Student or educator? Sign up with your school email and get 6 months of Pro free.
Cited by 254

The DOI matches its Crossref record; 4 of the 4 references it cites were found in a registry or an index.

Early complications of permanent pacemaker implantation: no difference between dual and single chamber systems.

Open access
Aggarwal, R. K., Connelly, D. T., Ray, S. G., Ball, J., Charles, R. G.·1995·Heart

Abstract

OBJECTIVE: To evaluate the incidence of intraoperative and early postoperative complications (up to two months after implant) of endocardial permanent pacemaker insertion in all patients under-going a first implant at a referral centre. METHODS: Prospective evaluation of all endocardial pacemaker implantation procedures performed from April 1992 to January 1994 carried out by completion of standard audit form at implant. Patients' demographic data, medical history, details of pacemaker hardware used, and any complications were noted. Follow up information was also collected prospectively onto standard forms at pacemaker outpatient clinic. SETTING: United Kingdom tertiary referral cardiothoracic centre. PATIENTS: 1088 consecutive patients underwent implantation of their first endocardial permanent pacemaker from April 1992 to January 1994. Implant and follow up data were available for 1059 (97.3%) patients at analysis. The median (range) age was 77 years (16-99); 51.2 % were male. RESULTS: Dual chamber units were implanted in 54.1% of patients, single chamber atrial in 5.2%, and ventricular in 40.7%. A temporary pacing lead was present at implant in 22.9% of patients. Most (93.6%) implants were performed via the subclavian vein. Immediate complications were rare: eight (0.8%) patients developed pneumothorax requiring medical treatment and 11 (1.0%) an insignificant pneumothorax. There was no significant difference in the pneumothorax rate for dual chamber (DDD) compared with single chamber systems. Arterial puncture without sequelae was documented in 2.7% of attempts at subclavian vein cannulation. A total of 35 patients (3.3%) required reoperation; the reoperation rate for dual chamber (3.5%) was similar to that for single chamber (3.1%) systems. Electrode displacement (n = 15, 1.4%) was the most common reason for reoperation. Atrial lead displacement (n = 10, 1.6% of atrial leads) was significantly more common than ventricular lead displacement (n = 5, 0.5% of ventricular leads, P = 0.047). There was no difference in electrode displacement rates for dual (1.6%) compared with single (1.2%) chamber systems. Pacemaker pocket infection led to reoperation in 10 patients (six dual, four single chamber, P = not significant) and was significantly more common in patients who had a temporary pacing lead in place at implant (2.9%) than in those who did not (0.4%, P = 0.0014). Five patients (0.5%) required reoperation for generator erosion (two dual, three single chamber, P = not significant). and a further five for drainage of haematoma or a serous fluid collection (three dual, two single chamber, P = not significant). Complications that did not require reoperation were also rare. Undersensing occurred in 10 patients (0.9%). Atrial undersensing (n = 8) was significantly more common than ventricular undersensing (n = 2, P = 0.017). All patients were successfully treated by reprogramming of sensitivity. Superficial wound infection was treated successfully with antibiotics in nine patients (six dual, three single chamber, P = not significant). Three patients with DDD generators developed sustained atrial fibrillation: two required reprogramming to VVI mode and one required cardioversion. CONCLUSIONS: Permanent pacing in a large tertiary referral centre with experienced operators carries a low risk. Infection rates are low, < 1% overall but significantly higher in patients who undergo temporary pacing before implantation. Lead displacement and undersensing are more likely to occur with atrial than ventricular leads. The overall complication rate for dual chamber pacing, however, is no higher than for single chamber pacing.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1136/hrt.73.6.571

Cites

4 of 4 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 4 references

Cited by

See all 254

Metadata

Authors
Aggarwal, R. K., Connelly, D. T., Ray, S. G., Ball, J., & Charles, R. G.
Venue
Heart
Publisher
BMJ
Published
1995-06-01
Volume
73
Issue
6
Pages
571-575
ISSN
1355-6037 (ISSN-L)
OpenAlex
W2135897962
Cited by
254
References
4
FWCI
5.91
Type
journal article
Language
en
Open access
bronze

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Early complications of permanent pacemaker implantation: no difference between dual and single chamber systems. https://app.cochat.ai/papers/10.1136/hrt.73.6.571
The paper
Aggarwal, R. K., Connelly, D. T., Ray, S. G., Ball, J., & Charles, R. G. (1995). Early complications of permanent pacemaker implantation: no difference between dual and single chamber systems. Heart. https://doi.org/10.1136/hrt.73.6.571

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.