Wear of Gamma-Crosslinked Polyethylene Acetabular Cups Against Roughened Femoral Balls
Abstract
Crosslinking of ultrahigh molecular weight polyethylene has been shown to markedly improve its wear resistance in clinical studies and laboratory tests using hip joint simulators. However, because most of the laboratory studies have been done under clean conditions using prosthesis-quality, highly polished counterfaces, there is concern regarding how well an intentionally crosslinked polyethylene acetabular cup will resist abrasion by a femoral ball that has been damaged by third-body abrasion in vivo. To investigate this, conventional and radiation crosslinked-remelted acetabular cups of ultra-high molecular weight polyethylene were tested in a hip joint simulator bearing against smooth femoral balls and against balls with moderate and severe roughening. Cups were tested with and without aging to accelerate any oxidative degradation. The crosslinked cups were produced by exposing extruded GUR 4150 bar stock of ultrahigh molecular weight polyethylene to 5 Mrad gamma radiation under a partial vacuum and then the bars were remelted to extinguish residual free radicals. Artificial aging at 70 degrees C under 5 atm oxygen for 14 days induced negligible oxidation in the crosslinked and remelted material. Against smooth balls, the wear of the crosslinked cups, with or without aging, averaged approximately 15% of that of the conventional cups. Against the moderately rough balls, the wear rate of the conventional cups was unchanged, whereas the wear rate increased slightly for the nonaged and aged crosslinked cups, but was still only 26% and 20% of that of the conventional cups, respectively. Against extremely rough balls, the mean wear rates increased markedly for each material such that during the final 1 million cycle interval, the average wear rates of the nonaged and the aged crosslinked cups were 72% and 47% of that of the conventional cups, respectively. That is, the crosslinked polyethylene showed substantially better wear resistance than conventional polyethylene across the range of ball roughnesses, with or without accelerated aging.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Moro, T., Takatori, Y., Ishihara, K., Konno, T., Takigawa, Y., Matsushita, T., Chung, U.-I., Nakamura, K., & Kawaguchi, H. (2004) · Nature Materials · doi:10.1038/nmat1233 · Cited by 572
Mano, J. F., Sousa, R. A., Boesel, L. F., Neves, N. M., & Reis, R. L. (2004) · Composites Science and Technology · doi:10.1016/j.compscitech.2003.09.001 · Cited by 450
Ultra high molecular weight polyethylene: Mechanics, morphology, and clinical behavior
Sobieraj, M. C., & Rimnac, C. M. (2009) · Journal of the Mechanical Behavior of Biomedical Materials · doi:10.1016/j.jmbbm.2008.12.006 · Cited by 289
McKELLOP, H., SHEN, F.-W., LU, B., CAMPBELL, P., & SALOVEY, R. (2000) · The Journal of Bone and Joint Surgery-American Volume · doi:10.2106/00004623-200012000-00004 · Cited by 264
Self-citation
Metadata
- Authors
- McKellop, H., Shen, F.-W., DiMaio, W., & Lancaster, J. G.
- Venue
- Clinical Orthopaedics and Related Research
- Published
- 1999-12
- Volume
- 369
- Issue
- 369
- Pages
- 73-82
- ISSN
- 0009-921X (ISSN-L)
- OpenAlex
- W1968617462
- Cited by
- 184
- References
- 0
- FWCI
- 6.49
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Wear of Gamma-Crosslinked Polyethylene Acetabular Cups Against Roughened Femoral Balls. https://app.cochat.ai/papers/10.1097/00003086-199912000-00008
- The paper
- McKellop, H., Shen, F.-W., DiMaio, W., & Lancaster, J. G. (1999). Wear of Gamma-Crosslinked Polyethylene Acetabular Cups Against Roughened Femoral Balls. Clinical Orthopaedics and Related Research. https://doi.org/10.1097/00003086-199912000-00008
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.