Evidence for the need of bedside accuracy of pulse oximetry in an intensive care unit

Seguin, Philippe, Le Rouzo, Anne, Tanguy, Michèle, Guillou, Yves Marie, Feuillu, Alain, Mallédant, Yannick · 2000 · Critical Care Medicine
Cited by 74

The DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.

Abstract

OBJECTIVE: To compare pulse oximetry saturation (Spo2 with arterial blood gas saturation (SaO2) obtained during clinical routine to determine the optimal lowest reliable value of SpO2 in ventilator-dependent patients before setting up a nurse-directed protocol of FIO2 titration. DESIGN: Prospective clinical study. SETTING: Surgical intensive care unit in a university hospital. PATIENTS: Thirty-three patients with a pulse oximeter probe in whom arterial blood gas was measured with a radial artery line. INTERVENTIONS: SPO2 was recorded by the nurses and compared with SaO2 obtained by blood gas analysis with a co-oximeter. Two sensors currently used in our surgical intensive care unit and connected to a monitor (HP OmniCare M1165/66A; Hewett Packard, Andover, MA) were tested. In group I, the Durensor DS 100A (Nellcor Puritan Bennett, Pleasanton, CA), a reusable sensor, was used. In group II, the Oxisensor D25L (Nellcor Puritan Bennett), a nonreusable sensor, was used. MEASUREMENTS AND MAIN RESULTS: In group 1, 64 data pairs were obtained. In this group, SaO2 ranged from 87 to 98% and SpO2 ranged from 92 to 100%. The bias was -1.90% and the limits of agreement ranged from -5.56 to 1.76%. In group 11, 47 data pairs were obtained. In this group, SaO2 ranged from 87 to 99% and SpO2 ranged from 92 to 100%. The bias was -2.49% and the limits of agreement ranged from -6.62 to 1.64%. CONCLUSIONS: In the range of SaO2 tested, regardless of the sensor used, SpO2 overestimated SaO2. Large limits of agreement were found. Based on this result, the authors concluded that before defining a nurse-directed protocol of FIO2 titration with SpO2, the material used daily must be evaluated. A minimum threshold SpO2 value of 96% in both groups I and II is more reliable to ensure SaO2 > or = 90%.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/00003246-200003000-00017

Cites

2 of 2 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 2 references

Cited by

See all 74

Metadata

Authors
Seguin, P., Le Rouzo, A., Tanguy, M., Guillou, Y. M., Feuillu, A., & Mallédant, Y.
Venue
Critical Care Medicine
Published
2000-03
Volume
28
Issue
3
Pages
703-706
ISSN
0090-3493 (ISSN-L)
OpenAlex
W2014653011
Cited by
74
References
2
FWCI
4.38
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Evidence for the need of bedside accuracy of pulse oximetry in an intensive care unit. https://app.cochat.ai/papers/10.1097/00003246-200003000-00017
The paper
Seguin, P., Le Rouzo, A., Tanguy, M., Guillou, Y. M., Feuillu, A., & Mallédant, Y. (2000). Evidence for the need of bedside accuracy of pulse oximetry in an intensive care unit. Critical Care Medicine. https://doi.org/10.1097/00003246-200003000-00017

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.