo, M. I. (2025) · South African Journal of Botany · doi:10.1016/j.sajb.2025.08.041 · Cited by 1
Houam, A., Zeghib, A., Tamfu, A. N., Anouar, E. H., Zeghib, F., Yılmaz, M. A., Ceylan, Ö., Orhan, İ. E., Cakir, O., Tarhan, A., et al. (2026) · Chemistry & Biodiversity · doi:10.1002/cbdv.202503675 · Cited by 2
Yousuf, S. M., Alater, A. M., Alabdulla, M., et al. (2023) · Case Reports in Psychiatry · doi:10.1155/2023/4370648 · Cited by 1
“Furthermore, it is recommended that faster and more resolving techniques in-lieu of SUD be made available at testing centers, since not all rehabilitation labs have super confirmatory tests like GC-MS or LC-MS-MS ([6]).”
Mfengwana, P.-M.-A. H., Manduna, I. T., & Mashele, S. S. (2023) · Indian Journal of Natural Products and Resources · doi:10.56042/ijnpr.v14i3.5405 · Cited by 1
Zhou, C., Cai, X., Zhang, Y., Pan, C., Wang, C., Zhai, S., Yang, L., Zhu, S., & Zhang, X. (2020) · Journal of Analytical Science & Technology · doi:10.1186/s40543-020-00222-w · Cited by 1
“compared LC-MS/MS and GC/MS method (Perez et al. 2016) to determine benzodiazepines in urine with solid-phase extraction (SPE).”
BackgroundMethodInfluentialZhao, Y., Araújo, M., Flynn, T. J., et al. (2021) · Food Additives & Contaminants Part A · doi:10.1080/19440049.2021.1873426 · Cited by 1
“Compared to GC-MS, LC-MS offers obvious advantages of ease and speed of sample preparation, shorter run time, and a potentially broader range of compounds that could be detected (Perez et al. 2016).”
BackgroundIntroduction to Bioanalytical Mass Spectrometry
Martins, D. N., Vendramini, P. H., Simionato, A. V. C., & Sussulini, A. (2021) · Springer eBooks · doi:10.1007/978-3-030-82381-8_23 · Cited by 2
Evolution of Multiplexing Technology for High‐Throughput LC/MS Analyses
Latawiec, A. (2023) · Wiley eBooks · doi:10.1002/9781119678496.ch3 · Cited by 1
Letter to the Editor Concerning the Use of Internal Standards: Are We Now Relying on It Too Much?
Tachibana, K. (2019) · Journal of Chromatographic Science · doi:10.1093/chromsci/bmz049
Yen, Y., Chang, Y., Lai, P.-J., et al. (2020) · Forensic Science International · doi:10.1016/j.forsciint.2020.110172
RetractedCNS depressants: benzodiazepines and barbiturates
Snozek, C. L. H. (2020) · Elsevier eBooks · doi:10.1016/b978-0-12-815846-3.00013-2 · Cited by 2
Binanto, I., Warnars, H. L. H. S., Sianipar, N. F., et al. (2021) · Universitas Sanata Dharma Repository (Universitas Sanata Dharma)
Brewers’ spent grains as potential novel functional food ingredients for food security
Tan, Y. X. (2020) · DR-NTU (Nanyang Technological University) · doi:10.32657/10356/146367 · Cited by 2
Mendes, F. D. A. (2021) · Digital Library of Theses and Dissertations (Universidade de São Paulo) · doi:10.11606/d.9.2021.tde-04022022-155512 · Cited by 1
Sphingolipids: A Roadmap from Biomarker Identification to Clinical Application
Bhadwal, P., & Agnihotri, N. (2022) · Springer eBooks · doi:10.1007/978-981-16-4345-3_27
การตรวจสอบความใชไดของวธการวเคราะหคดกรองสารเสพตดในนำวนลกตาดวยเทคนคการสกด QuEChERS และตรวจดวย LC-MS/MS
บุญเฉลียว, บ., & ตันศรีสวัสดิ์, ณ. (2019) · doi:10.58837/chula.the.2019.1079
An Overview of the Inborn Errors of Metabolism, Its Diagnosis, and Management
Saha, A., Aich, S., Mukherjee, K., Sur, T. K., Ghosh, S., & Vashisth, S. (2023) · Disease and Diagnosis · doi:10.34172/ddj.2023.501 · Cited by 2
“GC-MS and LC-MS are well suited to separate isomers, and detect and quantify the compounds compared to NMR or direct MS (47).”
Improving of anticancer compound identification model for medicinal plant’s LCMS data
Binanto, I. (2024) · AIP conference proceedings · doi:10.1063/5.0201202
Han, J. M., Jung, S. A., Song, E. H., Kang, S.-W., Yoo, H. H., Yu, J., & Kwak, Y. B. (2025) · Drug Targets and Therapeutics · doi:10.58502/dtt.25.0004 · Cited by 1
Shoffler, K. M., Barajas, D. A., Tomedi, M. R., et al. (2025) · Journal of Analytical Toxicology · doi:10.1093/jat/bkaf090
The stability of novel synthetic cannabinoids in blood samples in different storage conditions
Akbaba, M., ÖZDEMİR, O., & Isır, A. B. (2025) · Legal Medicine · doi:10.1016/j.legalmed.2025.102762 · Cited by 2
Workplace Drug Testing: SAMHSA and Non-SAMHSA Drugs
Dasgupta, A. (2026) · Springer eBooks · doi:10.1007/978-3-032-15015-8_3
Metadata
- Authors
- Perez, E. R., Knapp, J. A., Horn, C. K., Stillman, S. L., Evans, J. E., & Arfsten, D. P.
- Affiliations
- United States Department of the Navy (US)
- Corresponding author
- Arfsten, Darryl P.
- Venue
- Journal of Analytical Toxicology
- Publisher
- Oxford University Press (OUP)
- Published
- 2016-01-10
- Volume
- 40
- Issue
- 3
- Pages
- 201-207
- ISSN
- 0146-4760 (ISSN-L)
- 1945-2403
- PMC
- PMC4885916
- OpenAlex
- W2299389374
- Cited by
- 106
- References
- 7
- Citations by year
- 2026: 8 · 2025: 15 · 2024: 9 · 2023: 11 · 2022: 8 · 2021: 20 · 2020: 15 · 2019: 6 · 2018: 8 · 2017: 5 · 2016: 1
- FWCI
- 5.44
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Licence
- public-domain
- Locations
- Journal of Analytical Toxicology (publishedVersion, public-domain)
- PubMed Central (publishedVersion, public-domain)
- PubMed (publishedVersion)
- HighWire Press Open Archive (submittedVersion)
- Topics
- Life Sciences › Pharmacology, Toxicology and Pharmaceutics › Toxicology › Forensic Toxicology and Drug Analysis
- Physical Sciences › Chemistry › Analytical Chemistry › Analytical Methods in Pharmaceuticals
- Life Sciences › Agricultural and Biological Sciences › Food Science › Pesticide Residue Analysis and Safety
- Keywords
- benzodiazepines, LC-MS/MS, gas chromatography-mass spectrometry, urine analysis, oxazepam, lorazepam, nordiazepam, temazepam, matrix effects, deuterated internal standards, confirmatory drug testing
- MeSH terms
- Benzodiazepines, Chromatography, Liquid, Humans, Gas Chromatography-Mass Spectrometry, Tandem Mass Spectrometry, Limit of Detection
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.