Bradycardia and Asystole with the Use of Vagus Nerve Stimulation for the Treatment of Epilepsy: A Rare Complication of Intraoperative Device Testing
Open accessAbstract
Summary: Purposes: A 56‐year‐old man with mild mental retardation, right congenital hemiparesis, and refractory partial seizures was referred for vagus nerve stimulation (VNS). Methods: Routine lead diagnostic testing during the surgical procedure (1.0 mA, 20 Hz, and 500 μs, for ˜ 17 s) resulted, during the initial two stimulations, in a bradycardia of ˜30 beats/min. A third attempt led to transient asystole that required atropine and brief cardiopulmonary resuscitation. Results: The procedure was immediately terminated, the device removed, and the patient recovered completely. A postoperative cardiologic evaluation, including an ECG, 24‐h Holter monitor, echocardiogram, and a tilt‐table test, was normal. Conclusions: Possible mechanisms for the bradycardia/asystole include stimulation of cervical cardiac branches of the vagus nerve either by collateral current spread or directly by inadvertent placement of the electrodes on one of these branches; improper plugging of the electrodes into the pulse generator, resulting in erratic varying intensity of stimulation; reverse polarity; and idiosyncratic‐type reaction in a hypersusceptible individual. The manufacturer reports the occurrence rate in ˜3,500 implants for this intraoperative event to be approximately one in 875 cases or 0.1%.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
The anatomical basis for transcutaneous auricular vagus nerve stimulation
Butt, M. F., Albusoda, A., Farmer, A. D., & Aziz, Q. (2020) · Journal of Anatomy · doi:10.1111/joa.13122 · Cited by 603
Vagus-nerve stimulation for the treatment of epilepsy
Ben-Menachem, E. (2002) · The Lancet Neurology · doi:10.1016/s1474-4422(02)00220-x · Cited by 560
Vagus nerve stimulation: a new tool for brain research and therapy
George, M. S., Sackeim, H. A., Rush, A. J., Marangell, L. B., Nahas, Z., Husain, M. M., Lisanby, S., Burt, T., Goldman, J., & Ballenger, J. C. (2000) · Biological Psychiatry · doi:10.1016/s0006-3223(99)00308-x · Cited by 461
Morris, G. L., Gloss, D., Buchhalter, J., Mack, K. J., Nickels, K., & Harden, C. (2013) · Neurology · doi:10.1212/wnl.0b013e3182a393d1 · Cited by 450
Therapeutic mechanisms of vagus nerve stimulation
Henry, T. R. (2002) · Neurology · doi:10.1212/wnl.59.6_suppl_4.s3 · Cited by 415
Metadata
- Authors
- Asconapé, J. J., Moore, D. D., Zipes, D. P., Hartman, L. M., & Duffell, W. H.
- Venue
- Epilepsia
- Publisher
- Wiley
- Published
- 1999-10
- Volume
- 40
- Issue
- 10
- Pages
- 1452-1454
- ISSN
- 0013-9580 (ISSN-L)
- 1528-1167
- OpenAlex
- W2074243938
- Cited by
- 170
- References
- 0
- FWCI
- 3.59
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Bradycardia and Asystole with the Use of Vagus Nerve Stimulation for the Treatment of Epilepsy: A Rare Complication of Intraoperative Device Testing. https://app.cochat.ai/papers/10.1111/j.1528-1157.1999.tb02019.x
- The paper
- Asconapé, J. J., Moore, D. D., Zipes, D. P., Hartman, L. M., & Duffell, W. H. (1999). Bradycardia and Asystole with the Use of Vagus Nerve Stimulation for the Treatment of Epilepsy: A Rare Complication of Intraoperative Device Testing. Epilepsia. https://doi.org/10.1111/j.1528-1157.1999.tb02019.x
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.