A computer simulation study comparing lesion detection accuracy with digital mammography, breast tomosynthesis, and cone‐beam CT breast imaging
Abstract
Although conventional mammography is currently the best modality to detect early breast cancer, it is limited in that the recorded image represents the superposition of a three‐dimensional (3D) object onto a 2D plane. Recently, two promising approaches for 3D volumetric breast imaging have been proposed, breast tomosynthesis (BT) and CT breast imaging (CTBI). To investigate possible improvements in lesion detection accuracy with either breast tomosynthesis or CT breast imaging as compared to digital mammography (DM), a computer simulation study was conducted using simulated lesions embedded into a structured 3D breast model. The computer simulation realistically modeled x‐ray transport through a breast model, as well as the signal and noise propagation through a CsI based flat‐panel imager. Polyenergetic x‐ray spectra of for digital mammography, for BT, and for CTBI were modeled. For the CTBI simulation, the intensity of the x‐ray spectra for each projection view was determined so as to provide a total average glandular dose of , which is approximately equivalent to that given in conventional two‐view screening mammography. The same total dose was modeled for both the DM and BT simulations. Irregular lesions were simulated by using a stochastic growth algorithm providing lesions with an effective diameter of . Breast tissue was simulated by generating an ensemble of backgrounds with a power law spectrum, with the composition of 50% fibroglandular and 50% adipose tissue. To evaluate lesion detection accuracy, a receiver operating characteristic (ROC) study was performed with five observers reading an ensemble of images for each case. The average area under the ROC curves was 0.76 for DM, 0.93 for BT, and 0.94 for CTBI. Results indicated that for the same dose, a lesion embedded in a structured breast phantom was detected by the two volumetric breast imaging systems, BT and CTBI, with statistically significant higher confidence than with planar digital mammography, while the difference in lesion detection between BT and CTBI was not statistically significant.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Digital breast tomosynthesis versus digital mammography: a clinical performance study
Gennaro, G., Toledano, A., di Maggio, C., Baldan, E., Bezzon, E., La Grassa, M., Pescarini, L., Polico, I., Proietti, A., Toffoli, A., et al. (2010) · European Radiology · doi:10.1007/s00330-009-1699-5 · Cited by 257
Badano, A., Graff, C. G., Badal, A., Sharma, D., Zeng, R., Samuelson, F. W., Glick, S. J., & Myers, K. J. (2018) · JAMA Network Open · doi:10.1001/jamanetworkopen.2018.5474 · Cited by 235
Self-citation
Metadata
- Authors
- Gong, X., Glick, S. J., Liu, B., Vedula, A. A., & Thacker, S.
- Venue
- Medical Physics
- Publisher
- Wiley
- Published
- 2006-03-23
- Volume
- 33
- Issue
- 4
- Pages
- 1041-1052
- ISSN
- 0094-2405 (ISSN-L)
- 2473-4209
- OpenAlex
- W2004298112
- Cited by
- 161
- References
- 0
- FWCI
- 8.86
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: A computer simulation study comparing lesion detection accuracy with digital mammography, breast tomosynthesis, and cone‐beam CT breast imaging. https://app.cochat.ai/papers/10.1118/1.2174127
- The paper
- Gong, X., Glick, S. J., Liu, B., Vedula, A. A., & Thacker, S. (2006). A computer simulation study comparing lesion detection accuracy with digital mammography, breast tomosynthesis, and cone‐beam CT breast imaging. Medical Physics. https://doi.org/10.1118/1.2174127
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.