The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.
Signal-to-Noise Ratio in Doppler Radar System for Heart and Respiratory Rate Measurements
Abstract
A CMOS Doppler radar sensor has been developed and used to measure motion due to heart and respiration. The quadrature direct-conversion radar transceiver has been fully integrated in 0.25-mum CMOS, the baseband analog signal conditioning has been developed on a printed circuit board, and digital signal processing has been performed inMatlab. The theoretical signal-to-noise ratio (SNR) is derived based on the radar equation, the direct-conversion receiver's properties, oscillator phase noise, range correlation, and receiver noise. Heart and respiration signatures and rates have been measured at ranges from 0.5 to 2.0 m on 22 human subjects wearing normal T-shirts. The theoretical SNR expression was validated with this study. The heart rates found with the radar sensor were compared with a three-lead electrocardiogram, and they were within 5 beats/min with 95% confidence for 16 of 22 subjects at a 0.5-m range and 11 of 22 subjects at a 1.0-m range. The respiration rates found with the radar sensor were compared with those found using a piezoelectric respiratory effort belt, and the respiration rates were within five respirations per minute for 18 of 22 subjects at a 0.5-m range, 17 of 22 subjects at a 1.0-m range, and 19 of 22 subjects at a 1.5-m range.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Through signal processing, it is possible to extract biological information, such as breathing patterns, heart rate and full-body motion, from radio signals.
“Through signal processing, it is possible to extract biological information, such as breathing patterns, heart rate and full-body motion from these findings71,73–75.”
From The future of sleep health: a data-driven revolution in sleep science and medicine (2020) · cites this paper as [75]
High frequency and sub-millimetre wavelength radio technologies have demonstrated the ability to capture physiological signals without body contact.
“In the past decade, high frequency and sub-millimetre wavelength radio technologies have demonstrated the ability to capture physiological signals without body contact.”
From The future of sleep health: a data-driven revolution in sleep science and medicine (2020) · cites this paper as [75]
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Flagged first, then by citations
Research works printed but not linked: 26
- 10.1063/1.2798937 · DOI
- droitcour, 2002, 0.25 <formula formulatype="inline"><tex notation="tex">$\mu{\hbox{m}}$</tex></formula> cmos and bicmos single-chip direct-conversion doppler radars for remote sensing of vital signs, Int Solid-State Circuits Conf Dig, 1, 348
- 10.1109/TMTT.2004.823552 · DOI
- droitcour, 2004, chest motion sensing with modified silicon base station chips, IEICE Trans Electron, e87 c, 1524
- 10.1109/LMWC.2008.925113 · DOI
- 10.1016/S0190-9622(99)70126-X · DOI
- 10.1016/j.burns.2003.10.003 · DOI
- 2000, Code of federal regulations title 47 Telecommunications part 15 Radio frequency devices
- 10.1109/SECON.1993.465731 · DOI
- skolnik, 1990, Radar Handbook, 1 1
- host-madsen, 2008, Signal Processing Techniques for Knowledge Extraction and Information Fusion
- 10.1109/IEMBS.2009.5333206 · DOI
- 10.1109/TMTT.2007.895653 · DOI
- li, 2008, complex signal demodulation and random body movement cancellation techniques for non-contact vital sign detection, IEEE MTT-S Int Microw Symp Dig, 567
- 10.1063/1.2798937 · DOI
- oum, 2008, two frequency radar sensor for non-contact vital sign monitor, IEEE MTT-S Int Microw Symp Dig, 919
- 10.1109/MWSYM.2009.5165949 · DOI
- 10.1109/62.978359 · DOI
- 10.1109/MWSYM.2001.966866 · DOI
- 10.1002/bem.2250130610 · DOI
- 10.1183/09031936.97.10081865 · DOI
- 10.1007/BF02441473 · DOI
- degroote, 1997, chest wall motion during tidal breathing, J Appl Physiol, 83, 1531, 10.1152/jappl.1997.83.5.1531 · DOI
- 10.1109/10.362922 · DOI
- jollife, 2002, Principal Component Analysis
- 10.1109/MWSYM.2008.4633266 · DOI
Cited by
Sorted by citations
The future of sleep health: a data-driven revolution in sleep science and medicine
Perez-Pozuelo, I., Zhai, B., Palotti, J., Mall, R., Aupetit, M., Garcia-Gomez, J. M., Taheri, S., Guan, Y., & Fernandez-Luque, L. (2020) · npj Digital Medicine · doi:10.1038/s41746-020-0244-4 · Cited by 311
Metadata
- Authors
- Droitcour, A. D., Boric-Lubecke, O., & Kovacs, G. T. A.
- Affiliations
- Stanford University (US)
- Venue
- IEEE Transactions on Microwave Theory and Techniques
- Published
- 2009-10
- Volume
- 57
- Issue
- 10
- Pages
- 2498-2507
- ISSN
- 0018-9480 (ISSN-L)
- 1557-9670
- OpenAlex
- W2098649495
- Cited by
- 162
- References
- 1
- Citations by year
- 2026: 4 · 2025: 5 · 2024: 10 · 2023: 13 · 2022: 7 · 2021: 19 · 2020: 15 · 2019: 8 · 2018: 16 · 2017: 17 · 2016: 16 · 2015: 5 · 2014: 10 · 2013: 6 · 2012: 3
- FWCI
- 2.20
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- IEEE Transactions on Microwave Theory and Techniques (publishedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Physical Sciences › Engineering › Aerospace Engineering › Radar Systems and Signal Processing
- Physical Sciences › Engineering › Biomedical Engineering › Wireless Body Area Networks
- Keywords
- Doppler radar, signal-to-noise ratio, heart rate measurement, respiratory rate measurement, direct conversion receiver, oscillator phase noise, radar equation, non-contact vital sign monitoring
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review