Clinical evaluation of the Accutracker II ambulatory blood pressure monitor: assessment of performance in two countries and comparison with sphygmomanometry and intra-arterial blood pressure at rest and during exercise
The DOI matches its Crossref record.
Abstract
In order to assess the Accutracker II (Suntech Medical Instruments, Raleigh, North Carolina, USA), a relatively new ambulatory blood pressure (BP) monitor, versus standard forms of BP measurement, we compared same- and contralateral-arm measurements made, via a t-tube connected to a mercury column sphygmomanometer, by two clinicians using a teaching stethoscope and by intra-arterial recordings. Average systolic BP values obtained using the Accutracker II were similar to both the mercury column and intra-arterial determinations, but average diastolic BP values were lower than both the average mercury column (2.8 +/- 4.2 mmHg, P less than 0.001) and intra-arterial measurements (2.0 +/- 4.7 mmHg, P less than 0.02). During isometric exercise and 100-watt bicycle exercise, there were greater limits of agreement for the differences in BP between the Accutracker II and the intra-arterial transducer than were observed for the resting measurements, but these differences were no greater than those observed between intra-arterial and clinician-determined BP measurements. The clinical performance of the Accutracker II was assessed using 119 hypertensive subjects (84 in Norway and 35 in the USA) who wore the monitor for 24 h. While there was good-to-excellent data return in both countries, there were significantly less error codes secondary to excessive arm motion observed in Norway. Our data demonstrate that the Accutracker II is quite accurate compared with both the mercury column and intra-arterial methods of measuring BP, and performs well during 24 h outpatient activities. Our findings also indicate certain geographical differences which may be important in the performance of ambulatory BP-monitoring studies.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Blumenthal, J. A. (2010) · Archives of Internal Medicine · doi:10.1001/archinternmed.2009.470 · Cited by 578
Ben-Dov, I. Z., Kark, J. D., Ben-Ishay, D., Mekler, J., Ben-Arie, L., & Bursztyn, M. (2007) · Hypertension · doi:10.1161/hypertensionaha.107.087262 · Cited by 367
Nocturnal Blood Pressure Fall on Ambulatory Monitoring in a Large International Database
Staessen, J. A., Bieniaszewski, L., O'Brien, E., Gosse, P., Hayashi, H., Imai, Y., Kawasaki, T., Otsuka, K., Palatini, P., Thijs, L., et al. (1997) · Hypertension · doi:10.1161/01.hyp.29.1.30 · Cited by 310
Blumenthal, J. A., Sherwood, A., Gullette, E. C. D., Babyak, M., Waugh, R., Georgiades, A., Craighead, L. W., Tweedy, D., Feinglos, M., Appelbaum, M., et al. (2000) · Archives of Internal Medicine · doi:10.1001/archinte.160.13.1947 · Cited by 289
Preserved Endothelium-Dependent Vasodilatation in Patients with Essential Hypertension
Cockcroft, J. R., Chowienczyk, P. J., Benjamin, N., & Ritter, J. M. (1994) · New England Journal of Medicine · doi:10.1056/nejm199404143301502 · Cited by 281
Metadata
- Authors
- White, W. B., Lund-Johansen, P., McCabe, E. J., & Omvik, P.
- Venue
- Journal of Hypertension
- Published
- 1989-12
- Volume
- 7
- Issue
- 12
- Pages
- 967-975
- ISSN
- 0263-6352 (ISSN-L)
- OpenAlex
- W2085826045
- Cited by
- 74
- References
- 0
- FWCI
- 2.45
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Clinical evaluation of the Accutracker II ambulatory blood pressure monitor: assessment of performance in two countries and comparison with sphygmomanometry and intra-arterial blood pressure at rest and during exercise. https://app.cochat.ai/papers/10.1097/00004872-198912000-00007
- The paper
- White, W. B., Lund-Johansen, P., McCabe, E. J., & Omvik, P. (1989). Clinical evaluation of the Accutracker II ambulatory blood pressure monitor: assessment of performance in two countries and comparison with sphygmomanometry and intra-arterial blood pressure at rest and during exercise. Journal of Hypertension. https://doi.org/10.1097/00004872-198912000-00007
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.