Validation of the Oscar 2 oscillometric 24-h ambulatory blood pressure monitor according to the British Hypertension Society protocol
The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.
Abstract
OBJECTIVES: Accuracy of blood pressure measuring devices is of prime importance and should be validated before devices are used clinically. We carried out an independent evaluation of the Oscar 2 oscillometric ambulatory blood pressure monitor (SunTech Medical, Model 222) according to the British Hypertension Society (BHS) Protocol. METHODS: Validation of the Oscar 2 was carried out in accordance with Part 1 of the British Hypertension Society protocol. Having successfully completed the first three phases, in the static validation phase blood pressure measurements were made on 114 participants simultaneously by two observers blinded from each other's readings and those of the device, giving 255 data pairs for systolic and diastolic blood pressure. Readings were made using simultaneous same-arm measurement, each observer using a dual-head binaural stethoscope and a calibrated mercury sphygmomanometer. Data were recorded independently, to the nearest 2 mmHg and were checked by the supervisor who operated the device. RESULTS: The device mean and observer means (and standard deviations) were exceptionally close across a high range of pressures, with a mean difference for the better observer of 0+/-7 mmHg and -1+/-6 mmHg for systolic and diastolic blood pressure, respectively. With 62% of all systolic blood pressure standard vs. device differences equal or less than 5 mmHg, and 70% of all similar differences for diastolic blood pressure equal to or less than 5 mmHg, the device was categorized as Grade A for systolic and diastolic blood pressure. CONCLUSION: The Oscar 2 was graded A for both systolic and diastolic blood pressure by the British Hypertension Society protocol and can be recommended for clinical use in an adult population.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
The pathophysiology of hypertension
Beevers, G., Lip, G. Y. H., & O'Brien, E. (2001) · BMJ · doi:10.1136/bmj.322.7291.912 · Cited by 488
Cited by
Pimenta, E., Gaddam, K. K., Oparil, S., Aban, I., Husain, S., Dell'Italia, L. J., & Calhoun, D. A. (2009) · Hypertension · doi:10.1161/hypertensionaha.109.131235 · Cited by 531
Large Artery Stiffness Assessment Using SphygmoCor Technology
Butlin, M., & Qasem, A. (2016) · Pulse · doi:10.1159/000452448 · Cited by 244
Padden, C., & James, J. E. (2017) · Journal of Developmental and Physical Disabilities · doi:10.1007/s10882-017-9547-z · Cited by 215
“The reliability and validity of this monitor have been demonstrated (Goodwin et al. 2007).”
BackgroundBoudet, G., Walther, G., Courteix, D., Obert, P., Lesourd, B., Pereira, B., Chapier, R., Vinet, A., Chamoux, A., Naughton, G., et al. (2017) · European Journal of Preventive Cardiology · doi:10.1177/2047487316679523 · Cited by 39
“Blood pressure was measured after a 15-minute rest period in the supine position as the mean of three consecutive measurements (SunTech Medical, Model 222; North California, USA).(34)”
Method
Metadata
- Authors
- Goodwin, J., Bilous, M., Winship, S., Finn, P., & Jones, S. C.
- Venue
- Blood Pressure Monitoring
- Published
- 2007-04
- Volume
- 12
- Issue
- 2
- Pages
- 113-117
- ISSN
- 1359-5237 (ISSN-L)
- OpenAlex
- W2039014945
- Cited by
- 135
- References
- 1
- FWCI
- 2.59
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.