Validation of two automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension: the Omron M6 (HEM-7001-E) and the Omron R7 (HEM 637-IT)
Abstract
Background Two electronic devices for self-measurement of blood pressure – a brachial monitor, the Omron M6, and a wrist monitor, the Omron R7 – were evaluated in two separate studies according to the International Protocol of the European Society of Hypertension. Design The International Validation Protocol is divided into two phases: the first phase is performed on 15 selected participants (45 pairs of blood pressure measurements); if the device passes this phase, 18 supplementary participants are included (54 pairs of blood pressure measurements) making a total number of 33 participants (99 pairs of blood pressure measurements) on whom the final validation is performed. Methods The same methodology recommended by the European Society of Hypertension protocol was applied for both studies. In each study and for each participant, four blood pressure measurements were taken simultaneously by two trained observers using mercury sphygmomanometers alternately with three measurements taken by the tested device. The difference between the blood pressure value given by the device and that obtained by the two observers (mean of the two observers) was calculated for each measure. The 99 pairs of blood pressure differences were classified into three categories (≤5, ≤10 and ≤15 mmHg). The number of differences in each category was compared with the number required by the International Protocol. An individual analysis was then done to determine the number of comparisons ≤5 mmHg for each participant. At least 22 of the 33 participants should have two of their three comparisons ≤5 mmHg. Results In both studies, the two tested devices passed the first and the second phases of the validation process. The average differences between the device and mercury sphygmomanometer readings were 0.8±2.7 and −1.9±3.3 mmHg for systolic and diastolic blood pressure, respectively, for the Omron M6 device, and 0.2±4.2 and 0.2±2.9 mmHg for systolic and diastolic blood pressure, respectively, for the Omron R7 device. For both devices, readings differing by less than 5, 10 and 15 mmHg for systolic and diastolic blood pressure values fulfill the recommendation criteria of the International Protocol as well as the individual analysis. Conclusions The Omron M6 (HEM-7001-E) and the Omron R7 (HEM 637-IT) devices fulfilled the validation recommendations of the International Protocol.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Characterization of the oscillometric method for measuring indirect blood pressure
Geddes, L. A., Voelz, M., Combs, C., Reiner, D., & Babbs, C. F. (1982) · Annals of Biomedical Engineering · doi:10.1007/bf02367308 · Cited by 340
Reference Values for Self-recorded Blood Pressure: A Meta-analysis of Summary Data
Thijs, L., Staessen, J. A., Celis, H., de Gaudemaris, R., Imai, Y., Julius, S., & Fagard, R. (1998) · Archives of Internal Medicine · doi:10.1001/archinte.158.5.481 · Cited by 143
Cited by
Hypertension, Diabetes and Overweight: Looming Legacies of the Biafran Famine
Hult, M., Tornhammar, P., Ueda, P., Chima, C., Edstedt Bonamy, A.-K., Ozumba, B., & Norman, M. (2010) · PLoS ONE · doi:10.1371/journal.pone.0013582 · Cited by 306
Active Video Games for Youth: A Systematic Review
Barnett, A., Cerin, E., & Baranowski, T. (2011) · Journal of Physical Activity and Health · doi:10.1123/jpah.8.5.724 · Cited by 282
Uganda Genome Resource Enables Insights into Population History and Genomic Discovery in Africa
Gurdasani, D., Carstensen, T., Fatumo, S., Chen, G., Franklin, C. S., Prado-Martinez, J., Bouman, H., Abascal, F., Haber, M., Tachmazidou, I., et al. (2019) · Cell · doi:10.1016/j.cell.2019.10.004 · Cited by 255
Metadata
- Authors
- Topouchian, J. A., El Assaad, M. A., Orobinskaia, L. V., El Feghali, R. N., & Asmar, R. G.
- Venue
- Blood Pressure Monitoring
- Published
- 2006-06
- Volume
- 11
- Issue
- 3
- Pages
- 165-171
- ISSN
- 1359-5237 (ISSN-L)
- OpenAlex
- W2090415055
- Cited by
- 158
- References
- 2
- FWCI
- 5.03
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Validation of two automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension: the Omron M6 (HEM-7001-E) and the Omron R7 (HEM 637-IT). https://app.cochat.ai/papers/10.1097/01.mbp.0000209078.17246.34
- The paper
- Topouchian, J. A., El Assaad, M. A., Orobinskaia, L. V., El Feghali, R. N., & Asmar, R. G. (2006). Validation of two automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension: the Omron M6 (HEM-7001-E) and the Omron R7 (HEM 637-IT). Blood Pressure Monitoring. https://doi.org/10.1097/01.mbp.0000209078.17246.34
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.