Computational and human observer image quality evaluation of low dose, knowledge‐based CT iterative reconstruction
Open accessThe DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.
Abstract
Purpose: Aims in this study are to (1) develop a computational model observer which reliably tracks the detectability of human observers in low dose computed tomography (CT) images reconstructed with knowledge‐based iterative reconstruction (IMR™, Philips Healthcare) and filtered back projection (FBP) across a range of independent variables, (2) use the model to evaluate detectability trends across reconstructions and make predictions of human observer detectability, and (3) perform human observer studies based on model predictions to demonstrate applications of the model in CT imaging. Methods: Detectability (d′) was evaluated in phantom studies across a range of conditions. Images were generated using a numerical CT simulator. Trained observers performed 4‐alternative forced choice (4‐AFC) experiments across dose (1.3, 2.7, 4.0 mGy), pin size (4, 6, 8 mm), contrast (0.3%, 0.5%, 1.0%), and reconstruction (FBP, IMR), at fixed display window. A five‐channel Laguerre–Gauss channelized Hotelling observer (CHO) was developed with internal noise added to the decision variable and/or to channel outputs, creating six different internal noise models. Semianalytic internal noise computation was tested against Monte Carlo and used to accelerate internal noise parameter optimization. Model parameters were estimated from all experiments at once using maximum likelihood on the probability correct, P C. Akaike information criterion (AIC) was used to compare models of different orders. The best model was selected according to AIC and used to predict detectability in blended FBP‐IMR images, analyze trends in IMR detectability improvements, and predict dose savings with IMR. Predicted dose savings were compared against 4‐AFC study results using physical CT phantom images. Results: Detection in IMR was greater than FBP in all tested conditions. The CHO with internal noise proportional to channel output standard deviations, Model‐k4, showed the best trade‐off between fit and model complexity according to AICc. With parameters fixed, the model reasonably predicted detectability of human observers in blended FBP‐IMR images. Semianalytic internal noise computation gave results equivalent to Monte Carlo, greatly speeding parameter estimation. Using Model‐k4, the authors found an average detectability improvement of 2.7 ± 0.4 times that of FBP. IMR showed greater improvements in detectability with larger signals and relatively consistent improvements across signal contrast and x‐ray dose. In the phantom tested, Model‐k4 predicted an 82% dose reduction compared to FBP, verified with physical CT scans at 80% reduced dose. Conclusions: IMR improves detectability over FBP and may enable significant dose reductions. A channelized Hotelling observer with internal noise proportional to channel output standard deviation agreed well with human observers across a wide range of variables, even across reconstructions with drastically different image characteristics. Utility of the model observer was demonstrated by predicting the effect of image processing (blending), analyzing detectability improvements with IMR across dose, size, and contrast, and in guiding real CT scan dose reduction experiments. Such a model observer can be applied in optimizing parameters in advanced iterative reconstruction algorithms as well as guiding dose reduction protocols in physical CT experiments.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Strategies for Reducing Radiation Dose in CT
McCollough, C. H., Primak, A. N., Braun, N., Kofler, J., Yu, L., & Christner, J. (2009) · Radiologic Clinics of North America · doi:10.1016/j.rcl.2008.10.006 · Cited by 842
Cited by
Current Applications and Future Impact of Machine Learning in Radiology
Choy, G., Khalilzadeh, O., Michalski, M., Do, S., Samir, A. E., Pianykh, O. S., Geis, J. R., Pandharipande, P. V., Brink, J. A., & Dreyer, K. J. (2018) · Radiology · doi:10.1148/radiol.2018171820 · Cited by 800
“A model observer can be applied to optimize parameters and evaluate image quality of low-dose CT iterative reconstructions (65).”
Review of medical image quality assessment
Chow, L. S., & Paramesran, R. (2016) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2016.02.006 · Cited by 265
Automated reference-free detection of motion artifacts in magnetic resonance images
Küstner, T., Liebgott, A., Mauch, L., Martirosian, P., Bamberg, F., Nikolaou, K., Yang, B., Schick, F., & Gatidis, S. (2018) · Magnetic Resonance Materials in Physics Biology and Medicine · doi:10.1007/s10334-017-0650-z · Cited by 102
Pierre, K., Haneberg, A. G., Kwak, S., Peters, K. R., Hochhegger, B., Sananmuang, T., Tunlayadechanont, P., Tighe, P. J., Mancuso, A., & Forghani, R. (2023) · Seminars in Roentgenology · doi:10.1053/j.ro.2023.02.003 · Cited by 79
Metadata
- Authors
- Eck, B. L., Fahmi, R., Brown, K. M., Zabic, S., Raihani, N., Miao, J., & Wilson, D. L.
- Venue
- Medical Physics
- Publisher
- Wiley
- Published
- 2015-09-30
- Volume
- 42
- Issue
- 10
- Pages
- 6098-6111
- ISSN
- 0094-2405 (ISSN-L)
- 2473-4209
- OpenAlex
- W1849559129
- Cited by
- 46
- References
- 1
- FWCI
- 3.20
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Computational and human observer image quality evaluation of low dose, knowledge‐based CT iterative reconstruction. https://app.cochat.ai/papers/10.1118/1.4929973
- The paper
- Eck, B. L., Fahmi, R., Brown, K. M., Zabic, S., Raihani, N., Miao, J., & Wilson, D. L. (2015). Computational and human observer image quality evaluation of low dose, knowledge‐based CT iterative reconstruction. Medical Physics. https://doi.org/10.1118/1.4929973
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.