3D electron dose calculation using a Voxel based Monte Carlo algorithm (VMC)

Kawrakow, Iwan, Fippel, Matthias, Friedrich, Klaus·1996·Medical Physics

Abstract

A new model for calculating electron beam dose has been developed. The algorithm is based on a two‐ or three‐dimensional geometry defined by computerized tomography (CT) images. The Monte Carlo technique was used to solve the electron transport equation. However, in contrast to conventional Monte Carlo models (EGS4) several approximations and simplifications in the description of elementary electron processes were introduced reducing in this manner the computational time by a factor of about 35 without significant loss in accuracy. The Monte Carlo computer program does not need any precalculated data. The random access memory required is about 16 Mbytes for a 128 2×50 matrix, depending on the resolution of the CT cube. The Voxel Monte Carlo model (VMC) was tested in comparison to calculations by EGS4 and the “Hogstrom algorithm” (MDAH) using several fictive phantoms. In all cases a good coincidence has been found between EGS4 and VMC, especially near tissue inhomogeneities, whereas the MDAH algorithm has produced dose underestimations of up to 40%.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1118/1.597673

Cites

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

See all 322

Metadata

Authors
Kawrakow, I., Fippel, M., & Friedrich, K.
Venue
Medical Physics
Publisher
Wiley
Published
1996-04
Volume
23
Issue
4
Pages
445-457
ISSN
  • 0094-2405 (ISSN-L)
  • 2473-4209
OpenAlex
W2018372062
Cited by
322
References
0
FWCI
10.90
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: 3D electron dose calculation using a Voxel based Monte Carlo algorithm (VMC). https://app.cochat.ai/papers/10.1118/1.597673
The paper
Kawrakow, I., Fippel, M., & Friedrich, K. (1996). 3D electron dose calculation using a Voxel based Monte Carlo algorithm (VMC). Medical Physics. https://doi.org/10.1118/1.597673

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.