The DOI matches its Crossref record.
Polymerization Efficiency of LED Curing Lights
Abstract
ABSTRACT: Purpose: The purpose of this study was to compare the curing efficiency of three commercially available light‐emitting diode (LED)‐based curing lights with that of a quartz tungsten halogen (QTH) curing light by means of hardness testing. In addition, the power density (intensity) and spectral emission of each LED light was compared with the QTH curing light in both the 380‐to 520‐nm and the 450‐ to 500‐nm spectral ranges. Materials and Methods: A polytetrafluoroethylene mold 2 mm high and 8 mm in diameter was used to prepare five depth‐of‐cure test specimens for each combination of exposure duration, composite type (Silux Plus [microfill], Z‐100 [hybrid]), and curing light (ZAP Dual Curing™ Light, LumaCure™, VersaLux™, Optilux 401™). After 24 hours, Knoop hardness measurements were made for each side of the specimen, means were calculated, and a bottom/top Knoop hardness (B/T KH) percentage was determined. A value of at least 80% was used to indicate satisfactory polymerization. A linear regression of B/T KH percentage versus exposure duration was performed, and the resulting equation was used to predict the exposure duration required to produce a B/T KH percentage of 80% for the test conditions. The power densities (power/unit area) of the LED curing lights and the QTH curing light (Optilux 401™) were measured 1 mm from the target using a laboratory‐grade, laser power meter in both the full visible light spectrum range (380–780 nm) and the spectral range (between 450 and 500 nm), using a combination of long‐ and short‐wave edge filters. Results: The emission spectra of the LED lights more closely mirrored the absorption spectrum of the commonly used photoinitiator camphorquinone. Specifically, 95% of the emission spectrum of the VersaLux, 87% of the LumaCure, 84% of the ZAP LED, and 78% of the ZAP combination LED and QTH fell between 450 and 500 nm. In contrast, only 56% of the emission spectrum of the Optilux 401™ halogen lamp fell within this range. However, the power density between 450 and 500 nm was at least four times greater for the halogen lamp than for the purely LED lights. As I a result, the LED‐based curing lights required from 39 to 61 seconds to cure a 2‐mm thick hybrid I resin composite and between 83 and 131 seconds to adequately cure a microfill resin composite. By I comparison, the QTH light required only 21 and 42 seconds to cure the hybrid and microfill resin I composites, respectively.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Synthesis and Characterization of Photocurable Elastomers from Poly(glycerol-co-sebacate)
Nijst, C. L. E., Bruggeman, J. P., Karp, J. M., Ferreira, L., Zumbuehl, A., Bettinger, C. J., & Langer, R. (2007) · Biomacromolecules · doi:10.1021/bm070423u · Cited by 295
An Improvement of Led Lighting System Accuracy with Voltage Control System
Abdurohman, M., Nugraha, R., & Gautama Putrada, A. (2020) · 2020 Fifth International Conference on Informatics and Computing (ICIC) · doi:10.1109/icic50835.2020.9288576 · Cited by 7
Metadata
- Authors
- LEONARD, D. L., CHARLTON, D. G., ROBERTS, H. W., & COHEN, M. E.
- Venue
- Journal of Esthetic and Restorative Dentistry
- Publisher
- Wiley
- Published
- 2002-09
- Volume
- 14
- Issue
- 5
- Pages
- 286-295
- ISSN
- 1496-4155 (ISSN-L)
- 1708-8240
- OpenAlex
- W2169503522
- Cited by
- 170
- References
- 0
- FWCI
- 11.32
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Polymerization Efficiency of LED Curing Lights. https://app.cochat.ai/papers/10.1111/j.1708-8240.2002.tb00524.x
- The paper
- LEONARD, D. L., CHARLTON, D. G., ROBERTS, H. W., & COHEN, M. E. (2002). Polymerization Efficiency of LED Curing Lights. Journal of Esthetic and Restorative Dentistry. https://doi.org/10.1111/j.1708-8240.2002.tb00524.x
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.