Dual-energy X-Ray Absorptiometry: Interpreting Reports and Serial Measurements
Abstract
Physicians and other health care providers requesting dual-energy x-ray bone density studies must be able to critically review and interpret such studies. This requires an understanding of the computerized result components of the studies and how such data can be applied in clinical practice. The timing and interpretation of serial bone density studies is governed by the concepts of precision and least significant change. These statistical concepts are well within the purview of the practicing physician. Serial bone density studies cannot be interpreted without prior knowledge of the precision and least significant change of the test.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Changes in the BMD of less than 3–6 % at the hip and 2–4 % at the spine from test to test may be due to the precision error of the testing itself.”
From Clinician’s Guide to Prevention and Treatment of Osteoporosis (2014) · cites this paper as [97]
Citations
Cites
Still being checked.
Cited by
Clinician’s Guide to Prevention and Treatment of Osteoporosis
Cosman, F., de Beur, S. J., LeBoff, M. S., Lewiecki, E. M., Tanner, B., Randall, S., & Lindsay, R. (2014) · Osteoporosis International · doi:10.1007/s00198-014-2794-2 · Cited by 4,149
“This least significant change (LSC) varies with the specific instrument used, patient population being assessed, measurement site, technologist’s skill with patient positioning and test analysis, and the confidence intervals used [97].”
BackgroundThe clinician's guide to prevention and treatment of osteoporosis
LeBoff, M. S., Greenspan, S. L., Insogna, K. L., Lewiecki, E. M., Saag, K. G., Singer, A. J., & Siris, E. S. (2022) · Osteoporosis International · doi:10.1007/s00198-021-05900-y · Cited by 1,487
Long-term and sequential treatment for osteoporosis
Foessl, I., Dimai, H. P., & Obermayer-Pietsch, B. (2023) · Nature Reviews Endocrinology · doi:10.1038/s41574-023-00866-9 · Cited by 273
Comparison of Central and Peripheral Bone Mineral Density Measurements in Postmenopausal Women
Abdelmohsen, A. M. (2017) · Journal of Chiropractic Medicine · doi:10.1016/j.jcm.2017.08.001 · Cited by 24
Bell, W., Davies, J. S., Evans, W. D., & Scanlon, M. F. (1999) · Annals of Human Biology · doi:10.1080/030144699282985 · Cited by 23
Metadata
- Authors
- BONNICK, S. L.
- Affiliations
- University of North Texas (US)
- Corresponding author
- BONNICK, SYDNEY L.
- Venue
- Clinical Obstetrics & Gynecology
- Published
- 2013-12
- Volume
- 56
- Issue
- 4
- Pages
- 677-685
- ISSN
- 0009-9201 (ISSN-L)
- OpenAlex
- W2082039678
- Cited by
- 9
- References
- 0
- Citations by year
- 2023: 1 · 2022: 2 · 2017: 1 · 2016: 2 · 2015: 1 · 2014: 1
- FWCI
- 1.37
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Clinical Obstetrics & Gynecology (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Orthopedics and Sports Medicine › Bone health and osteoporosis research
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Health Sciences › Medicine › Physiology (medical) › Body Composition Measurement Techniques
- Keywords
- dual-energy X-ray absorptiometry, bone mineral density, osteoporosis diagnosis, measurement precision, least significant change, test interpretation, clinical practice
- MeSH terms
- Female, Humans, Radiographic Image Interpretation, Computer-Assisted, Absorptiometry, Photon, Osteoporosis, Postmenopausal
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Dual-energy X-Ray Absorptiometry: Interpreting Reports and Serial Measurements. https://app.cochat.ai/papers/10.1097/grf.0b013e3182a8240c
- The paper
- BONNICK, S. L. (2013). Dual-energy X-Ray Absorptiometry: Interpreting Reports and Serial Measurements. Clinical Obstetrics & Gynecology. https://doi.org/10.1097/grf.0b013e3182a8240c
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.