Abstract
Aims To assess the performance of a handheld bedside ketone sensor in the face of likely metabolic disturbances in diabetic ketoacidosis, namely: pH, glucose and acetoacetate. Methods The effects of pH (7.44–6.83), glucose (5–50 mmol/l) and acetoacetate (0–5 mmol/l) were examined in venous blood to investigate the accuracy of betahydroxybutyrate measurement (0–5 mmol/l) by a handheld ketone sensor. Sensor results were compared with a reference method. Linear regression models were fitted to the difference between the methods with the concentration of metabolite as the explanatory factor. Results Decreasing pH and increasing glucose had no effect on the accuracy of the handheld ketone sensor; the gradients of the fitted lines were −0.14 and −0.003, respectively. The 95% confidence intervals were −0.7–0.4 and −0.01–0.004, respectively (P = 0.59 and 0.4, respectively). In the acetoacetate study, a positive relationship between the sensor and reference method results was found, the gradient was 0.09. The 95% confidence interval was 0.05–0.14 (P ≤ 0.001), indicating that high concentrations of acetoacetate interfere with the sensor performance. Conclusions Acidosis and hyperglycaemia have minimal effects on the sensor performance. However, high concentrations of acetoacetate result in some overestimation of betahydroxybutyrate. This bedside ketone sensor provides useful data over a broad range of conditions likely to be encountered during moderate to severe diabetic ketoacidosis.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Relationship of blood acetoacetate and 3-hydroxybutyrate in diabetes
Stephens, J. M., Sulway, M. J., & Watkins, P. J. (1971) · Diabetes · doi:10.2337/diab.20.7.485 · Cited by 99
Umpierrez, G. E., Watts, N. B., & Phillips, L. S. (1995) · Diabetes Care · doi:10.2337/diacare.18.1.137 · Cited by 54
Laboratory and Clinical Evaluation of Assays for β-Hydroxybutyrate
Porter, W. H., Yao, H. H., & Karounos, D. G. (1997) · American Journal of Clinical Pathology · doi:10.1093/ajcp/107.3.353 · Cited by 30
Cited by
Joint British Diabetes Societies guideline for the management of diabetic ketoacidosis
Savage, M. W., Dhatariya, K. K., Kilvert, A., Rayman, G., Rees, J. A. E., Courtney, C. H., Hilton, L., Dyer, P. H., Hamersley, M. S., & Societies, F. T. J. B. D. (2011) · Diabetic Medicine · doi:10.1111/j.1464-5491.2011.03246.x · Cited by 338
Teymourian, H., Moonla, C., Tehrani, F., Vargas, E., Aghavali, R., Barfidokht, A., Tangkuaram, T., Mercier, P. P., Dassau, E., & Wang, J. (2020) · Analytical Chemistry · doi:10.1021/acs.analchem.9b05109 · Cited by 278
Berard, L. D., Siemens, R., & Woo, V. (2018) · Canadian Journal of Diabetes · doi:10.1016/j.jcjd.2017.10.007 · Cited by 69
Noyes, K. J., Crofton, P., Bath, L. E., Holmes, A., Stark, L., Oxley, C. D., & Kelnar, C. J. H. (2007) · Pediatric Diabetes · doi:10.1111/j.1399-5448.2007.00240.x · Cited by 65
Ceriotti, F., Kaczmarek, E., Guerra, E., Mastrantonio, F., Lucarelli, F., Valgimigli, F., & Mosca, A. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814563351 · Cited by 45
“Similarly it was recently reported that Optium BHB measurements were adversely influenced by acetoacetate, suggesting poor specificity of the BHB dehydrogenase enzyme used.(32) This was contradicted in this study as acetoacetate at the concentrations tested did not influence the accuracy of BHB measurements (Figure 4).”
Result
Metadata
- Authors
- Khan, A. S. A., Talbot, J. A., Tieszen, K. L., Gardener, E. A., Gibson, J. M., & New, J. P.
- Venue
- Diabetic Medicine
- Publisher
- Wiley
- Published
- 2004-06-21
- Volume
- 21
- Issue
- 7
- Pages
- 782-785
- ISSN
- 0742-3071 (ISSN-L)
- 1464-5491
- OpenAlex
- W2124935320
- Cited by
- 32
- References
- 4
- FWCI
- 1.14
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Evaluation of a bedside blood ketone sensor: the effects of acidosis, hyperglycaemia and acetoacetate on sensor performance. https://app.cochat.ai/papers/10.1111/j.1464-5491.2004.01233.x
- The paper
- Khan, A. S. A., Talbot, J. A., Tieszen, K. L., Gardener, E. A., Gibson, J. M., & New, J. P. (2004). Evaluation of a bedside blood ketone sensor: the effects of acidosis, hyperglycaemia and acetoacetate on sensor performance. Diabetic Medicine. https://doi.org/10.1111/j.1464-5491.2004.01233.x
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.