Sensitivity of near-infrared spectroscopy and diffuse correlation spectroscopy to brain hemodynamics: simulations and experimental findings during hypercapnia
Open accessAbstract
Near-infrared spectroscopy (NIRS) and diffuse correlation spectroscopy (DCS) are two diffuse optical technologies for brain imaging that are sensitive to changes in hemoglobin concentrations and blood flow, respectively. Measurements for both modalities are acquired on the scalp, and therefore hemodynamic processes in the extracerebral vasculature confound the interpretation of cortical hemodynamic signals. The sensitivity of NIRS to the brain versus the extracerebral tissue and the contrast-to-noise ratio (CNR) of NIRS to cerebral hemodynamic responses have been well characterized, but the same has not been evaluated for DCS. This is important to assess in order to understand their relative capabilities in measuring cerebral physiological changes. We present Monte Carlo simulations on a head model that demonstrate that the relative brain-to-scalp sensitivity is about three times higher for DCS (0.3 at 3 cm) than for NIRS (0.1 at 3 cm). However, because DCS has higher levels of noise due to photon-counting detection, the CNR is similar for both modalities in response to a physiologically realistic simulation of brain activation. Even so, we also observed higher CNR of the hemodynamic response during graded hypercapnia in adult subjects with DCS than with NIRS.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Ferrari, M., & Quaresima, V. (2012) · NeuroImage · doi:10.1016/j.neuroimage.2012.03.049 · Cited by 2,274
Neuroimaging-aided differential diagnosis of the depressive state
Takizawa, R., Fukuda, M., Kawasaki, S., Kasai, K., Mimura, M., Pu, S., Noda, T., Niwa, S.-I., & Okazaki, Y. (2014) · NeuroImage · doi:10.1016/j.neuroimage.2013.05.126 · Cited by 337
Cited by
Tachtsidis, I., & Scholkmann, F. (2016) · Neurophotonics · doi:10.1117/1.nph.3.3.031405 · Cited by 607
Dreier, J. P., Fabricius, M., Ayata, C., Sakowitz, O. W., William Shuttleworth, C., Dohmen, C., Graf, R., Vajkoczy, P., Helbok, R., Suzuki, M., et al. (2017) · Journal of Cerebral Blood Flow & Metabolism · doi:10.1177/0271678x16654496 · Cited by 350
Modified Beer-Lambert law for blood flow
Baker, W. B., Parthasarathy, A. B., Busch, D. R., Mesquita, R. C., Greenberg, J. H., & Yodh, A. G. (2014) · Biomedical Optics Express · doi:10.1364/boe.5.004053 · Cited by 283
Exploring early human brain development with structural and physiological neuroimaging
Vasung, L., Abaci Turk, E., Ferradal, S. L., Sutin, J., Stout, J. N., Ahtam, B., Lin, P.-Y., & Grant, P. E. (2019) · NeuroImage · doi:10.1016/j.neuroimage.2018.07.041 · Cited by 177
Sensor space group analysis for fNIRS data
Tak, S., Uga, M., Flandin, G., Dan, I., & Penny, W. D. (2016) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2016.03.003 · Cited by 83
Metadata
- Authors
- Selb, J., Boas, D. A., Chan, S.-T., Evans, K. C., Buckley, E. M., & Carp, S. A.
- Venue
- Neurophotonics
- Publisher
- SPIE-Intl Soc Optical Eng
- Published
- 2014-08-06
- Volume
- 1
- Issue
- 1
- Pages
- 015005
- ISSN
- 2329-423X (ISSN-L)
- OpenAlex
- W2125379859
- Cited by
- 165
- References
- 2
- FWCI
- 6.01
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Sensitivity of near-infrared spectroscopy and diffuse correlation spectroscopy to brain hemodynamics: simulations and experimental findings during hypercapnia. https://app.cochat.ai/papers/10.1117/1.nph.1.1.015005
- The paper
- Selb, J., Boas, D. A., Chan, S.-T., Evans, K. C., Buckley, E. M., & Carp, S. A. (2014). Sensitivity of near-infrared spectroscopy and diffuse correlation spectroscopy to brain hemodynamics: simulations and experimental findings during hypercapnia. Neurophotonics. https://doi.org/10.1117/1.nph.1.1.015005
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.