Quality‐of‐Life During DDD and Dual Sensor VVIR Pacing
The DOI matches its Crossref record.
Abstract
Twenty‐one patients (mean age 68 ± 8 years) with dual‐sensor (QT + activity) DDDR pacemaker were randomly assigned to a crossover, double‐blind study in order to evaluate their quality‐of‐life scores. All pacemakers were implanted for sick sinus syndrome (8 patients) or complete heart block (13 patients). The pacemakers were randomly programmed to VVIR or DDD pacing modes for 2‐week periods and then the pacing mode was switched for another 2‐week period. At the end of each period, the quality‐of‐life was evaluated by a questionnaire with regard to cardiovascular symptoms, physical activity, psychosocial and emotional functioning, and self‐perceived health. Nineteen questions were scored 0–5 points each. Significant improvement in the mean total quality‐of‐life score (20.5 ± 14.9 vs 34.8 ± 17.4) as well as in dyspnea on effort, dizzy spells, palpitation, sweating, fatigue, lethargy, emotional functioning, and self‐perceived health was observed during DDD compared to VVIR pacing. No question was scored in favor of VVIR pacing mode. Significant improvements during DDD pacing was demonstrated in all subgroups of patients (sick sinus syndrome, chronotropically competent and incompetent patients, and patients with high degree AV block). Eighteen patients preferred DDD pacing mode, while only one preferred WIR pacing mode. Two remaining patients expressed no preference. The results suggest that DDD pacing offers better quality‐of‐life than dual sensor VVIR pacing in all subgroups of patients commonly indicated for pacemaker implantation.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
No references could be resolved yet.
Research works printed but not linked: 18
- Kailof I., 1975, Haemodynamic effect of atrial triggered versus fixed rate pacing at rest and during exercise in complete heart block, Acta Med Scand, 197, 195, 10.1111/j.0954-6820.1975.tb04902.x · doi:10.1111/j.0954-6820.1975.tb04902.x
- 10.1016/0002-9149(88)90938-1 · DOI
- 10.1111/j.1540-8159.1989.tb01895.x · doi:10.1111/j.1540-8159.1989.tb01895.x
- Lukl J, 1992, In which patients with atrioventricular block is optimal indication for DDD pacing, Vnitrni Lek, 38, 1011
- Shumaker SA, 1990, Quality of Life Assessments in Clinical Trials, 95
- 10.1111/j.1540-8159.1985.tb05726.x · DOI
- 10.1136/hrt.50.2.112 · DOI · Paper page
- 10.1136/hrt.58.6.613 · DOI
- 10.1111/j.1540-8159.1989.tb02671.x · doi:10.1111/j.1540-8159.1989.tb02671.x
- 10.1002/clc.4960120907 · DOI
- 10.1111/j.1540-8159.1991.tb04110.x · DOI
- Bubien RS, 1990, A randomized comparison of quality of life and exercise capacity with DDD and VVIR Pacing (abstract), PACE, 13, 524
- 10.1136/hrt.65.4.188 · DOI
- 10.1111/j.1540-8159.1990.tb06896.x · DOI
- 10.1111/j.1540-8159.1992.tb02920.x · DOI
- 10.1136/hrt.64.5.313 · DOI
- 10.1016/S0735-1097(10)80186-X · DOI
- 10.1111/j.1540-8159.1993.tb01818.x · doi:10.1111/j.1540-8159.1993.tb01818.x
Cited by
Kusumoto, F. M., Schoenfeld, M. H., Barrett, C., Edgerton, J. R., Ellenbogen, K. A., Gold, M. R., Goldschlager, N. F., Hamilton, R. M., Joglar, J. A., Kim, R. J., et al. (2019) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2018.10.044 · Cited by 1,008
Kusumoto, F. M., Schoenfeld, M. H., Barrett, C., Edgerton, J. R., Ellenbogen, K. A., Gold, M. R., Goldschlager, N. F., Hamilton, R. M., Joglar, J. A., Kim, R. J., et al. (2019) · Circulation · doi:10.1161/cir.0000000000000628 · Cited by 603
Kusumoto, F. M., Schoenfeld, M. H., Barrett, C., Edgerton, J. R., Ellenbogen, K. A., Gold, M. R., Goldschlager, N. F., Hamilton, R. M., Joglar, J. A., Kim, R. J., et al. (2019) · Heart Rhythm · doi:10.1016/j.hrthm.2018.10.037 · Cited by 131
Dretzke, J., Toff, W. D., Lip, G. Y., Raftery, J., Fry-Smith, A., & Taylor, R. S. (2004) · Cochrane Database of Systematic Reviews · doi:10.1002/14651858.cd003710.pub2 · Cited by 89
Metadata
- Authors
- LUKL, J., DOUPAL, V., & HEINC, P.
- Venue
- Pacing and Clinical Electrophysiology
- Publisher
- Wiley
- Published
- 1994-11
- Volume
- 17
- Issue
- 11
- Pages
- 1844-1848
- ISSN
- 0147-8389 (ISSN-L)
- 1540-8159
- OpenAlex
- W1985527955
- Cited by
- 49
- References
- 0
- FWCI
- 2.10
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Quality‐of‐Life During DDD and Dual Sensor VVIR Pacing. https://app.cochat.ai/papers/10.1111/j.1540-8159.1994.tb03760.x
- The paper
- LUKL, J., DOUPAL, V., & HEINC, P. (1994). Quality‐of‐Life During DDD and Dual Sensor VVIR Pacing. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/j.1540-8159.1994.tb03760.x
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.