The DOI matches its Crossref record.
The Effect of Cage Stiffness on the Rate of Lumbar Interbody Fusion: An In Vivo Model Using Poly(L-Lactic Acid) and Titanium Cages
Abstract
STUDY DESIGN: A goat interbody fusion model using poly-(L-lactic acid) and titanium cages was designed to evaluate the effect of cage stiffness on lumbar interbody fusion. OBJECTIVE: To investigate the effect of cage stiffness on the rate of interbody fusion. SUMMARY OF BACKGROUND DATA: Various types of cages considerably exceed the stiffness of vertebral bone, which ultimately may lead to postoperative complications. To avoid these complications, poly-(L-lactic acid) cages with limited stiffness have been designed. The mechanical integrity of the cages remains intact for at least 6 months. METHODS: Interbody fusions were performed at L3-L4 of 15 Dutch milk goats, and one of three cages was randomly implanted: 1) a titanium cage (n = 3), 2) a stiff poly-(L-lactic acid) cage (n = 6), or 3) a flexible poly-(L-lactic acid) cage (n = 6). Interbody fusion was assessed radiographically by three independent observers 3 and 6 months after surgery. RESULTS: At 3 months, all the poly-(L-lactic acid) specimens showed ingrowth of new bone, but with radiolucency in the fusion mass. At 6 months, solid arthrodesis was observed in four of six poly-(L-lactic acid) specimens, advanced ingrowth in one specimen, and infection in one specimen. Titanium cages showed ingrowth of bone, but with radiolucency in the fusion mass. Interbody fusion using poly-(L-lactic acid) cages showed a significantly higher rate statistically (P = 0.016) and more complete fusion than titanium cages of the same design. CONCLUSIONS: The reduced stiffness of poly-(L-lactic acid) cages showed enhanced interbody fusion, as compared with titanium cages after 6 months. Bioabsorbable poly-(L-lactic acid) cages thus may be a viable alternative to current interbody cage devices, thereby avoiding the concomitant problems related to their excessive stiffness. However, the bioabsorbability of the poly-(L-lactic acid) cages awaits investigation in a long-term study currently underway.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Navarro, M., Michiardi, A., Castaño, O., & Planell, J. A. (2008) · Journal of The Royal Society Interface · doi:10.1098/rsif.2008.0151 · Cited by 1,513
Davachi, S. M., & Kaffashi, B. (2015) · Polymer-Plastics Technology and Engineering · doi:10.1080/03602559.2014.979507 · Cited by 156
“PLLA cages thus may be a viable alternative to current inter body cage devices([196]).”
BackgroundIn vitro and in vivo degradation of bioabsorbable PLLA spinal fusion cages
van Dijk, M., Tunc, D. C., Smit, T. H., Higham, P., Burger, E. H., & Wuisman, P. I. J. M. (2002) · Journal of Biomedical Materials Research · doi:10.1002/jbm.10466 · Cited by 78
Self-citation
Metadata
- Authors
- van Dijk, M., Smit, T. H., Sugihara, S., Burger, E. H., & Wuisman, P. I.
- Venue
- Spine
- Published
- 2002-04
- Volume
- 27
- Issue
- 7
- Pages
- 682-688
- ISSN
- 0362-2436 (ISSN-L)
- OpenAlex
- W2054121842
- Cited by
- 165
- References
- 0
- FWCI
- 6.16
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: The Effect of Cage Stiffness on the Rate of Lumbar Interbody Fusion: An In Vivo Model Using Poly(L-Lactic Acid) and Titanium Cages. https://app.cochat.ai/papers/10.1097/00007632-200204010-00003
- The paper
- van Dijk, M., Smit, T. H., Sugihara, S., Burger, E. H., & Wuisman, P. I. (2002). The Effect of Cage Stiffness on the Rate of Lumbar Interbody Fusion: An In Vivo Model Using Poly(L-Lactic Acid) and Titanium Cages. Spine. https://doi.org/10.1097/00007632-200204010-00003
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.