CoChat

An Implantable Telemetry Platform System for In Vivo Monitoring of Physiological Parameters

Valdastri, P., Menciassi, A., Arena, A., Caccamo, C., Dario, P. · 2004 · IEEE Transactions on Information Technology in Biomedicine
Cited by 117

The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.

Abstract

This paper describes a microcontroller-based multichannel telemetry system, suitable for in vivo monitoring of physiological parameters. The device can digitalize and transmit up to three analog signals coming from different sensors. The telemetry transmission is obtained by using a carrier frequency of 433.92 MHz and an amplitude-shift keying modulation. The signal data rate is 13 kb/s per channel. The digital microcontroller provides good flexibility and interesting performance, such as the threshold monitoring, the transmission error detection, and a low power consumption, thanks to the implementation of a sleep mode. The small overall size (less than 1 cm3), the power density compatible with current regulations for the design of implantable devices, and the dedicated packaging make the system suitable for in vivo monitoring in humans. The design, fabrication, operation, packaging, and performance of the system are described in this paper. An in vivo pressure monitoring case study is described as well.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1109/titb.2004.834389

Cites

1 of 1 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Cited by

See all 117

Metadata

Authors
Valdastri, P., Menciassi, A., Arena, A., Caccamo, C., & Dario, P.
Venue
IEEE Transactions on Information Technology in Biomedicine
Published
2004-09
Volume
8
Issue
3
Pages
271-278
ISSN
1089-7771 (ISSN-L)
OpenAlex
W2106872146
Cited by
117
References
1
FWCI
4.29
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.