Sampling frequency of the electrocardiogram for spectral analysis of the heart rate variability
Abstract
The R-R interval measurement from digitized electrocardiograms (ECG) contains an error due to the finite sampling frequency which may jeopardize the beat-to-beat analysis of the heart rate. In this paper, we develop a model to describe and quantitate this error. The "measured" R-R interval is modeled as the sum of the "true" R-R interval and of the error of measurement. The first and second order statistics of the error are computed in order to investigate its influence on the heart rate variability (HRV) power spectrum. They are found to be only functions of the ECG sampling frequency and, in particular, the power spectrum of the error contributes an additive high-pass filter-like term (colored noise) to the power spectrum of the HRV. The accuracy of the model is tested via a simulation procedure. The model indicates that the relative balance between the HRV and the error power spectra is important and should be checked before any variability analysis on the heart rate. This balance may be favorable to the error when 1) the sampling frequency of the ECG is too low, and/or 2) the variability of the heart rate is too little. In these cases, the HRV spectrum analysis may not give reliable results. Two tests are proposed in order to evaluate the error influence either in specific frequency bands or in the total frequency range.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Akselrod, S., Gordon, D., Ubel, F. A., Shannon, D. C., Berger, A. C., & Cohen, R. J. (1981) · Science · doi:10.1126/science.6166045 · Cited by 5,079
Pagani, M., Lombardi, F., Guzzetti, S., Rimoldi, O., Furlan, R., Pizzinelli, P., Sandrone, G., Malfatto, G., Dell'Orto, S., & Piccaluga, E. (1986) · Circulation Research · doi:10.1161/01.res.59.2.178 · Cited by 4,130
Kleiger, R. E., Miller, J. P., Bigger, J. T., & Moss, A. J. (1987) · The American Journal of Cardiology · doi:10.1016/0002-9149(87)90795-8 · Cited by 3,995
Cited by
Heart Rate Variability: Standards of Measurement, Physiological Interpretation, and Clinical Use
Electrophysiology, T. F. O. T. E. S. O. C. T. N. A. (1996) · Circulation · doi:10.1161/01.cir.93.5.1043 · Cited by 17,159
Heart rate variability: Standards of measurement, physiological interpretation, and clinical use
Malik, M., Bigger, J. T., Camm, A. J., Kleiger, R. E., Malliani, A., Moss, A. J., & Schwartz, P. J. (1996) · European Heart Journal · doi:10.1093/oxfordjournals.eurheartj.a014868 · Cited by 15,289
An overview of heart rate variability metrics and norms
Shaffer, F., & Ginsberg, J. P. (2017) · Frontiers in Public Health · doi:10.3389/fpubh.2017.00258 · Cited by 7,185
Heart rate variability: origins, methods, and interpretive caveats
BERNTSON, G. G., THOMAS BIGGER, J., ECKBERG, D. L., GROSSMAN, P., KAUFMANN, P. G., MALIK, M., NAGARAJA, H. N., PORGES, S. W., SAUL, J. P., STONE, P. H., et al. (1997) · Psychophysiology · doi:10.1111/j.1469-8986.1997.tb02140.x · Cited by 3,962
Malik, M. (1996) · Annals of Noninvasive Electrocardiology · doi:10.1111/j.1542-474x.1996.tb00275.x · Cited by 850
Metadata
- Authors
- Merri, M., Farden, D. C., Mottley, J. G., & Titlebaum, E. L.
- Venue
- IEEE Transactions on Biomedical Engineering
- Published
- 1990
- Volume
- 37
- Issue
- 1
- Pages
- 99-106
- ISSN
- 0018-9294 (ISSN-L)
- OpenAlex
- W2011565562
- Cited by
- 248
- References
- 3
- FWCI
- 4.40
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Sampling frequency of the electrocardiogram for spectral analysis of the heart rate variability. https://app.cochat.ai/papers/10.1109/10.43621
- The paper
- Merri, M., Farden, D. C., Mottley, J. G., & Titlebaum, E. L. (1990). Sampling frequency of the electrocardiogram for spectral analysis of the heart rate variability. IEEE Transactions on Biomedical Engineering. https://doi.org/10.1109/10.43621
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.