APPLANATION TONOMETRY AND CENTRAL CORNEAL THICKNESS
The DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.
Abstract
Readings with the Goldmann applanation tonometer were made at various intraocular hydrostatic pressures and compared with central corneal thickness and radius in rabbit and in man. Linear correlations were established between hydrostatic pressure and applanation readings, with correlation coefficients close to 1.0. In rabbits the tonometer readings were generally too low. In human eyes with a normal corneal thickness tonometer readings and hydrostatic pressure coincided, with thick corneas the readings were too high, with thin corneas too low. The correlation between corneal thickness and the error of applanation tonometry (ΔP) was statistically highly significant (P < 0.001). No statistical correlation could be established between corneal radius and ΔP. Multiple regression, taking thickness as well as corneal radius into consideration, revealed only slightly higher correlation coefficients. It is concluded that the central corneal thickness is a parameter which should be taken into consideration when evaluating applanation tonometer readings. A Table is presented showing the correction to be added to the applanation reading at differing corneal thickness.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Goldmann, H., & Schmidt, T. (1957) · Ophthalmologica · doi:10.1159/000303213 · Cited by 607
BIOMETRIC CORRELATIONS OF CORNEAL THICKNESS
EHLERS, N., HANSEN, F. K., & AASVED, H. (1975) · Acta Ophthalmologica · doi:10.1111/j.1755-3768.1975.tb01784.x · Cited by 161
Self-citation
Cited by
Human Corneal Thickness and Its Impact on Intraocular Pressure Measures
Doughty, M. J., & Zaman, M. L. (2000) · Survey of Ophthalmology · doi:10.1016/s0039-6257(00)00110-7 · Cited by 1,362
The prevalence of primary open-angle glaucoma in Japanese
Iwase, A., Suzuki, Y., Araie, M., Yamamoto, T., Abe, H., Shirato, S., Kuwayama, Y., Mishima, H. K., Shimizu, H., Tomita, G., et al. (2004) · Ophthalmology · doi:10.1016/j.ophtha.2004.03.029 · Cited by 872
Liu, J., & Roberts, C. J. (2005) · Journal of Cataract and Refractive Surgery · doi:10.1016/j.jcrs.2004.09.031 · Cited by 753
Sources of error with use of Goldmann-type tonometers
Whitacre, M. M., & Stein, R. (1993) · Survey of Ophthalmology · doi:10.1016/0039-6257(93)90053-a · Cited by 747
Central corneal thickness in the ocular hypertension treatment study (OHTS)
Brandt, J. D., Beiser, J. A., Kass, M. A., & Gordon, M. O. (2001) · Ophthalmology · doi:10.1016/s0161-6420(01)00760-6 · Cited by 626
Metadata
- Authors
- EHLERS, N., BRAMSEN, T., & SPERLING, S.
- Venue
- Acta Ophthalmologica
- Publisher
- Wiley
- Published
- 1975-03
- Volume
- 53
- Issue
- 1
- Pages
- 34-43
- ISSN
- 0001-639X (ISSN-L)
- 1755-375X
- 1755-3768
- OpenAlex
- W2033867289
- Cited by
- 931
- References
- 2
- FWCI
- 3.69
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.