The development of a "reduced logMAR" visual acuity chart for use in routine clinical practice

Open access
Rosser, D A·2001·British Journal of Ophthalmology

Abstract

BACKGROUND/AIMS: The advantages of logMAR acuity data over the Snellen fraction are well known, and yet existing logMAR charts have not been adopted into routine ophthalmic clinical use. As this may be due in part to the time required for a logMAR measurement, this study was performed to determine whether an abbreviated logMAR chart design could combine the advantages of existing charts with a clinically acceptable measurement time. METHODS: The test-retest variability, agreement (with the gold standard), and time taken for "single letter" (interpolated) acuity measurements taken using three prototype "reduced logMAR" (RLM) charts and the Snellen chart were compared with those of the ETDRS chart which acted as the gold standard. The Snellen chart was also scored with the more familiar "line assignment" method. The subjects undergoing these measurements were drawn from a typical clinical outpatient population exhibiting a range of acuities. RESULTS: The RLM A prototype chart achieved a test-retest variability of +/-0.24 logMAR compared with +/-0.18 for the ETDRS chart. Test-retest variability for the Snellen chart was +/-0.24 logMAR using clinically prohibitive "single letter" scoring increasing to +/-0.33 with the more usual "line assignment" method. All charts produced acuity data which agreed well with those of the ETDRS chart. "Single letter" acuity measurements using the prototype RLM charts were completed in approximately half the time of those taken using the ETDRS and Snellen charts. The duration of a Snellen "line assignment" measurement was not evaluated. CONCLUSION: The RLM A chart offers an acceptable level of test-retest variability when compared with the gold standard ETDRS chart, while reducing the measurement time by half. Also, by allowing a faster, less variable acuity measurement than the Snellen chart, the RLM A chart can bring the benefits of logMAR acuity to routine clinical practice.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1136/bjo.85.4.432

Cites

2 of 2 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 2 references

Cited by

See all 201

Metadata

Authors
Rosser, D. A.
Venue
British Journal of Ophthalmology
Publisher
BMJ
Published
2001-04-01
Volume
85
Issue
4
Pages
432-436
ISSN
0007-1161 (ISSN-L)
OpenAlex
W2157639616
Cited by
201
References
2
FWCI
3.80
Type
journal article
Language
en
Open access
bronze

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: The development of a "reduced logMAR" visual acuity chart for use in routine clinical practice. https://app.cochat.ai/papers/10.1136/bjo.85.4.432
The paper
Rosser, D. A. (2001). The development of a "reduced logMAR" visual acuity chart for use in routine clinical practice. British Journal of Ophthalmology. https://doi.org/10.1136/bjo.85.4.432

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.