Evidence for the need of bedside accuracy of pulse oximetry in an intensive care unit
The DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.
Abstract
OBJECTIVE: To compare pulse oximetry saturation (Spo2 with arterial blood gas saturation (SaO2) obtained during clinical routine to determine the optimal lowest reliable value of SpO2 in ventilator-dependent patients before setting up a nurse-directed protocol of FIO2 titration. DESIGN: Prospective clinical study. SETTING: Surgical intensive care unit in a university hospital. PATIENTS: Thirty-three patients with a pulse oximeter probe in whom arterial blood gas was measured with a radial artery line. INTERVENTIONS: SPO2 was recorded by the nurses and compared with SaO2 obtained by blood gas analysis with a co-oximeter. Two sensors currently used in our surgical intensive care unit and connected to a monitor (HP OmniCare M1165/66A; Hewett Packard, Andover, MA) were tested. In group I, the Durensor DS 100A (Nellcor Puritan Bennett, Pleasanton, CA), a reusable sensor, was used. In group II, the Oxisensor D25L (Nellcor Puritan Bennett), a nonreusable sensor, was used. MEASUREMENTS AND MAIN RESULTS: In group 1, 64 data pairs were obtained. In this group, SaO2 ranged from 87 to 98% and SpO2 ranged from 92 to 100%. The bias was -1.90% and the limits of agreement ranged from -5.56 to 1.76%. In group 11, 47 data pairs were obtained. In this group, SaO2 ranged from 87 to 99% and SpO2 ranged from 92 to 100%. The bias was -2.49% and the limits of agreement ranged from -6.62 to 1.64%. CONCLUSIONS: In the range of SaO2 tested, regardless of the sensor used, SpO2 overestimated SaO2. Large limits of agreement were found. Based on this result, the authors concluded that before defining a nurse-directed protocol of FIO2 titration with SpO2, the material used daily must be evaluated. A minimum threshold SpO2 value of 96% in both groups I and II is more reliable to ensure SaO2 > or = 90%.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Measurement in Medicine: The Analysis of Method Comparison Studies
Altman, D. G., & Bland, J. M. (1983) · The Statistician · doi:10.2307/2987937 · Cited by 4,300
Jubran, A., & Tobin, M. J. (1990) · Chest · doi:10.1378/chest.97.6.1420 · Cited by 296 · cited 3×
“Like Joubran and Tobin (12), we did not analyze carboxyhemoglobin systematically.”
BackgroundMethodInfluential
Cited by
European Resuscitation Council Guidelines for Resuscitation 2010 Section 1. Executive summary
Nolan, J. P., Soar, J., Zideman, D. A., Biarent, D., Bossaert, L. L., Deakin, C., Koster, R. W., Wyllie, J., & Böttiger, B. (2010) · Resuscitation · doi:10.1016/j.resuscitation.2010.08.021 · Cited by 1,375
European Resuscitation Council Guidelines for Resuscitation 2015
Perkins, G. D., Handley, A. J., Koster, R. W., Castrén, M., Smyth, M. A., Olasveengen, T., Monsieurs, K. G., Raffay, V., Gräsner, J.-T., Wenzel, V., et al. (2015) · Resuscitation · doi:10.1016/j.resuscitation.2015.07.015 · Cited by 1,287
European Resuscitation Council Guidelines for Resuscitation 2015
Monsieurs, K. G., Nolan, J. P., Bossaert, L. L., Greif, R., Maconochie, I. K., Nikolaou, N. I., Perkins, G. D., Soar, J., Truhlář, A., Wyllie, J., et al. (2015) · Resuscitation · doi:10.1016/j.resuscitation.2015.07.038 · Cited by 915
European Resuscitation Council Guidelines for Resuscitation 2005
Soar, J., Deakin, C. D., Nolan, J. P., Abbas, G., Alfonzo, A., Handley, A. J., Lockey, D., Perkins, G. D., & Thies, K. (2005) · Resuscitation · doi:10.1016/j.resuscitation.2005.10.004 · Cited by 832
European Resuscitation Council Guidelines for Resuscitation 2015
Maconochie, I. K., Bingham, R., Eich, C., López-Herce, J., Rodríguez-Núñez, A., Rajka, T., Van de Voorde, P., Zideman, D. A., Biarent, D., Monsieurs, K. G., et al. (2015) · Resuscitation · doi:10.1016/j.resuscitation.2015.07.028 · Cited by 450
Metadata
- Authors
- Seguin, P., Le Rouzo, A., Tanguy, M., Guillou, Y. M., Feuillu, A., & Mallédant, Y.
- Venue
- Critical Care Medicine
- Published
- 2000-03
- Volume
- 28
- Issue
- 3
- Pages
- 703-706
- ISSN
- 0090-3493 (ISSN-L)
- OpenAlex
- W2014653011
- Cited by
- 74
- References
- 2
- FWCI
- 4.38
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Evidence for the need of bedside accuracy of pulse oximetry in an intensive care unit. https://app.cochat.ai/papers/10.1097/00003246-200003000-00017
- The paper
- Seguin, P., Le Rouzo, A., Tanguy, M., Guillou, Y. M., Feuillu, A., & Mallédant, Y. (2000). Evidence for the need of bedside accuracy of pulse oximetry in an intensive care unit. Critical Care Medicine. https://doi.org/10.1097/00003246-200003000-00017
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.